WHO Multicenter Evaluation of FACSCount CD4 and Pima CD4 T-Cell Count Systems: Instrument Performance and Misclassification of HIV-Infected Patients

被引:0
作者
Wade, Djibril [1 ,2 ,3 ]
Daneau, Geraldine [1 ]
Aboud, Said [4 ]
Vercauteren, Gaby H. [5 ]
Urassa, Willy S. K. [5 ]
Kestens, Luc [1 ,3 ]
机构
[1] Inst Trop Med Antwerp, Dept Biomed Sci, Immunol Lab, Antwerp, Belgium
[2] Univ Cheikh Anta Diop, Le Dantec Univ Teaching Hosp, Lab Bacteriol Virol, Immunol Unit, Dakar 92000, Senegal
[3] Univ Antwerp, Dept Biomed Sci, B-2020 Antwerp, Belgium
[4] Muhimbili Univ Hlth & Allied Sci, Dept Microbiol & Immunol, Dar Es Salaam, Tanzania
[5] WHO, CH-1211 Geneva, Switzerland
关键词
CD4; count; CD4%; resource-limited settings; Pima CD4; FACSCount CD4; POINT-OF-CARE; FLOW-CYTOMETRY; VIRAL LOAD; LYMPHOCYTE COUNTS; ENUMERATION; ANALYZER; SETTINGS; BLOOD; INDIVIDUALS; VALIDATION;
D O I
暂无
中图分类号
R392 [医学免疫学]; Q939.91 [免疫学];
学科分类号
100102 ;
摘要
Background: CD4(+) T-cell counts are used to screen and follow-up HIV-infected patients during treatment. As part of the World Health Organization prequalification program of diagnostics, we conducted an independent multicenter evaluation of the FACSCount CD4 and the Pima CD4, using the FACSCalibur as reference method. Methods: A total of 440 paired capillary and venous blood samples were collected from HIV-infected patients attending the HIV outpatient clinic in Antwerp, Belgium, and the HIV care and treatment center in Dar es Salam, Tanzania. Capillary blood was run on Pima analyzer, whereas venous blood was analyzed on FACSCount, Pima, and FACSCalibur instruments. Precision and agreement between methods were assessed. Results: The FACSCount CD4 results were in agreement with the FACSCalibur results with relative bias of 0.4% and 3.1% on absolute CD4 counts and an absolute bias of -0.6% and -1.1% on CD4% in Antwerp and Dar es Salam, respectively. The Pima CD4 results were in agreement with the FACSCalibur results with relative bias of -4.1% and -9.4% using venous blood and of -9.5% and -0.9% using capillary blood in Antwerp and Dar es Salam, respectively. At the threshold of 350 cells per microliter, the FACSCount CD4 and Pima CD4 using venous and capillary blood misclassified 7%, 9%, and 13% of patients, respectively. Conclusions: The FACSCount CD4 provides reliable CD4 counts and CD4% and is suitable for monitoring adult and pediatric HIV patients in moderate-volume settings. The Pima CD4 is more suitable for screening eligible adult HIV patients for antiretroviral treatment initiation in low-volume laboratories.
引用
收藏
页码:E98 / E107
页数:10
相关论文
共 51 条
[1]  
[Anonymous], HIV AIDS DIAGN TECHN
[2]   A reliable and inexpensive EasyCD4 assay for monitoring HIV-infected individuals in resource-limited settings [J].
Balakrishnan, Pachamuthu ;
Solomon, Suniti ;
Mohanakrishnan, Janardhanan ;
Cecelia, Anitha J. ;
Kumarasamy, Nagalingeswaran ;
Murugavel, Kailapuri G. ;
Venkatakrishnan, Bhavani ;
Solomon, Sunil S. ;
Crowe, Suzanne M. ;
Ganesh, Aylur K. ;
Thyagarajan, Sadras P. ;
Flanigan, Timothy ;
Mayer, Kenneth H. .
JAIDS-JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2006, 43 (01) :23-26
[3]  
Bekele Yonas, 2011, Ethiop Med J, V49, P187
[4]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[5]   REVISED SCHEME FOR EVALUATION OF AUTOMATIC INSTRUMENTS FOR USE IN CLINICAL CHEMISTRY [J].
BROUGHTON, PM ;
GOWENLOCK, AH ;
MCCORMACK, JJ ;
NEILL, DW .
ANNALS OF CLINICAL BIOCHEMISTRY, 1974, 11 (NOV) :207-218
[6]  
Cassens U, 2004, ANTIVIR THER, V9, P395
[7]  
Cheng XH, 2007, JAIDS-J ACQ IMM DEF, V45, P257
[8]  
Chesher Douglas, 2008, Clin Biochem Rev, V29 Suppl 1, pS23
[9]   A North American multilaboratory study of CD4 counts using flow cytometric PanLeukogating (PLG): A NIAID-DAIDS immunology quality assessment program study [J].
Denny, Thomas N. ;
Gelman, Rebecca ;
Bergeron, Michele ;
Landay, Alan ;
Lam, Lee ;
Louzao, Raul ;
Mandy, Frank F. ;
Schmitz, John ;
Spira, Thomas ;
Wilkening, Cindy ;
Glencross, Deborah K. .
CYTOMETRY PART B-CLINICAL CYTOMETRY, 2008, 74B :S52-S64
[10]   Multisite Evaluation of a Point-of-Care Instrument for CD4+ T-Cell Enumeration Using Venous and Finger-Prick Blood: The PIMA CD4 [J].
Diaw, Papa Alassane ;
Daneau, Geraldine ;
Coly, Abdoul Aziz ;
Ndiaye, Birahim Pierre ;
Wade, Djibril ;
Camara, Makhtar ;
Mboup, Souleymane ;
Kestens, Luc ;
Dieye, Tandakha Ndiaye .
JAIDS-JOURNAL OF ACQUIRED IMMUNE DEFICIENCY SYNDROMES, 2011, 58 (04) :E103-E111