Monitoring of heart rate and inter-beat intervals with wrist plethysmography in patients with atrial fibrillation

被引:46
作者
Harju, Jarkko [1 ]
Tarniceriu, Adrian [2 ]
Parak, Jakub [3 ,4 ]
Vehkaoja, Antti [3 ,4 ]
Yli-Hankala, Arvi [1 ,5 ]
Korhonen, Ilkka [3 ,4 ]
机构
[1] Tampere Univ Hosp, Dept Anesthesia, Tampere, Finland
[2] PulseOn SA, Neuchatel, Switzerland
[3] Tampere Univ Technol, Tampere, Finland
[4] Pulseon Oy, Espoo, Finland
[5] Univ Tampere, Med Sch, Tampere, Finland
关键词
heart rate; pulse rate; photoplethysmography; atrial fibrillation; perioperative monitoring; RATE-VARIABILITY; ACCURACY; PREVALENCE; FAILURE; WORN;
D O I
10.1088/1361-6579/aac9a9
中图分类号
Q6 [生物物理学];
学科分类号
071011 ;
摘要
Objective: Atrial fibrillation (AF) causes marked risk for patients, while silent fibrillation may remain unnoticed if not suspected and screened. Development of comfortable yet accurate beat-to-beat heart rate (HR) monitoring with good AF detection sensitivity would facilitate screening and improve treatment. The purpose of this study was to evaluate whether a wrist-worn photoplethysmography (PPG) device can be used to monitor beat-to-beat HR accurately during post-operative treatment in patients suffering from AF and whether wrist-PPG can be used to distinguish AF from sinus rhythm (SR). Approach: Twenty-nine patients (14 with AF, 15 with SR, mean age 71.5 years) with multiple comorbidities were monitored during routine post-operative treatment. The monitoring included standard ECG, finger PPG monitoring and a wrist-worn PPG monitor with green and infrared light sources. The HR from PPG sensors was compared against ECG-derived HR. Main results: The wrist PPG technology had very good HR and beat detection accuracy when using green light. For the SR group, the mean absolute error (MAE) for HR was 1.50 bpm, and for the interbeat intervals (IBI), the MAE was 7.64 ms. For the AF group, the MAE for HR was 4.28 bpm and for IBI, the MAE was 14.67 ms. Accuracy for the infrared (IR) channel was worse. Finger PPG provided similar accuracy for HR and better accuracy for the IBI. AF detection sensitivity using green light was 99.0% and the specificity was 93.0%. Performance can be improved by discarding unreliable IBI periods. Significance: Results suggest that wrist PPG measurement allows accurate HR and beat-to-beat HR monitoring also in AF patients, and could be used for differentiating between SR and AF with very good sensitivity.
引用
收藏
页数:13
相关论文
共 41 条
  • [1] Agreement between methods of measurement with multiple observations per individual
    Bland, J. Martin
    Altman, Douglas G.
    [J]. JOURNAL OF BIOPHARMACEUTICAL STATISTICS, 2007, 17 (04) : 571 - 582
  • [2] Photoplethysmography variability as an alternative approach to obtain heart rate variability information in chronic pain patient
    Chuang, Chiung-Cheng
    Ye, Jing-Jhao
    Lin, Wan-Chun
    Lee, Kuan-Ting
    Tai, Yu-Ting
    [J]. JOURNAL OF CLINICAL MONITORING AND COMPUTING, 2015, 29 (06) : 801 - 806
  • [3] Detection of atrial fibrillation using an earlobe photoplethysmographic sensor
    Conroy, Thomas
    Guzman, Jairo Hernandez
    Hall, Burr
    Tsouri, Gill
    Couderc, Jean-Philippe
    [J]. PHYSIOLOGICAL MEASUREMENT, 2017, 38 (10) : 1906 - 1918
  • [4] Accuracy of smartphone apps for heart rate measurement
    Coppetti, Thomas
    Brauchlin, Andreas
    Mueggler, Simon
    Attinger-Toller, Adrian
    Templin, Christian
    Schoenrath, Felix
    Hellermann, Jens
    Luescher, Thomas F.
    Biaggi, Patric
    Wyss, Christophe A.
    [J]. EUROPEAN JOURNAL OF PREVENTIVE CARDIOLOGY, 2017, 24 (12) : 1287 - 1293
  • [5] FPGA-oriented HW/SW implementation of ECG beat detection and classification algorithm
    Cvikl, Matej
    Zemva, Andrej
    [J]. DIGITAL SIGNAL PROCESSING, 2010, 20 (01) : 238 - 248
  • [6] Automatic Real Time Detection of Atrial Fibrillation
    Dash, S.
    Chon, K. H.
    Lu, S.
    Raeder, E. A.
    [J]. ANNALS OF BIOMEDICAL ENGINEERING, 2009, 37 (09) : 1701 - 1709
  • [7] Delgado-Gonzalo R, 2015, IEEE ENG MED BIO, P430, DOI 10.1109/EMBC.2015.7318391
  • [8] Technical challenges and future directions in lead reconstruction for reduced-lead systems
    Feild, Dirk Q.
    Zhou, Sophia H.
    Helfenbein, Eric D.
    Gregg, Richard E.
    Lindauer, James M.
    [J]. JOURNAL OF ELECTROCARDIOLOGY, 2008, 41 (06) : 466 - 473
  • [9] Fridman V, 2016, J ATR FIBRILLATION, V8, P29
  • [10] Fukushima H, 2012, IEEE ENG MED BIO, P2901, DOI 10.1109/EMBC.2012.6346570