The purpose of this study was to determine whether steady-state free precession (SSFP) could improve accuracy of geometric models for evaluation of left ventricular (LV)function in comparison to turbo gradient echo (TGrE) and thereby reduce the acquisition and post-processing times, which are commonly long by use of the Simpson's Rule. In 25 subjects, cine loops of the complete heart in short and horizontal long-axis planes were acquired using TGrE (TR/TE/flip = 5.0/1.9/25) compared with SSFP (TR/TE/flip = 3.2/1.2/60). LV volumes and EF were measured with various geometric models for TGrE and SSFP. With three-dimensional data, the LV volumes were higher and the resulting EF lower for SSFP in contrast to TGrE (51 +/- 15% vs. 57 +/- 15%, p < 0.001). With SSFP, various geometric models yielded good to excellent correlations for LV volumes and LVEF compared to volumetric data (r = 0.94 - 0.98, mean relative difference 7.0-11.4%). In contrast, correlations were low using biplane or single-plane ellipsoid models in TGrE (r = 0.71 - 0.75, mean relative difference 15.9-30.2%). A new combined geometric model, taking all three dimensions into account, yielded the highest accuracy for SSFP in comparison to volumetric data (r = 0.99, mean relative difference 4.7%). Geometric models for assessment of LV volumes and EF yield higher accuracy and reproducibility by use of the SSFP sequence than by standard TGrE. This may increase clinical utility of magnetic resonance by shorter acquisition and processing times.
引用
收藏
页码:327 / 339
页数:13
相关论文
共 50 条
  • [1] Quantitative assessment of left ventricular function: Steady-state free precession MR imaging with or without sensitivity encoding
    Kacere, RD
    Pereyra, M
    Nemeth, MA
    Muthupillai, R
    Flamm, SD
    RADIOLOGY, 2005, 235 (03) : 1031 - 1035
  • [2] Understanding steady-state free precession: A geometric perspective
    Dharmakumar, R
    Wright, GA
    CONCEPTS IN MAGNETIC RESONANCE PART A, 2005, 26A (01) : 1 - 10
  • [3] LEFT-VENTRICULAR EJECTION FRACTION DURING INCREMENTAL AND STEADY-STATE EXERCISE
    FOSTER, C
    GAL, RA
    PORT, SC
    SCHMIDT, DH
    MEDICINE AND SCIENCE IN SPORTS AND EXERCISE, 1995, 27 (12): : 1602 - 1606
  • [4] Quantitative diffusion imaging with steady-state free precession
    Deoni, SCL
    Peters, TM
    Rutt, BK
    MAGNETIC RESONANCE IN MEDICINE, 2004, 51 (02) : 428 - 433
  • [5] Validating the Gold-standard: Steady-State Free Precession Sequence for Assessment of Left Ventricular and Right Ventricular Mass
    Santos-Gallego, Carlos G.
    Njerve, Ida U.
    Ishikawa, Kiyotake
    Aguero, Jaime
    Vahl, Torsten
    Picatoste, Belen
    Hammoudi, Nadjib
    Najjar, Roger J.
    Fuster, Valentin
    Badimon, Juan
    Sanz, Javier
    CIRCULATION, 2014, 130
  • [6] Fat in left ventricular myocardium assessed by steady-state free precession pulse sequences
    Giovanni Donato Aquaro
    Gaetano Nucifora
    Laura Pederzoli
    Elisabetta Strata
    Daniele De Marchi
    Giancarlo Todiere
    Barison Andrea
    Alessandro Pingitore
    Massimo Lombardi
    The International Journal of Cardiovascular Imaging, 2012, 28 : 813 - 821
  • [7] Fat in left ventricular myocardium assessed by steady-state free precession pulse sequences
    Aquaro, Giovanni Donato
    Nucifora, Gaetano
    Pederzoli, Laura
    Strata, Elisabetta
    De Marchi, Daniele
    Todiere, Giancarlo
    Andrea, Barison
    Pingitore, Alessandro
    Lombardi, Massimo
    INTERNATIONAL JOURNAL OF CARDIOVASCULAR IMAGING, 2012, 28 (04): : 813 - 821
  • [8] Steady state free precession cardiovascular magnetic resonance: accuracy of left and right ventricular functional assessment in children
    Jennifer A Bryant
    Keith Godfrey
    Mark A Hanson
    Lucy Davies
    Alison Fletcher
    Charles Peebles
    Journal of Cardiovascular Magnetic Resonance, 15 (Suppl 1)
  • [9] Comparison of left ventricular volumetric assessment by standard steady-state free precession and real time cine imaging
    Parke, K.
    Wormmeighton, J.
    Mccann, G.
    Xue, H.
    Kellman, P.
    Arnold, R.
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2019, 20 : 332 - 332
  • [10] Improved reliability in left ventricular ejection fraction assessment by adopting quantitative methods
    Hassan, Saamir A.
    Lopez-Mattei, Juan
    Sanchez, Liza
    Banchs, Jose
    ECHOCARDIOGRAPHY-A JOURNAL OF CARDIOVASCULAR ULTRASOUND AND ALLIED TECHNIQUES, 2016, 33 (12): : 1936 - 1937