Fracture of the GORE HELEX Septal Occluder: Associated Factors and Clinical Outcomes

被引:27
作者
Fagan, Thomas [1 ]
Dreher, Dennis [2 ]
Cutright, Warren [2 ]
Jacobson, Joth [2 ]
Latson, Larry [3 ]
机构
[1] Univ Colorado, Childrens Hosp, Div Pediat Cardiol, Aurora, CO USA
[2] WL Gore & Associates, Flagstaff, AZ USA
[3] Cleveland Clin Fdn, Div Pediat Cardiol, Cleveland, OH 44195 USA
关键词
congenital heart disease; atrial septal defect; septal occluder device; frame fracture; catheterization; PATENT FORAMEN OVALE; TRANSCATHETER CLOSURE; PERCUTANEOUS CLOSURE; CATHETER ABLATION; CARDIAC CYCLE; ATRIAL; DEVICE; DEFECT; OCCLUSION; COMPLICATIONS;
D O I
10.1002/ccd.21929
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Objectives: This study examines factors associated with, and clinical effects of, wire frame fractures (WFF) of the GORE HELEX (TM) Septal Occluder (HELEX). Methods: Investigator-reported data from every HELEX implanted in the United States between 4/2000 and 4/2005 were reviewed. Clinical and procedural data from patients who experienced WFF were compared with data from patients who did not. The echocardiographic and fluoroscopic images for HELEX with WFF were reviewed for predictors of WFF and alterations in HELEX function. Results: With 90% of subjects followed for >12 months, 19/298 (6.4%) HELEX implanted were found to have WFF. Thirty WFFs were observed, with multiple WFFs occurring in 8/19 HELEX. Univariate predictors of WFF were large device size (P = 0.0003) and balloon defect size (P = 0.001); however, large device size (P = 0.0003) was the only significant predictor of WFF by multivariate analysis. WFF of the 30-mm and 35-mm HELEX accounted for 84% (16/19) of all WFFs. Review of HELEX images with WFF revealed all WFF occurred along the circumferential wire, except one (straight portion of locking loop). Seventeen of 29 (59%) circumferential WFFs were on the right disk. Residual defect status remained clinically insignificant (6/19) or became completely occluded (13/19) during follow-up. There were no clinical sequelae due to WFF; however, secondary to right atrial disk mobility, the HELEX with locking loop WFF was percutaneously removed 6 weeks after implantation. Conclusions: WFF occurred in 6.4% of HELEX and were most common in large devices. With the exception of one device removed for theoretical risks, no clinical sequelae were related to WFF. WFF did not alter the function of the HELEX. (C) 2009 Wiley-Liss, Inc.
引用
收藏
页码:941 / 948
页数:8
相关论文
共 31 条
[1]   Erosion of amplatzer septal occluder device after closure of secundum atrial septal defects: Review of registry of complications and recommendations to minimize future risk [J].
Amin, Z ;
Hijazi, ZM ;
Bass, JL ;
Cheatham, JP ;
Hellenbrand, WE ;
Kleinman, CS .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2004, 63 (04) :496-502
[2]   Surgical removal of atrial septal defect occlusion system-devices [J].
Bohm, J ;
Bittigau, K ;
Kohler, F ;
Baumann, G ;
Konertz, W .
EUROPEAN JOURNAL OF CARDIO-THORACIC SURGERY, 1997, 12 (06) :869-872
[3]  
Carminati M, 2000, CARDIOL YOUNG, V10, P519
[4]  
Carminati M, 2001, J Interv Cardiol, V14, P319, DOI 10.1111/j.1540-8183.2001.tb00339.x
[5]   Development of aorta-to-right atrial fistula following closure of secundum atrial septal defect using the Amplatzer septal occluder [J].
Chun, DS ;
Turrentine, MW ;
Moustapha, A ;
Hoyer, MH .
CATHETERIZATION AND CARDIOVASCULAR INTERVENTIONS, 2003, 58 (02) :246-251
[6]  
Delaney Jeffrey W, 2006, Congenit Heart Dis, V1, P202, DOI 10.1111/j.1747-0803.2006.00036.x
[7]   Cardiac perforation after device closure of atrial septal defects with the Amplatzer septal occluder [J].
Divekar, A ;
Gaamangwe, T ;
Shaikh, N ;
Raabe, M ;
Ducas, J .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 45 (08) :1213-1218
[8]   Migration of an amplatzer septal occluder device for closure of atrial septal defect into the ascending aorta with formation of an aorta-to-right atrial fistula [J].
Grayburn, PA ;
Schwartz, B ;
Anwar, A ;
Hebeler, RF .
AMERICAN JOURNAL OF CARDIOLOGY, 2005, 96 (11) :1607-1609
[9]   Transcatheter closure of atrial-septal defects and patent foramen ovale in adults:: Optimal anatomic adaptation of occlusion device [J].
Hoepp, HW ;
Deutsch, HJ ;
La Rosée, K ;
Schnabel, P ;
Terheggen, G ;
Schneider, CA ;
Korsten, J ;
Babic, UU .
AMERICAN HEART JOURNAL, 1999, 138 (05) :941-949
[10]   Closure of patent foramen ovale for paradoxical emboli: Intermediate-term risk of recurrent neurological events following transcatheter device placement [J].
Hung, J ;
Landzberg, MJ ;
Jenkins, KJ ;
King, MEE ;
Lock, JE ;
Palacios, IF ;
Lang, P .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (05) :1311-1316