Precordial electrode placement for optimal ECG monitoring: Implications for ambulatory monitor devices and event recorders

被引:20
作者
Nedios, Sotirios [1 ]
Romero, Inaki [2 ]
Gerds-Li, Jin-Hong [3 ]
Fleck, Eckard [3 ]
Kriatselis, Charalampos [3 ]
机构
[1] Univ Leipzig, Ctr Heart, Dept Electrophysiol, D-04289 Leipzig, Germany
[2] Holst Ctr IMEC, Eindhoven, Netherlands
[3] German Heart Inst Berlin, Dept Internal Med & Cardiol, Berlin, Germany
关键词
Ambulatory; Patient monitoring; Electrocardiography (ECG); Atrial fibrillation; ATRIAL-FIBRILLATION; SURFACE ELECTROCARDIOGRAMS; DIAGNOSIS;
D O I
10.1016/j.jelectrocard.2014.04.003
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Introduction: Detection of QRS complexes, P-waves and atrial fibrillation f-waves in electrocardiographic (ECG) signals is critical for the correct diagnosis of arrhythmias. We aimed to find the best bipolar lead (BL) with the highest signal amplitude and shortest inter-electrode spacing. Methods: ECG signals (120 seconds) were recorded in 36 patients with 16 precordial electrodes placed in a standardized pattern. An average signal was analysed for each of 120 possible BLs obtained by calculating the difference between pairs of unipolar leads. Peak-to-peak amplitudes of QRS waves (50 ms around R-peak) and P waves (270-70 ms before R-peak) were calculated. For patients with atrial fibrillation, power of the fibrillatory (f) wave was used instead. Maximum values at each distance were considered and differentiation analysis was performed based on incremental changes (amplitude to distance). Results: There was a significant correlation between distance and QRS-amplitude (r = 0.78, p < 0.001), P-wave amplitude (r = 0.60, p < 0.01) and f-wave power (r = 0.79, p < 0.001). The range of values was: QRS-amplitude 0.7-2.33 mV, P-wave amplitude 0.07-0.18 mV, and f-wave power 0.55-2.12 mV(2)/s. The maximum value for the shortest distance was on a heart-aligned axis over the left ventricle for the QRS complex (1.9 mV at 8.7 cm) and over the atria for the P-wave (0.98 mV) and f-waves (1.45 mV(2)/s at 8 cm, respectively). Conclusion: There is a strong positive correlation between electrode distance and ECG signal-amplitude. Distance of 8 cm on a heart-aligned axis and over the relevant heart-chamber provides the highest signal amplitude for the shortest distance. These findings are essential for the design and use of ambulatory monitoring devices. (C) 2014 Elsevier Inc. All rights reserved.
引用
收藏
页码:669 / 676
页数:8
相关论文
共 27 条
  • [21] DIAGNOSIS OF ATRIAL-FIBRILLATION FROM SURFACE ELECTROCARDIOGRAMS BASED ON COMPUTER-DETECTED ATRIAL ACTIVITY
    SLOCUM, J
    SAHAKIAN, A
    SWIRYN, S
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 1992, 25 (01) : 1 - 8
  • [22] Diagnostic Utility of a Novel Leadless Arrhythmia Monitoring Device
    Turakhia, Mintu P.
    Hoang, Donald D.
    Zimetbaum, Peter
    Miller, Jared D.
    Froelicher, Victor F.
    Kumar, Uday N.
    Xu, Xiangyan
    Yang, Felix
    Heidenreich, Paul A.
    [J]. AMERICAN JOURNAL OF CARDIOLOGY, 2013, 112 (04) : 520 - 524
  • [23] Symptomatic and asymptomatic atrial fibrillation in patients undergoing radiofrequency catheter ablation
    Vasamreddy, CR
    Dalal, D
    Dong, J
    Cheng, A
    Spragg, D
    Lamiy, SZ
    Meininger, G
    Henrikson, CA
    Marine, JE
    Berger, R
    Calkins, H
    [J]. JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2006, 17 (02) : 134 - 139
  • [24] Optimum lead positioning for recording bipolar atrial electrocardiograms during sinus rhythm and atrial fibrillation
    Waktare, JEP
    Gallagher, MM
    Murtagh, A
    Camm, AJ
    Malik, M
    [J]. CLINICAL CARDIOLOGY, 1998, 21 (11) : 825 - 830
  • [25] Evaluation and outcomes of patients with palpitations
    Weber, BE
    Kapoor, WN
    [J]. AMERICAN JOURNAL OF MEDICINE, 1996, 100 (02) : 138 - 148
  • [26] Ultra-Low-Power Wearable Biopotential Sensor Nodes
    Yazicioglu, R. F.
    Torfs, T.
    Penders, J.
    Romero, I.
    Kim, H.
    Merken, P.
    Gyselinckx, B.
    Yoo, H. J.
    Van Hoof, C.
    [J]. 2009 ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-20, 2009, : 3205 - 3208
  • [27] Ambulatory Arrhythmia Monitoring Choosing the Right Device
    Zimetbaum, Peter
    Goldman, Alena
    [J]. CIRCULATION, 2010, 122 (16) : 1629 - 1636