High-resolution transthoracic real-time three-dimensional echocardiography -: Quantitation of cardiac volumes and function using semi-automatic border detection and comparison with cardiac magnetic resonance imaging

被引:216
作者
Kuhl, HP
Schreckenberg, M
Rulands, D
Katoh, M
Schäfer, W
Schummers, G
Bücker, A
Hanrath, P
Franke, A
机构
[1] Univ Hosp, Med Klin 1, Aachen, Germany
[2] Univ Hosp, Dept Diagnost Radiol, Aachen, Germany
[3] Univ Hosp, Dept Nucl Med, Aachen, Germany
[4] TomTec Imaging Syst GmbH, Unterschleissheim, Germany
关键词
D O I
10.1016/j.jacc.2004.01.037
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
OBJECTIVES We sought to validate high-resolution transthoracic. real-time (RT) three-dimensional echocardiography (3DE), in combination with a novel semi-automatic contour detection algorithm, for the assessment of left ventricular (LV) volumes and function in patients. BACKGROUND Quantitative RT-3DE has been limited by impaired image quality and time-consuming manual data analysis. METHODS Twenty-four subjects with abnormal (n = 14) or normal (n = 10) LVs were investigated. The results for end-diastolic volume (EDV), end-systolic volume (ESV), and ejection fraction (EF) obtained by manual tracing were compared with the results determined by the semi-automatic border detection algorithm. Moreover, the results of the semi-automatic method were compared with volumes and EF obtained by cardiac magnetic resonance imaging (CMRI). RESULTS Excellent correlation coefficients (r = 0.98 to 0.99) and low variability (EDV - 1.3 +/- 8.6 ml; ESV -0.2 +/- 5.4 ml; EF -0.1 +/- 2.7%; p = NS) were observed between the semiautomatically and manually assessed data. The RT-3DE data correlated highly with CMRI (r = 0.98). However, LV volumes were underestimated by RT-3DE compared with CMRI (EDV - 13.6 +/- 18.9 ml, p = 0.002; ESV - 12.8 +/- 20.5 ml, p = 0.005). The difference for EF was not significant between the two methods (EF 0.9 +/- 4.4%, p = NS). Observer variability was acceptable, and repeatability of the method was excellent. CONCLUSIONS The RT-3DE, in combination with a semi-automatic contour tracing algorithm, allows accurate determination of cardiac volumes and function compared with both manual tracing and CMRI. High repeatability suggests applicability of the method for the serial follow-up of patients with cardiac disease. (C) 2004 by the American College of Cardiology Foundation.
引用
收藏
页码:2083 / 2090
页数:8
相关论文
共 30 条
[1]   Real-time three-dimensional dobutamine stress echocardiography in assessment of ischemia: Comparison with two-dimensional dobutamine stress echocardiography [J].
Ahmad, M ;
Xie, TR ;
McCulloch, M ;
Abreo, G ;
Runge, M .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2001, 37 (05) :1303-1309
[2]  
[Anonymous], 1990, SPLINE MODELS OBSERV
[3]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[4]  
Buck T, 1997, CIRCULATION, V96, P4286
[5]   ACC/AHA/ASE 2003 guideline update for the clinical application of echocardiography: Summary article [J].
Cheitlin, MD ;
Armstrong, WF ;
Aurigemma, GP ;
Beller, GA ;
Bierman, FZ ;
Davis, JL ;
Douglas, PS ;
Faxon, DP ;
Gillam, LD ;
Kimball, TR ;
Kussmaul, WG ;
Pearlman, AS ;
Philbrick, JT ;
Rakowski, H ;
Thys, DM ;
Antman, EM ;
Smith, SC ;
Alpert, JS ;
Gregoratos, G ;
Anderson, JL ;
Hiratzka, LF ;
Faxon, DP ;
Hunt, SA ;
Fuster, V ;
Jacobs, AK ;
Gibbons, RJ ;
Russell, RO .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2003, 42 (05) :954-970
[6]   Left ventricular volume estimation for real-time three-dimensional echocardiography [J].
Corsi, C ;
Saracino, G ;
Sarti, A ;
Lamberti, C .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2002, 21 (09) :1202-1208
[7]   Mortality and morbidity of aortic regurgitation in clinical practice - A long-term follow-up study [J].
Dujardin, KS ;
Enriquez-Sarano, M ;
Schaff, HV ;
Bailey, KR ;
Seward, JB ;
Tajik, AJ .
CIRCULATION, 1999, 99 (14) :1851-1857
[8]   Transthoracic high-resolution real-time three-dimensional echocardiography for the analysis of regional left-ventricular wall motion patterns in patients with cardiac resynchronization therapy [J].
Franke, A ;
Rulands, D ;
Breithardt, OA ;
Brehmer, K ;
Sinha, AM ;
Markus, KU ;
Stellbrink, C ;
Kühl, HP .
EUROPEAN HEART JOURNAL, 2003, 24 :396-396
[9]   Freehand three-dimensional echocardiography for determination of left ventricular volume and mass in patients with abnormal ventricles: Comparison with magnetic resonance imaging [J].
Gopal, AS ;
Schnellbaecher, MJ ;
Shen, ZQ ;
Boxt, LM ;
Katz, J ;
King, DL .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 1997, 10 (08) :853-861
[10]   ASSESSMENT OF CARDIAC-FUNCTION BY 3-DIMENSIONAL ECHOCARDIOGRAPHY COMPARED WITH CONVENTIONAL NONINVASIVE METHODS [J].
GOPAL, AS ;
SHEN, ZQ ;
SAPIN, PM ;
KELLER, AM ;
SCHNELLBAECHER, MJ ;
LEIBOWITZ, DW ;
AKINBOBOYE, OO ;
RODNEY, RA ;
BLOOD, DK ;
KING, DL .
CIRCULATION, 1995, 92 (04) :842-853