Do manual cutting guides for total knee arthroplasty introduce systematic error?

被引:11
作者
Hohmann, Erik [1 ,2 ]
Tetsworth, Kevin [3 ,4 ]
机构
[1] Univ Queensland, Dept Orthopaed Surg, Sch Clin Med, Brisbane, Qld, Australia
[2] CQ Univ, Musculoskeletal Res Unit, POB 4045, Rockhampton, Qld 4700, Australia
[3] Royal Brisbane Hosp, Dept Orthopaed Surg, Herston, Qld, Australia
[4] Univ Queensland, Sch Med, Dept Surg, Brisbane, Qld, Australia
关键词
Accuracy of cutting guides; Total knee arthroplasty; Computer-assisted navigation; FEMORAL COMPONENTS; NAVIGATION SYSTEM; ALIGNMENT; ACCURACY; REPLACEMENT; BONE; PLACEMENT; VARUS;
D O I
10.1007/s00264-015-2963-8
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Introduction Conventional cutting guides in total knee arthroplasty can potentially cause unintentional deviation from the planned direction and depth of bone resection resulting in malaligned components. The purpose of this study was therefore to investigate the accuracy of bone cutting jigs for both the femur and tibia using imageless navigation. Material and methods A total of 125 patients with a mean age of 66.7 +/- 9.9 years underwent primary total knee arthroplasty with a Stryker Triathlon (TM) fixed bearing posterior cruciate retaining implant using imageless navigation. Coronal and sagittal position of the secured cutting jig was recorded and bone resection was checked with a rectangular probe attached to a navigation tracker. Results There were significant within group differences for the femoral sagittal cut (mean delta = 0.9 degrees [31 %]; p = 0.00001), femoral depth medial compartment (mean delta = 0.5 mm [5 %]; p = 0.001), femoral depth lateral compartment (mean delta = 0.7 mm [7 %]; p = 0.00001), proximal tibial cut (mean delta = 0.3 mm [25 %]; p = 0.001), tibial depth medial compartment (mean delta = 0.6 mm [10 %]; p = 0.0001) and tibia depth lateral cut (mean delta = 0.4 mm [5 %]; p = 0.002). Deviation of more than 2 degrees was observed for the distal cut in the sagittal plane in 17 % and in 9.6 % for the proximal tibial cut in the sagittal plane of all patients. Conclusion The results of this study demonstrated significant differences between the dialed in cut and "actual" bone resection achieved for all planes for both the femur and tibia. The femur sagittal cut demonstrated a tendency for an extended cut and the tibia showed a tendency for varus.
引用
收藏
页码:277 / 284
页数:8
相关论文
共 23 条
  • [1] Intraoperative cutting errors in total knee arthroplasty
    Bäthis, H
    Perlick, L
    Tingart, M
    Perlick, C
    Lüring, C
    Grifka, J
    [J]. ARCHIVES OF ORTHOPAEDIC AND TRAUMA SURGERY, 2005, 125 (01) : 16 - 20
  • [2] The accuracy of bone resections made during computer navigated total knee replacement. Do we resect what the computer plans we resect?
    Biant, Leela C.
    Yeoh, Kwan
    Walker, Peter M.
    Bruce, Warwick J. M.
    Walsh, William R.
    [J]. KNEE, 2008, 15 (03) : 238 - 241
  • [3] Alignment deviation between bone resection and final implant positioning in computer-navigated total knee arthroplasty
    Catani, Fabio
    Biasca, Nicola
    Ensini, Andrea
    Leardini, Alberto
    Bianchi, Luca
    Digennaro, Vitantonio
    Giannini, Sandro
    [J]. JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A (04) : 765 - 771
  • [4] Randomized control trial comparing radiographic total knee arthroplasty implant placement using computer navigation versus conventional technique
    Chin, PL
    Yang, KY
    Yeo, SJ
    Lo, NN
    [J]. JOURNAL OF ARTHROPLASTY, 2005, 20 (05) : 618 - 626
  • [5] DUNBAR MJ, 1995, INT ORTHOP, V19, P265
  • [6] What is the optimal alignment of the tibial and femoral components in knee arthroplasty? An overview of the literature
    Gromov, Kirill
    Korchi, Mounim
    Thomsen, Morten G.
    Husted, Henrik
    Troelsen, Anders
    [J]. ACTA ORTHOPAEDICA, 2014, 85 (05) : 480 - 487
  • [7] Computer-assisted navigation increases precision of component placement in total knee arthroplasty
    Haaker, RG
    Stockheim, M
    Kamp, M
    Proff, G
    Breitenfelder, J
    Ottersbach, A
    [J]. CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2005, (433) : 152 - 159
  • [8] Cutting and implanting errors in minimally invasive total knee arthroplasty using a navigation system
    Hasegawa, Masahiro
    Yoshida, Kakunoshin
    Wakabayashi, Hiroki
    Sudo, Akihiro
    [J]. INTERNATIONAL ORTHOPAEDICS, 2013, 37 (01) : 27 - 30
  • [9] Accuracy of manual instrumentation of tibial cutting guide in total knee arthroplasty
    Iorio, R.
    Bolle, G.
    Conteduca, F.
    Valeo, L.
    Conteduca, J.
    Mazza, D.
    Ferretti, A.
    [J]. KNEE SURGERY SPORTS TRAUMATOLOGY ARTHROSCOPY, 2013, 21 (10) : 2296 - 2300
  • [10] James TP, 2013, J MED DEVICES, V7