Application of antibody nanogold ionic liquid carbon paste electrode for sensitive electrochemical immunoassay of thyroid-stimulating hormone

被引:113
作者
Beitollahi, Hadi [1 ]
Ivari, Susan Ghofrani [2 ]
Torkzadeh-Mahani, Masoud [3 ]
机构
[1] Grad Univ Adv Technol, Environm Dept, Inst Sci & High Technol & Environm Sci, Kerman, Iran
[2] Grad Univ Adv Technol, Dept Chem, Kerman, Iran
[3] Grad Univ Adv Technol, Biotechnol Dept, Inst Sci & High Technol & Environm Sci, Kerman, Iran
关键词
Electrochemical immunoassay; Thyroid stimulating hormone; Carbon paste electrode; Gold nanoparticle; Ionic liquid; O-Aminophenol; CHEMILUMINESCENCE IMMUNOASSAY; AMPEROMETRIC IMMUNOSENSOR; GOLD NANOPARTICLES; GRAPHENE; ENZYME; NANOCOMPOSITE; PROSPECTS; SENSOR; POINT; CARE;
D O I
10.1016/j.bios.2018.03.003
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
A novel electrochemical immunosensor based on carbon paste electrode (CPE) composed of ionic liquid (IL) and graphite was constructed. It demonstrated good efficiency for quick (each test in 30 s) determination of thyroid stimulating hormone (TSH). Electrode surface was modified by gold nanoparticles in order to immobilize of the thyroid stimulating hormone antibody (anti-TSH) on the CPE. The immunoassay structure was established by sandwiching the antigen (TSH) between the thyroid stimulating hormone antibody on the CPE surface modified with gold nanoparticles and the secondary antibody, polyclonal anti-human-TSH labeled with horseradish peroxidase (HRP-labeled anti-TSH). The signal of differential pulse voltammetry (DPV) was used as a basis for the determination of TSH concentration. This signal is generated by the reaction between O-aminophenol (OAP) and H2O2 catalyzed by HRP. The proposed immunosensor is able to measure the concentration of TSH in a linear range between 0.2 and 90.0 ng/mL with a detection limit 0.1 +/- 0.02 ng/mL. In addition, high sensitivity and acceptable stability were achieved by this immunosensor which is promising in the clinical assay of TSH.
引用
收藏
页码:97 / 102
页数:6
相关论文
共 39 条
  • [11] Electrochemical immunosensors for the detection of survival motor neuron (SMN) protein using different carbon nanomaterials-modified electrodes
    Eissa, Shimaa
    Alshehri, Nawal
    Rahman, Anas M. Abdel
    Dasouki, Majed
    Abu Salah, Khalid M.
    Zourob, Mohammed
    [J]. BIOSENSORS & BIOELECTRONICS, 2018, 101 : 282 - 289
  • [12] Disposable amperometric immunosensor for the detection of polycyclic aromatic hydrocarbons (PAHs) using screen-printed electrodes
    Fähnrich, KA
    Pravda, M
    Guilbault, GG
    [J]. BIOSENSORS & BIOELECTRONICS, 2003, 18 (01) : 73 - 82
  • [13] Garcia-Gonzalez Elena, 2016, Pract Lab Med, V4, P1, DOI 10.1016/j.plabm.2015.11.001
  • [14] Direct electrochemistry and electrocatalysis of hemoglobin on carbon ionic liquid electrode
    Huang, Ke-Jing
    Sun, Jun-Yong
    Niu, De-Jun
    Xie, Wan-Zhen
    Wang, Wei
    [J]. COLLOIDS AND SURFACES B-BIOINTERFACES, 2010, 78 (01) : 69 - 74
  • [15] AUTOMATIC-APPARATUS FOR HETEROGENEOUS ENZYME IMMUNOASSAYS BASED ON ELECTROCATALYTIC DETECTION OF THE ENZYME AND ELECTROCHEMICAL REGENERATION OF THE SOLID-PHASE
    HUET, D
    BOURDILLON, C
    [J]. ANALYTICA CHIMICA ACTA, 1993, 272 (02) : 205 - 212
  • [16] An innovative sample-to-answer polymer lab-on-a-chip with on-chip reservoirs for the POCT of thyroid stimulating hormone (TSH)
    Jung, Wooseok
    Han, Jungyoup
    Kai, Junhai
    Lim, Ji-Youn
    Sul, Donggeun
    Ahn, Chong H.
    [J]. LAB ON A CHIP, 2013, 13 (23) : 4653 - 4662
  • [17] Disposable multichannel immunosensors for 2,4-dichlorophenoxyacetic acid using acetylcholinesterase as an enzyme label
    Kalab, T
    Skladal, P
    [J]. ELECTROANALYSIS, 1997, 9 (04) : 293 - 297
  • [18] Kaur M. V. Amandeep, 2014, INT J BIOASSAYS, V3, P3040
  • [19] Gold-film array-electrode for electrochemical ELISA
    Kim, SD
    Chung, JW
    Kim, JT
    Krause, H
    Pyun, JC
    [J]. SENSORS AND ACTUATORS B-CHEMICAL, 2005, 111 : 463 - 469
  • [20] Renewable amperometric immunosensor for Schistosoma japonium antibody assay
    Liu, GD
    Wu, ZY
    Wang, SP
    Shen, GL
    Yu, RQ
    [J]. ANALYTICAL CHEMISTRY, 2001, 73 (14) : 3219 - 3226