Optical surface guidance for submillimeter monitoring of patient position during frameless stereotactic radiotherapy

被引:38
作者
Covington, Elizabeth L. [1 ]
Fiveash, John B. [1 ]
Wu, Xingen [1 ]
Brezovich, Ivan [1 ]
Willey, Christopher D. [1 ]
Riley, Kristen [2 ]
Popple, Richard A. [1 ]
机构
[1] Univ Alabama Birmingham, Dept Radiat Oncol, Birmingham, AL 35294 USA
[2] Univ Alabama Birmingham, Dept Neurosurg, Birmingham, AL USA
关键词
surface guided radiotherapy; stereotactic radiosurgery; intrafraction motion; optical surface imaging; frameless radiosurgery; LINEAR-ACCELERATOR; GUIDED RADIOSURGERY; REAL-TIME; ACCURACY; SYSTEM;
D O I
10.1002/acm2.12611
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose To evaluate the accuracy of monitoring intrafraction motion during stereotactic radiotherapy with the optical surface monitoring system. Prior studies showing a false increase in the magnitude of translational offsets at non-coplanar couch positions prompted the vendor to implement software changes. This study evaluated two software improvements intended to address false offsets. Methods The vendor implemented two software improvements: a volumetric (ACO) rather than planar calibration and, approximately 6 months later, an improved calibration workflow (CIB) designed to better compensate for thermal drift. Offsets relative to the reference position, obtained at table angle 0 following image-guided setup, were recorded before beam-on at each table position and at the end of treatment the table returned to 0 degrees for patients receiving SRT. Results Prior to ACO, between ACO and CIB, and after CIB, 223, 155, and 436 fractions were observed respectively. The median magnitude of translational offsets at the end of treatment was similar for all three intervals: 0.29, 0.33, and 0.27 mm. Prior to ACO, the offset magnitude for non-zero table positions had a median of 0.79 mm and was found to increase with increasing distance from isocenter to the anterior patient surface. After ACO, the median magnitude was 0.74 mm, but the dependence on surface-to-isocenter distance was eliminated. After CIB, the median magnitude for non-zero table positions was reduced to 0.57 mm. Conclusion Ongoing improvements in software and calibration procedures have decreased reporting of false offsets at non-zero table angles. However, the median magnitude for non-zero table angles is larger than that observed at the end of treatment, indicating that accuracy remains better when the table is not rotated.
引用
收藏
页码:91 / 98
页数:8
相关论文
共 22 条
[1]   Intrafraction variations in linac-based image-guided radiosurgery of intracranial lesions [J].
Badakhshi, H. ;
Barelkowski, T. ;
Wust, P. ;
Budach, V. ;
Boehmer, D. ;
Graf, R. .
CANCER RADIOTHERAPIE, 2013, 17 (07) :664-667
[2]   A phantom evaluation of a stereo-vision surface imaging system for radiotherapy patient setup [J].
Bert, C ;
Metheany, KG ;
Doppke, K ;
Chen, GTY .
MEDICAL PHYSICS, 2005, 32 (09) :2753-2762
[3]   The university of Florida frameless high-precision stereotactic radiotherapy system [J].
Bova, FJ ;
Buatti, JM ;
Friedman, WA ;
Mendenhall, WM ;
Yang, CC ;
Liu, C .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1997, 38 (04) :875-882
[4]   Stereotactic radiosurgery with MLC-defined arcs: Verification of dosimetry, spatial accuracy, and end-to-end tests [J].
Brezovich, Ivan A. ;
Wu, Xingen ;
Popple, Richard A. ;
Covington, Elizabeth ;
Cardan, Rex ;
Shen, Sui ;
Fiveash, John ;
Bredel, Markus ;
Guthrie, Barton .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2019, 20 (05) :84-98
[5]   Initial clinical experience with a frameless and maskless stereotactic radiosurgery treatment [J].
Cervino, Laura I. ;
Detorie, Nicole ;
Taylor, Matthew ;
Lawson, Joshua D. ;
Harry, Taylor ;
Murphy, Kevin T. ;
Mundt, Arno J. ;
Jiang, Steve B. ;
Pawlicki, Todd A. .
PRACTICAL RADIATION ONCOLOGY, 2012, 2 (01) :54-62
[6]   Frame-less and mask-less cranial stereotactic radiosurgery: a feasibility study [J].
Cervino, Laura I. ;
Pawlicki, Todd ;
Lawson, Joshua D. ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (07) :1863-1873
[7]   Using surface imaging and visual coaching to improve the reproducibility and stability of deep-inspiration breath hold for left-breast-cancer radiotherapy [J].
Cervino, Laura I. ;
Gupta, Sonia ;
Rose, Mary A. ;
Yashar, Catheryn ;
Jiang, Steve B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2009, 54 (22) :6853-6865
[8]   Evaluation of radiotherapy setup accuracy for head and neck cancer using a 3-D surface imaging system [J].
Cho, H. -L. ;
Park, E. -T. ;
Kim, J. -Y. ;
Kwak, K. -S. ;
Kim, C. -J. ;
Ahn, K. -J. ;
Suh, T. -S. ;
Lee, Y. -K. ;
Kim, S. -W. ;
Kim, J. -K. ;
Lim, S. ;
Choi, Y. -M. ;
Park, S. -K. .
JOURNAL OF INSTRUMENTATION, 2013, 8
[9]   FEASIBILITY OF SINGLE-ISOCENTER VOLUMETRIC MODULATED ARC RADIOSURGERY FOR TREATMENT OF MULTIPLE BRAIN METASTASES [J].
Clark, Grant M. ;
Popple, Richard A. ;
Young, P. Edward ;
Fiveash, John B. .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2010, 76 (01) :296-302
[10]   The spatial accuracy of two frameless, linear accelerator-based systems for single-isocenter, multitarget cranial radiosurgery [J].
Ezzell, Gary A. .
JOURNAL OF APPLIED CLINICAL MEDICAL PHYSICS, 2017, 18 (02) :37-43