The value of ElastPQ for the evaluation of liver stiffness in patients with B and C chronic hepatopathies

被引:25
作者
Mare, Ruxandra [1 ]
Sporea, Loan [1 ]
Lupusoru, Raluca [1 ]
Sirli, Roxana [1 ]
Popescu, Alina [1 ]
Danila, Mirela [1 ]
Pienar, Corina [1 ]
机构
[1] Victor Babes Univ Med & Pharm Timisoara, Dept Gastroenterol & Hepatol, 156 Liviu Rebreanu Bv, Timisoara 300723, Romania
关键词
Transient Elastography; ElastPQ; Liver stiffness; Chronic hepatitis B; Chronic hepatitis C; SHEAR-WAVE ELASTOGRAPHY; CHRONIC HEPATITIS-C; TRANSIENT ELASTOGRAPHY; FIBROSIS ASSESSMENT; GOLD STANDARD; ACCURACY; BIOPSY; FEASIBILITY; DIAGNOSIS; PROBE;
D O I
10.1016/j.ultras.2017.02.005
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
The aim of this study was to evaluate the diagnostic performance of a point shear wave elastography using ARFI technique - ElastPQ in patients with B and C chronic hepatopathies, using Transient Elastography (TE) as the reference method, since it is a validated method for liver fibrosis assessment. Subjects and methods: the study included 228 consecutive subjects with chronic hepatopathies (26% HBV, 74% HCV) from whom 51% had liver cirrhosis. Liver stiffness (LS) was evaluated in the same session by means of 2 elastographic methods: TE (FibroScan, EchoSens) and ElastPQ (Affinity, Philips) techniques. For TE 10 valid LS measurements were performed for each patient and the median value was calculated. Reliable LS measurements by TE (M or XL probe) were considered the median value of 10 LS measurements with a success rate >60% and an interquartile range <30%. For ElastPQ we calculated the median value of 10 LS measurements in the liver parenchyma, at least 1 cm below the capsule, avoiding large vessels. For differentiating between stages of liver fibrosis we used the TE cut-off values published in the Tsochatzis meta -analysis: significant fibrosis (F >= 2)- 7.0 kPa, severe fibrosis (F >= 3)- 9.5 kPa and for liver cirrhosis (F = 4)-12 kPa (Tsochatzis et al., 2011). The areas under the receiver operating characteristic curve (AUROCs) were used to assess the diagnostic performance of ElastPQ correlations between ElastPQ and TE were evaluated. Results: Valid LS measurements were obtained in 90.7% (207/228) cases by means of TE and in 98.7% (225/228) cases with ElastPQ. In the final analysis 205 patients were included. The ElastPQ values ranged from 232 to 44.07 kPa (median = 10.42 kPa). Based on TE cut-off values (Tsochatzis et al., 2011) we divided our cohort into 4 groups: F0-F1:61/205 (29.8%); F2: 14/205 (6.8%); F3: 15/205 (7.3%); F = 4: 115/205 (56.1%). The best cut-off values for discriminating, significant, severe fibrosis and cirrhosis were 7.2, 8.5 and 8.9 kPa, respectively. The AUROCs were calculated considering TE as the reference method: 0.94 for significant fibrosis (F >= 2), 0.97 for severe fibrosis (F >= 3) and 0.97 for cirrhosis (F = 4). In our cohort there was a strong correlation between measurements obtained by Transient Elastography and ElastPQ (r = 0.85, p < 0.001). Conclusions: ElastPQ seems to have a good diagnostic accuracy for staging liver fibrosis. (C) 2017 Elsevier B.V. All rights reserved.
