Photon Counting CT Angiography of the Head and Neck: Image Quality Assessment of Polyenergetic and Virtual Monoenergetic Reconstructions

被引:22
作者
Michael, Arwed Elias [1 ]
Boriesosdick, Jan [1 ]
Schoenbeck, Denise [1 ]
Lopez-Schmidt, Ingo [1 ]
Kroeger, Jan Robert [1 ]
Moenninghoff, Christoph [1 ]
Horstmeier, Sebastian [1 ]
Pennig, Lenhard [2 ,3 ]
Borggrefe, Jan [1 ]
Niehoff, Julius Henning [1 ]
机构
[1] Ruhr Univ Bochum, Johannes Wesling Univ Hosp, Dept Radiol Neuroradiol & Nucl Med, D-32429 Minden, Germany
[2] Univ Cologne, Inst Diagnost & Intervent Radiol, Fac Med, D-50923 Cologne, Germany
[3] Univ Cologne, Univ Hosp Cologne, D-50923 Cologne, Germany
关键词
computed tomography; photon counting detector; CT angiography of the head and neck; virtual monoenergetic reconstructions; KILOELECTRON-VOLT-SETTINGS; LAYER DETECTOR CT; OPTIMIZATION; PERFORMANCE; DIAGNOSIS;
D O I
10.3390/diagnostics12061306
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background: The purpose of the present study was the evaluation of the image quality of polyenergetic and monoenergetic reconstructions (PERs and MERs) of CT angiographies (CTAs) of the head and neck acquired with the novel photon counting CT (PCCT) method in clinical routine. Methods: Thirty-seven patients were enrolled in this retrospective study. Quantitative image parameters of the extracranial, intracranial and cerebral arteries were evaluated for the PER and MER (40-120 keV). Additionally, two radiologists rated the perceived image quality. Results: The mean CTDIvol used in the PCCT was 8.31 +/- 1.19 mGy. The highest signal within the vessels was detected in the 40 keV MER, whereas the lowest noise was detected in the 115 keV MER. The most favorable contrast-to-noise-ratio (CNR) and signal-to-noise-ratio (SNR) were detected in the PER and low keV MER. In the qualitative image analysis, the PER was superior to the MER in all rated criteria. For MER, 60-65 keV was rated as best image quality. Conclusion: Overall, PCCT offers excellent image quality for CTAs of the head and neck. At the current state, the PER of the PCCT seems to be the most favorable reconstruction for diagnostic reporting.
引用
收藏
页数:15
相关论文
共 37 条
[1]   Advanced image-based virtual monoenergetic dual-energy CT angiography of the abdomen: optimization of kiloelectron volt settings to improve image contrast [J].
Albrecht, Moritz H. ;
Scholtz, Jan-Erik ;
Huesers, Kristina ;
Beeres, Martin ;
Bucher, Andreas M. ;
Kaup, Moritz ;
Martin, Simon S. ;
Fischer, Sebastian ;
Bodelle, Boris ;
Bauer, Ralf W. ;
Lehnert, Thomas ;
Vogl, Thomas J. ;
Wichmann, Julian L. .
EUROPEAN RADIOLOGY, 2016, 26 (06) :1863-1870
[2]   ENERGY-SELECTIVE RECONSTRUCTIONS IN X-RAY COMPUTERIZED TOMOGRAPHY [J].
ALVAREZ, RE ;
MACOVSKI, A .
PHYSICS IN MEDICINE AND BIOLOGY, 1976, 21 (05) :733-744
[3]   Value of monoenergetic low-kV dual energy CT datasets for improved image quality of CT pulmonary angiography [J].
Apfaltrer, Paul ;
Sudarski, Sonja ;
Schneider, David ;
Nance, John W., Jr. ;
Haubenreisser, Holger ;
Fink, Christian ;
Schoenberg, Stefan O. ;
Henzler, Thomas .
EUROPEAN JOURNAL OF RADIOLOGY, 2014, 83 (02) :322-328
[4]   Advances in endovascular treatment of acute ischaemic stroke [J].
Asadi, H. ;
Dowling, R. ;
Yan, B. ;
Wong, S. ;
Mitchell, P. .
INTERNAL MEDICINE JOURNAL, 2015, 45 (08) :798-805
[5]   Segments of the internal carotid artery: A new classification [J].
Bouthillier, A ;
vanLoveren, HR ;
Keller, JT .
NEUROSURGERY, 1996, 38 (03) :425-432
[6]   Dual energy computed tomography virtual monoenergetic imaging: technique and clinical applications [J].
D'Angelo, Tommaso ;
Cicero, Giuseppe ;
Mazziotti, Silvio ;
Ascent, Giorgio ;
Albrecht, Moritz H. ;
Martin, Simon S. ;
Othman, Ahmed E. ;
Vogl, Thomas J. ;
Wichmann, Julian L. .
BRITISH JOURNAL OF RADIOLOGY, 2019, 92 (1098)
[7]   Photon-counting x-ray detectors for CT [J].
Danielsson, Mats ;
Persson, Mats ;
Sjolin, Martin .
PHYSICS IN MEDICINE AND BIOLOGY, 2021, 66 (03)
[8]   High-Pitch Photon-Counting Detector Computed Tomography Angiography of the Aorta Intraindividual Comparison to Energy-Integrating Detector Computed Tomography at Equal Radiation Dose [J].
Euler, Andre ;
Higashigaito, Kai ;
Mergen, Victor ;
Sartoretti, Thomas ;
Zanini, Bettina ;
Schmidt, Bernhard ;
Flohr, Thomas G. ;
Ulzheimer, Stefan ;
Eberhard, Matthias ;
Alkadhi, Hatem .
INVESTIGATIVE RADIOLOGY, 2022, 57 (02) :115-121
[9]   First performance evaluation of a dual-source CT (DSCT) system [J].
Flohr, TG ;
McCollough, CH ;
Bruder, H ;
Petersilka, M ;
Gruber, K ;
Süss, C ;
Grasruck, M ;
Stierstorfer, K ;
Krauss, B ;
Raupach, R ;
Primak, AN ;
Küttner, A ;
Achenbach, S ;
Becker, C ;
Kopp, A ;
Ohnesorge, BM .
EUROPEAN RADIOLOGY, 2006, 16 (02) :256-268
[10]   Photon-counting CT review [J].
Flohr, Thomas ;
Petersilka, Martin ;
Henning, Andre ;
Ulzheimer, Stefan ;
Ferda, Jiri ;
Schmidt, Bernhard .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 79 :126-136