Biomechanical evaluation of two commonly used external fixators in the treatment of open subtalar dislocation-A finite element analysis

被引:31
作者
Ramlee, Muhammad Hanif [1 ]
Kadir, Mohammed Rafiq Abdul [1 ]
Murali, Malliga Raman [2 ]
Kamarul, Tunku [2 ]
机构
[1] Univ Teknol Malaysia, Fac Biosci & Med Engn, Med Devices & Technol Grp MEDITEG, Johor Baharu 81310, Johor, Malaysia
[2] Univ Malaya, Fac Med, NOCERAL, Dept Orthopaed Surg,TEG, Kuala Lumpur 50603, Malaysia
关键词
Finite element analysis; External fixator; Subtalar dislocation; Stability; Biomechanics; ARTICULAR-CARTILAGE; PROXIMAL FEMUR; FIXATION; ANKLE; FRACTURES; FOOT; PIN; TOPOGRAPHY; THICKNESS; LIGAMENTS;
D O I
10.1016/j.medengphy.2014.07.001
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Subtalar dislocation is a rare injury caused by high-energy trauma. Current treatment strategies include leg casts, internal fixation and external fixation. Among these, external fixators are the most commonly used as this method is believed to provide better stabilization. However, the biomechanical stability provided by these fixators has not been demonstrated. This biomechanical study compares two commonly used external fixators, i.e. Mitkovic and Delta. CT imaging data were used to reconstruct three-dimensional models of the tibia, fibula, talus, calcaneus, navicular, cuboid, three cuneiforms and five metatarsal bones. The 3D models of the bones and cartilages were then converted into four-noded linear tetrahedral elements, whilst the ligaments were modelled with linear spring elements. Bones and cartilage were idealized as homogeneous, isotropic and linear. To simulate loading during walking, axial loading (70 N during the swing and 350 N during the stance phase) was applied at the end of diaphyseal tibia. The results demonstrate that the Mitkovic fixator produced greater displacement (peak 3.0 mm and 15.6 mm) compared to the Delta fixator (peak 0.8 mm and 3.9 mm), in both the swing and stance phase, respectively. This study demonstrates that the Delta external fixator provides superior stability over the Mitkovic fixator. The Delta fixator may be more effective in treating subtalar dislocation. (C) 2014 IPEM. Published by Elsevier Ltd. All rights reserved.
引用
收藏
页码:1358 / 1366
页数:9
相关论文
共 64 条
[1]   Three-dimensional distribution of articular cartilage thickness in the elderly talus and calcaneus analyzing the subchondral bone plate density [J].
Akiyama, K. ;
Sakai, T. ;
Sugimoto, N. ;
Yoshikawa, H. ;
Sugamoto, K. .
OSTEOARTHRITIS AND CARTILAGE, 2012, 20 (04) :296-304
[2]   Static and dynamic optimization solutions for gait are practically equivalent [J].
Anderson, FC ;
Pandy, MG .
JOURNAL OF BIOMECHANICS, 2001, 34 (02) :153-161
[3]  
Ansah M, 2000, FOOT ANKLE SURG, V6, P179
[4]  
Athanasiou KA, 1998, CLIN ORTHOP RELAT R, P269
[5]   Mechanical and functional assessment of the wrist affected by rheumatoid arthritis: A finite element analysis [J].
Bajuri, M. N. ;
Kadir, Mohammed Rafiq Abdul ;
Raman, Murali Malliga ;
Kamarul, T. .
MEDICAL ENGINEERING & PHYSICS, 2012, 34 (09) :1294-1302
[6]  
Bali Kamal, 2011, Foot Ankle Surg, V17, pe40, DOI 10.1016/j.fas.2011.04.005
[7]   Hip contact forces and gait patterns from routine activities [J].
Bergmann, G ;
Deuretzbacher, G ;
Heller, M ;
Graichen, F ;
Rohlmann, A ;
Strauss, J ;
Duda, GN .
JOURNAL OF BIOMECHANICS, 2001, 34 (07) :859-871
[8]   A biomechanical evaluation of the tibiofibular and tibiotalar ligaments of the ankle [J].
Beumer, A ;
van Hemert, WLW ;
Swierstra, BA ;
Jasper, LE ;
Belkoff, SM .
FOOT & ANKLE INTERNATIONAL, 2003, 24 (05) :426-429
[9]   Finite element analysis of a novel pin-sleeve system for external fixation of distal limb fractures in horses [J].
Brianza, Stefano ;
Brighenti, Vittoria ;
Lansdowne, Jennifer L. ;
Schwieger, Karsten ;
Boure, Ludovic .
VETERINARY JOURNAL, 2011, 190 (02) :260-267
[10]  
Carroll EA, 2010, J SURG ORTHOP ADV, V20, P74