引用
收藏
页码:144 / 151
页数:8
相关论文
共 30 条
  • [1] [Anonymous], 2015, J HEPATOL, V63, P237, DOI 10.1016/j.jhep.2015.04.006
  • [2] NONINVASIVE IN VIVO LIVER FIBROSIS EVALUATION USING SUPERSONIC SHEAR IMAGING: A CLINICAL STUDY ON 113 HEPATITIS C VIRUS PATIENTS
    Bavu, Eric
    Gennisson, Jean-Luc
    Couade, Mathieu
    Bercoff, Jeremy
    Mallet, Vincent
    Fink, Mathias
    Badel, Anne
    Vallet-Pichard, Anais
    Nalpas, Bertrand
    Tanter, Mickael
    Pol, Stanislas
    [J]. ULTRASOUND IN MEDICINE AND BIOLOGY, 2011, 37 (09) : 1361 - 1373
  • [3] Sampling variability of liver fibrosis in chronic hepatitis C
    Bedossa, P
    Dargère, D
    Paradis, V
    [J]. HEPATOLOGY, 2003, 38 (06) : 1449 - 1457
  • [4] Value of acoustic radiation force impulse elastography for the assessment of ascites syndrome
    Bota, Simona
    Sporea, Ioan
    Sirli, Roxana
    Popescu, Alina
    Danila, Mirela
    Sendroiu, Madalina
    [J]. WORLD JOURNAL OF RADIOLOGY, 2011, 3 (08): : 205 - 209
  • [5] Current concepts: Liver biopsy.
    Bravo, AA
    Sheth, SG
    Chopra, S
    [J]. NEW ENGLAND JOURNAL OF MEDICINE, 2001, 344 (07) : 495 - 500
  • [6] Prospective comparison of transient elastography, fibrotest, APRI, and liver biopsy for the assessment of fibrosis in chronic hepatitis C
    Castéra, L
    Vergniol, J
    Foucher, J
    Le Bail, B
    Chanteloup, E
    Haaser, M
    Darriet, M
    Couzigou, P
    De Lédinghen, V
    [J]. GASTROENTEROLOGY, 2005, 128 (02) : 343 - 350
  • [7] Pitfalls of Liver Stiffness Measurement: A 5-Year Prospective Study of 13,369 Examinations
    Castera, Laurent
    Foucher, Juliette
    Bernard, Pierre-Henri
    Carvalho, Francoise
    Allaix, Daniele
    Merrouche, Wassil
    Couzigou, Patrice
    de Ledinghen, Victor
    [J]. HEPATOLOGY, 2010, 51 (03) : 828 - 835
  • [8] EFSUMB Guidelines and Recommendations on the Clinical Use of Ultrasound Elastography. Part 2: Clinical Applications
    Cosgrove, D.
    Piscaglia, F.
    Bamber, J.
    Bojunga, J.
    Correas, J. -M.
    Gilja, O. H.
    Klauser, A. S.
    Sporea, I.
    Calliada, F.
    Cantisani, V.
    D'Onofrio, M.
    Drakonaki, E. E.
    Fink, M.
    Friedrich-Rust, M.
    Fromageau, J.
    Havre, R. F.
    Jenssen, C.
    Ohlinger, R.
    Saftoiu, A.
    Schaefer, F.
    Dietrich, C. F.
    [J]. ULTRASCHALL IN DER MEDIZIN, 2013, 34 (03): : 238 - 253
  • [9] Feasibility of liver transient elastography with FibroScan® using a new probe for obese patients
    de Ledinghen, Victor
    Vergniol, Julien
    Foucher, Juliette
    El-Hajbi, Farid
    Merrouche, Wassil
    Rigalleau, Vincent
    [J]. LIVER INTERNATIONAL, 2010, 30 (07) : 1043 - 1048
  • [10] Accuracy of the ElastPQ® Technique for the Assessment of Liver Fibrosis in Patients with Chronic Hepatitis C: a "Real Life" Single Center Study
    Ferraioli, Giovanna
    Maiocchi, Laura
    Lissandrin, Raffaella
    Tinelli, Carmine
    De Silvestri, Annalisa
    Filice, Carlo
    [J]. JOURNAL OF GASTROINTESTINAL AND LIVER DISEASES, 2016, 25 (03) : 331 - 335