Glucose concentrations in blood and tissue - a pilot study on variable time lag

被引:5
作者
Chlup, Rudolf [1 ,2 ,3 ,4 ]
Krejci, Jan [5 ]
O'Connell, Mark [6 ]
Sebestova, Blanka [5 ]
Plicka, Robert [5 ]
Jezova, Lucie [5 ]
Brozova, Tereza [5 ,7 ]
Doubravova, Blanka [4 ,5 ]
Zalesakova, Hana [4 ,5 ]
Durajkova, Emilia [4 ,5 ]
Vojtek, Jiri [5 ]
Bartek, Josef [8 ]
机构
[1] Palacky Univ, Fac Med & Dent, Dept Physiol, CR-77147 Olomouc, Czech Republic
[2] Palacky Univ, Fac Med & Dent, Dept Internal Med Gastroenterol & Hepatol 2, CR-77147 Olomouc, Czech Republic
[3] Univ Hosp, Olomouc, Czech Republic
[4] Inst Neurol & Geriatr, Moravsky Beroun, Czech Republic
[5] BVT Technol As, Strazek, Czech Republic
[6] Probe Sci Ltd, Bedford, England
[7] Brno Univ Technol, Fac Mech Engn, Heat Transfer & Fluid Flow Lab, CS-61090 Brno, Czech Republic
[8] Palacky Univ, Fac Med & Dent, Dept Med Chem & Biochem, CR-77147 Olomouc, Czech Republic
来源
BIOMEDICAL PAPERS-OLOMOUC | 2015年 / 159卷 / 04期
关键词
diabetes mellitus; continuous glucose monitoring; lag phase; glucose transport; calibration; TYPE-2; DIABETIC-PATIENTS; GLYCEMIC VARIABILITY; MONITORING DATA; CLOSED-LOOP; MANAGEMENT; QUALITY; SYSTEM; HBA1C;
D O I
10.5507/bp.2015.007
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Aim. The aim of this pilot study was to acquire insight into the parameters of glycaemic control, especially, (1) the time delay (lag phase) between plasma and tissue glucose concentrations in relation to rise and fall in glucose levels and (2) the rate of glucose increase and decrease. Methods. Four healthy people (HP), 4 people with type 1diabetes (DM1) and 4 with type 2 diabetes (DM2) underwent concurrent glucose measurements by means of (1) the continuous glucose monitoring system (CGMS-Medtronic), Medtronic-Minimed, CA, USA, calibrated by the glucometer Calla, Wellion, Austria, and, (2) the Beckman II analyser to measure glucose concentrations in venous plasma. Samples were taken on 4 consecutive days in the fasting state and 4 times after consumption of 50 g glucose. Carelink Personal, MS Excel, Maple and Mat lab were applied to plot the evolution of glucose concentration and analyse the results. The time difference between increase and decrease was calculated for HP, DM 1 and DM 2. Results. In DM1 and DM2, glucose tolerance testing (GTT) resulted in slower transport of glucose into subcutaneous tissue than in HP where the lag phase lasted up to 12 min. The maximum increase/decrease rates in DM1 and DM2 vs HP were 0.25 vs < 0.1 mmol/L/min. Conclusion. CGMS is shown to provide reliable plasma glucose concentrations provided the system is calibrated during a steady state. The analysis of glucose change rates improves understanding of metabolic processes better than standard GTT.
引用
收藏
页码:527 / 534
页数:8
相关论文
共 35 条
  • [1] Outpatient assessment of Karlsburg diabetes management system-based decision support
    Augstein, Petra
    Vogt, Lutz
    Kohnert, Klaus-Dieter
    Freyse, Ernst-Joachim
    Heinke, Peter
    Salzsieder, Eckhard
    [J]. DIABETES CARE, 2007, 30 (07) : 1704 - 1708
  • [2] The use and efficacy of continuous glucose monitoring in type 1 diabetes treated with insulin pump therapy: a randomised controlled trial
    Battelino, T.
    Conget, I.
    Olsen, B.
    Schuetz-Fuhrmann, I.
    Hommel, E.
    Hoogma, R.
    Schierloh, U.
    Sulli, N.
    Bolinder, J.
    [J]. DIABETOLOGIA, 2012, 55 (12) : 3155 - 3162
  • [3] Sustained Benefit of Continuous Glucose Monitoring on A1C, Glucose Profiles, and Hypoglycemia in Adults With Type 1 Diabetes
    Beck, Roy W.
    Xing, Dongyuan
    Gilliam, Lisa
    Hirsch, Irl
    Kollman, Craig
    Laffel, Lori
    Ruedy, Katrina J.
    Tamborlane, William V.
    Weinzimer, Stuart
    Wolpert, Howard
    [J]. DIABETES CARE, 2009, 32 (11) : 2047 - 2049
  • [4] Recommendations for Standardizing Glucose Reporting and Analysis to Optimize Clinical Decision Making in Diabetes: The Ambulatory Glucose Profile (AGP)
    Bergenstal, Richard M.
    Ahmann, Andrew J.
    Bailey, Timothy
    Beck, Roy W.
    Bissen, Joan
    Buckingham, Bruce
    Deeb, Larry
    Dolin, Robert H.
    Garg, Satish K.
    Goland, Robin
    Hirsch, Irl B.
    Klonoff, David C.
    Kruger, Davida F.
    Matfin, Glenn
    Mazze, Roger S.
    Olson, Beth A.
    Parkin, Christopher
    Peters, Anne
    Powers, Margaret A.
    Rodriguez, Henry
    Southerland, Phil
    Strock, Ellie S.
    Tamborlane, William
    Wesley, David M.
    [J]. DIABETES TECHNOLOGY & THERAPEUTICS, 2013, 15 (03) : 198 - 211
  • [5] Timing of changes in interstitial and venous blood glucose measured with a continuous subcutaneous glucose sensor
    Boyne, MS
    Silver, DM
    Kaplan, J
    Saudek, CD
    [J]. DIABETES, 2003, 52 (11) : 2790 - 2794
  • [6] Continuous glucose monitoring a novel approach to the determination of the glycaemic index of foods (DEGIF 1) -: Determination of the glycaemic index of foods by means of the CGMS
    Chlup, R
    Jelenová, D
    Kudlová, P
    Chlupová, K
    Bartek, J
    Zapletalová, J
    Langová, K
    Chlupová, L
    [J]. EXPERIMENTAL AND CLINICAL ENDOCRINOLOGY & DIABETES, 2006, 114 (02) : 68 - 74
  • [7] Effective Assessment of Diabetes Control Using Personal Glucometers (CONTOURLINK, Bayer, Germany; CALLA, Wellion, Austria; LINUS, Agamatrix, USA)
    Chlup, R.
    Doubravova, B.
    Bartek, J.
    Zapletalova, J.
    Krystynik, O.
    Prochazka, V.
    [J]. DISEASE MARKERS, 2013, 2013 : 895 - 905
  • [8] Chlup R, 2014, KLIN FARMAKOL FARMAC, V28, P72
  • [9] Chlup Rudolf, 2010, J Diabetes Sci Technol, V4, P615
  • [10] Cohen Ohad, 2009, J Diabetes Sci Technol, V3, P804