Multiparameter techniques for non-invasive measurement of blood glucose

被引:21
作者
Amaral, Carlos F. [1 ]
Brischwein, Martin [1 ]
Wolf, Bernhard [1 ]
机构
[1] Tech Univ Munich, Heinz Nixdorf Lehrstuhl Med Elekt, D-80333 Munich, Germany
关键词
Bioimpedance; Blood glucose monitoring; Non-invasive measurement; Absorption spectroscopy; NEAR-INFRARED SPECTROSCOPY; PARTIAL LEAST-SQUARES; TISSUE;
D O I
10.1016/j.snb.2009.04.023
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Although painless glycemic control could improve life quality of patients, enabling a better regulation of hyper- and hypoglycaemia episodes, none of the available commercial devices offer enough precision to replace lancet approaches. The aim of this work is to describe the technologies of absorption spectroscopy and bioimpedance applied simultaneously to transcutaneous glycemic measurement. Basic electrical and optic theories were studied systematically and in following experiments glucose characteristics were measured in blood. Light spectroscopy studies were done in MIR ranges (5-40 mu m), while complex bioimpedance measurements were scanned from 100 Hz to 30 MHz. Parallel to above parameters, information of epidermis temperature (conductance and IR radiation) was also collected, improving prediction quality. Cutaneous assays showed correlation of 0.43 for impedance and 0.46 for optic spectroscopy alone. The multiparameter analysis allowed the improvement of correlation to 0.57, after data treatment with Partial Least Square and neural networks method in Matlab. (C) 2009 Elsevier B.V. All rights reserved.
引用
收藏
页码:12 / 16
页数:5
相关论文
共 26 条
[1]   Noninvasive glucose sensing [J].
Arnold, MA ;
Small, GW .
ANALYTICAL CHEMISTRY, 2005, 77 (17) :5429-5439
[2]  
Burmeister JJ, 1999, CLIN CHEM, V45, P1621
[3]   First human experiments with a novel non-invasive, non-optical continuous glucose monitoring system [J].
Caduff, A ;
Hirt, E ;
Feldman, Y ;
Ali, Z ;
Heinemann, L .
BIOSENSORS & BIOELECTRONICS, 2003, 19 (03) :209-217
[4]   Influence of contact state on NIR diffuse reflectance spectroscopy in vivo [J].
Chen, WL ;
Liu, R ;
Xu, KX ;
Wang, RKK .
JOURNAL OF PHYSICS D-APPLIED PHYSICS, 2005, 38 (15) :2691-2695
[5]   Current development in non-invasive glucose monitoring [J].
do Amaral, Carlos Eduardo Ferrante ;
Wolf, Benhard .
MEDICAL ENGINEERING & PHYSICS, 2008, 30 (05) :541-549
[6]  
GEDDES LA, 1989, PRINCIPLES APPL BIOM, P537
[7]   Multivariate determination of glucose concentrations from optimally filtered frequency-warped NIR spectra of human blood serum [J].
Ham, FM ;
Cohen, GM ;
Kostanic, I ;
Gooch, BR .
PHYSIOLOGICAL MEASUREMENT, 1996, 17 (01) :1-20
[8]  
HEISE HM, 2000, ENCY ANAL CHEM, V1, P1
[9]  
JOSSINET J, 1992, P 14 ANN INT C IEEE, P1712
[10]   Improvement of partial least squares models for in vitro and in vivo glucose quantifications by using near-infrared spectroscopy and searching combination moving window partial least squares [J].
Kasemsumran, Sumaporn ;
Du, Yi Ping ;
Maruo, Katsuhiko ;
Ozaki, Yukhiro .
CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2006, 82 (1-2) :97-103