Echocardiographic Assessment of Prosthetic Heart Valves

被引:13
作者
Blauwet, Lori A. [1 ]
Miller, Fletcher A., Jr. [1 ]
机构
[1] Mayo Clin, Div Cardiovasc Dis, Rochester, MN 55905 USA
关键词
Echocardiography; Doppler; Prosthesis; Heart valves; LONG-TERM SURVIVAL; COMPREHENSIVE HEMODYNAMIC ASSESSMENT; MEDICAL MECHANICAL PROSTHESES; VELOCITY INTEGRAL RATIO; PATIENT MISMATCH; DOPPLER-ECHOCARDIOGRAPHY; EUROPEAN ASSOCIATION; REGISTERED BRANCH; AMERICAN SOCIETY; CANADIAN SOCIETY;
D O I
10.1016/j.pcad.2014.05.001
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Valvular heart disease is a global health problem. It is estimated that more than 280,000 prosthetic heart valves are implanted worldwide each year. As the world's population is aging, the incidence of prosthetic heart valve implantation and the prevalence of prosthetic heart valves continue to increase. Assessing heart valve prosthesis function remains challenging, as prosthesis malfunction is unpredictable but not uncommon. Transthoracic two-dimensional and Doppler echocardiography is the preferred method for assessing prosthetic valve function. Clinically useful Doppler-derived measures for assessing prosthetic valve hemodynamic profiles have been reported for aortic, mitral, and tricuspid valve prostheses, but echocardiographic data regarding pulmonary valve prostheses remain limited. Complete prosthetic valve evaluation by transthoracic echocardiography (TTE) is sometimes challenging due to acoustic shadowing and artifacts. In these cases, further imaging with transesophageal echocardiography, fluoroscopy and/or gated CT may be warranted, particularly if prosthetic valve dysfunction is suspected. Being able to differentiate pathologic versus functional obstruction of an individual prosthesis is extremely important, as this distinction affects management decisions. Transprosthetic and periprosthetic regurgitation may be difficult to visualize on TTE, so careful review of Doppler-derived data combined with a high index of suspicion is warranted, particularly in symptomatic patients. A baseline TTE soon after valve implantation is indicated in order to "fingerprint" the prosthesis hemodynamic profile. It remains unclear how frequently serial imaging should be performed in order to assess prosthetic valve function, as this issue has not been systematically studied. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:100 / 110
页数:11
相关论文
共 50 条
[21]   Cinefluoroscopy for assessment of mechanical heart valves with suspected dysfunction [J].
Derda, Anselm A. ;
Marquardt, Marvin M. ;
Martens, Andreas ;
Mirena, Elion ;
Vogel-Claussen, Jens ;
Kempf, Tibor ;
Haverich, Axel ;
Bauersachs, Johann ;
Napp, L. Christian .
FRONTIERS IN CARDIOVASCULAR MEDICINE, 2022, 9
[22]   Lack of Accessible Data on Prosthetic Heart Valves [J].
Frank, Michelle ;
Ganzoni, Giulia ;
Starck, Christoph ;
Gruenenfelder, Juerg ;
Corti, Roberto ;
Gruner, Christiane ;
Huerlimann, David ;
Tanner, Felix C. ;
Jenni, Rolf ;
Greutmann, Matthias ;
Biaggi, Patric .
INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2016, 32 (03) :439-447
[23]   Evaluation of hemolysis in patients with prosthetic heart valves [J].
Maraj, R ;
Jacobs, LE ;
Ioli, A ;
Kotler, MN .
CLINICAL CARDIOLOGY, 1998, 21 (06) :387-392
[24]   Assessment of Complex Multi-Valve Disease and Prosthetic Valves [J].
Venneri, Lucia ;
Khattar, Rajdeep S. ;
Senior, Roxy .
HEART LUNG AND CIRCULATION, 2019, 28 (09) :1436-1446
[25]   Comprehensive Hemodynamic Assessment of 368 Normal St. Jude Medical Mechanical Mitral Valve Prostheses Based on Early Postimplantation Echocardiographic Studies [J].
Blauwet, Lori A. ;
Malouf, Joseph F. ;
Connolly, Heidi M. ;
Hodge, David O. ;
Herges, Regina M. ;
Suri, Rakesh M. ;
Miller, Fletcher A., Jr. .
JOURNAL OF THE AMERICAN SOCIETY OF ECHOCARDIOGRAPHY, 2013, 26 (04) :381-389
[26]   Echocardiographic assessment of the right heart in adults: a practical guideline from the British Society of Echocardiography [J].
Zaidi, Abbas ;
Knight, Daniel S. ;
Augustine, Daniel X. ;
Harkness, Allan ;
Oxborough, David ;
Pearce, Keith ;
Ring, Liam ;
Robinson, Shaun ;
Stout, Martin ;
Willis, James ;
Sharma, Vishal .
ECHO RESEARCH AND PRACTICE, 2020, 7 (01) :G19-G41
[27]   Echocardiographic Assessment of Pulmonary Artery Systolic Pressure and Outcomes in Ambulatory Heart Failure Patients [J].
Kalogeropoulos, Andreas P. ;
Siwamogsatham, Sarawut ;
Hayek, Salim ;
Li, Song ;
Deka, Anjan ;
Marti, Catherine N. ;
Georgiopoulou, Vasiliki V. ;
Butler, Javed .
JOURNAL OF THE AMERICAN HEART ASSOCIATION, 2014, 3 (01) :e000363
[28]   Echocardiographic Assessment of the Right Heart in Mice [J].
Brittain, Evan ;
Penner, Niki L. ;
West, James ;
Hemnes, Anna .
JOVE-JOURNAL OF VISUALIZED EXPERIMENTS, 2013, (81)
[29]   Echocardiographic Evaluation of Hemodynamics in Patients With Decompensated Systolic Heart Failure [J].
Nagueh, Sherif F. ;
Bhatt, Rajat ;
Vivo, Rey P. ;
Krim, Selim R. ;
Sarvari, Sebastian Imre ;
Russell, Kristoffer ;
Edvardsen, Thor ;
Smiseth, Otto A. ;
Estep, Jerry D. .
CIRCULATION-CARDIOVASCULAR IMAGING, 2011, 4 (03) :220-227
[30]   Assessment of magnetic field (4.7 T) induced forces on prosthetic heart valves and annuloplasty rings [J].
Edwards, MB ;
Ordidge, RJ ;
Hand, JW ;
Taylor, KM ;
Young, IR .
JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 22 (02) :311-317