Monitoring of heart rate and inter-beat intervals with wrist plethysmography in patients with atrial fibrillation

被引:46
作者
Harju, Jarkko [1 ]
Tarniceriu, Adrian [2 ]
Parak, Jakub [3 ,4 ]
Vehkaoja, Antti [3 ,4 ]
Yli-Hankala, Arvi [1 ,5 ]
Korhonen, Ilkka [3 ,4 ]
机构
[1] Tampere Univ Hosp, Dept Anesthesia, Tampere, Finland
[2] PulseOn SA, Neuchatel, Switzerland
[3] Tampere Univ Technol, Tampere, Finland
[4] Pulseon Oy, Espoo, Finland
[5] Univ Tampere, Med Sch, Tampere, Finland
关键词
heart rate; pulse rate; photoplethysmography; atrial fibrillation; perioperative monitoring; RATE-VARIABILITY; ACCURACY; PREVALENCE; FAILURE; WORN;
D O I
10.1088/1361-6579/aac9a9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective: Atrial fibrillation (AF) causes marked risk for patients, while silent fibrillation may remain unnoticed if not suspected and screened. Development of comfortable yet accurate beat-to-beat heart rate (HR) monitoring with good AF detection sensitivity would facilitate screening and improve treatment. The purpose of this study was to evaluate whether a wrist-worn photoplethysmography (PPG) device can be used to monitor beat-to-beat HR accurately during post-operative treatment in patients suffering from AF and whether wrist-PPG can be used to distinguish AF from sinus rhythm (SR). Approach: Twenty-nine patients (14 with AF, 15 with SR, mean age 71.5 years) with multiple comorbidities were monitored during routine post-operative treatment. The monitoring included standard ECG, finger PPG monitoring and a wrist-worn PPG monitor with green and infrared light sources. The HR from PPG sensors was compared against ECG-derived HR. Main results: The wrist PPG technology had very good HR and beat detection accuracy when using green light. For the SR group, the mean absolute error (MAE) for HR was 1.50 bpm, and for the interbeat intervals (IBI), the MAE was 7.64 ms. For the AF group, the MAE for HR was 4.28 bpm and for IBI, the MAE was 14.67 ms. Accuracy for the infrared (IR) channel was worse. Finger PPG provided similar accuracy for HR and better accuracy for the IBI. AF detection sensitivity using green light was 99.0% and the specificity was 93.0%. Performance can be improved by discarding unreliable IBI periods. Significance: Results suggest that wrist PPG measurement allows accurate HR and beat-to-beat HR monitoring also in AF patients, and could be used for differentiating between SR and AF with very good sensitivity.
引用
收藏
页数:13
相关论文
共 41 条
  • [31] Clinical relevance of silent atrial fibrillation: Prevalence, prognosis, quality of life, and management
    Savelieva, I
    Camm, AJ
    [J]. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 2000, 4 (02) : 369 - 382
  • [32] Accuracy in Wrist-Worn, Sensor-Based Measurements of Heart Rate and Energy Expenditure in a Diverse Cohort
    Shcherbina, Anna
    Mattsson, C. Mikael
    Waggott, Daryl
    Salisbury, Heidi
    Christle, Jeffrey W.
    Hastie, Trevor
    Wheeler, Matthew T.
    Ashley, Euan A.
    [J]. JOURNAL OF PERSONALIZED MEDICINE, 2017, 7 (02):
  • [33] Photoplethysmography: Beyond the Calculation of Arterial Oxygen Saturation and Heart Rate
    Shelley, Kirk H.
    [J]. ANESTHESIA AND ANALGESIA, 2007, 105 : S31 - S36
  • [34] How should we optimize cardiac resynchronization therapy?
    Stanton, Tony
    Hawkins, Nathaniel M.
    Hogg, Kerry J.
    Goodfield, Nicholas E. R.
    Petrie, Mark C.
    McMurray, John J. V.
    [J]. EUROPEAN HEART JOURNAL, 2008, 29 (20) : 2458 - 2472
  • [35] Tanev S., 2012, INT J BIOAUTOMATION, V16, P273
  • [36] Terbizan D., 2002, Measurement in Physical Education and Exercise Science, V6, P243, DOI DOI 10.1207/S15327841MPEE0604_3
  • [37] Incidence of Atrial Fibrillation and Relationship With Cardiovascular Events, Heart Failure, and Mortality A Community-Based Study From the Netherlands
    Vermond, Rob A.
    Geelhoed, Bastiaan
    Verweij, Niek
    Tieleman, Robert G.
    Van der Harst, Pim
    Hillege, Hans L.
    Van Gilst, Wiek H.
    Van Gelder, Isabelle C.
    Rienstra, Michiel
    [J]. JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2015, 66 (09) : 1000 - 1007
  • [38] Accuracy of Wrist-Worn Heart Rate Monitors
    Wang, Robert
    Blackburn, Gordon
    Desai, Milind
    Phelan, Dermot
    Gillinov, Lauren
    Houghtaling, Penny
    Gillinov, Marc
    [J]. JAMA CARDIOLOGY, 2017, 2 (01) : 104 - 106
  • [39] Comparison of three mobile devices for measuring R-R intervals and heart rate variability: Polar S810i, Suunto t6 and an ambulatory ECG system
    Weippert, Matthias
    Kumar, Mohit
    Kreuzfeld, Steffi
    Arndt, Dagmar
    Rieger, Annika
    Stoll, Regina
    [J]. EUROPEAN JOURNAL OF APPLIED PHYSIOLOGY, 2010, 109 (04) : 779 - 786
  • [40] Highly wearable cuff-less blood pressure and heart rate monitoring with single-arm electrocardiogram and photoplethysmogram signals
    Zhang, Qingxue
    Zhou, Dian
    Zeng, Xuan
    [J]. BIOMEDICAL ENGINEERING ONLINE, 2017, 16