The dosimetric effect of residual breath-hold motion in pencil beam scanned proton therapy - An experimental study

被引:7
作者
Gorgisyan, Jenny [1 ,2 ,3 ,7 ]
Lomax, Antony J. [1 ,4 ]
af Rosenschold, Per Munck [3 ,5 ]
Persson, Gitte F. [2 ,8 ,9 ]
Krieger, Miriam [1 ,4 ]
Colvill, Emma [1 ]
Scherman, Jonas [5 ]
Gagnon-Moisan, Francis [1 ]
Egloff, Martina [1 ]
Fattori, Giovanni [1 ]
Engelholm, Svend Aage [2 ]
Weber, Damien C. [1 ,6 ]
Perrin, Rosalind [1 ,10 ]
机构
[1] ETH Domain, Paul Scherrer Inst, Zurich, Switzerland
[2] Copenhagen Univ Hosp, Rigshosp, Dept Oncol, Copenhagen, Denmark
[3] Univ Copenhagen, Niels Bohr Inst, Herlev, Denmark
[4] Swiss Fed Inst Technol, Dept Phys, Zurich, Switzerland
[5] Skane Univ Hosp, Dept Hematol Oncol & Radiat Phys, Radiat Phys, Lund, Sweden
[6] Univ Hosp Zurich, Radiat Oncol Dept, Zurich, Switzerland
[7] Lausanne Univ Hosp, Inst Radiat Phys, Lausanne, Switzerland
[8] Univ Copenhagen, Herlev Hosp, Dept Oncol, Herlev, Denmark
[9] Univ Copenhagen, Dept Clin Med, Fac Med Sci, Herlev, Denmark
[10] Univ Klinikum Erlangen, Dept Med Radiat Phys, Strahlenklin, Univ Str 29, D-91054 Erlangen, Germany
关键词
4D phantom; PBS proton therapy; Breath-hold; Lung tumours; LUNG-TUMORS; RESPIRATORY MOTION; RADIOTHERAPY; UNCERTAINTIES; MITIGATION; PHANTOM; CANCER; ROBUSTNESS; TARGETS; THORAX;
D O I
10.1016/j.radonc.2019.01.033
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background and purpose: Motion management in the treatment of lung cancer is necessary to assure highest quality of the delivered radiation therapy. In this study, the breath-hold technique is experimentally investigated for pencil beam scanned (PBS) proton therapy, with respect to the dosimetric effect of residual breath-hold motion. Material and methods: Three-dimensional (3D)-printed tumours extracted from CT scans of three patients were inserted into a dynamic anthropomorphic breathing phantom. The target was set up to move with the individual patient's tumour motion during breath-hold as previously assessed on fluoroscopy. Target dose was measured with radio-chromic film, and both single field uniform dose (SFUD) and intensity-modulated proton therapy (IMPT) plans were delivered. Experiments were repeated for each patient without any motion, to compute the relative dose deviation between static and breath-hold cases. Results: SFUD plans showed small dose deviations between static and breath-hold cases, as evidenced by the gamma pass rate (3%, 3 mm) of 85% or higher. Dose deviation was more evident for IMPT plans, with gamma pass rate reduced to 50-70%. Conclusions: The breath-hold technique is robust to residual intra-breath-hold motion for SFUD treatment plans, based on our experimental study. IMPT was less robust with larger detected dose deviations. (C) 2019 Elsevier B.V. All rights reserved.
引用
收藏
页码:135 / 142
页数:8
相关论文
共 37 条
[11]   Geometric uncertainties in voluntary deep inspiration breath hold radiotherapy for locally advanced lung cancer [J].
Josipovic, Mirjana ;
Persson, Gitte F. ;
Dueck, Jenny ;
Bangsgaard, Jens Peter ;
Westman, Gunnar ;
Specht, Lena ;
Aznar, Marianne C. .
RADIOTHERAPY AND ONCOLOGY, 2016, 118 (03) :510-514
[12]   Carbon ion radiotherapy in Japan: an assessment of 20 years of clinical experience [J].
Kamada, Tadashi ;
Tsujii, Hirohiko ;
Blakely, Eleanor A. ;
Debus, Juergen ;
De Neve, Wilfried ;
Durante, Marco ;
Jaekel, Oliver ;
Mayer, Ramona ;
Orecchia, Roberto ;
Poetter, Richard ;
Vatnitsky, Stanislav ;
Chu, William T. .
LANCET ONCOLOGY, 2015, 16 (02) :E93-E100
[13]   Impact of Real-Time Image Gating on Spot Scanning Proton Therapy for Lung Tumors: A Simulation Study [J].
Kanehira, Takahiro ;
Matsuura, Taeko ;
Takao, Seishin ;
Matsuzaki, Yuka ;
Fujii, Yusuke ;
Fujii, Takaaki ;
Ito, Yoichi M. ;
Miyamoto, Naoki ;
Inoue, Tetsuya ;
Katoh, Norio ;
Shimizu, Shinichi ;
Umegaki, Kikuo ;
Shirato, Hiroki .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2017, 97 (01) :173-181
[14]   The management of respiratory motion in radiation oncology report of AAPM Task Group 76 [J].
Keall, Paul J. ;
Mageras, Gig S. ;
Balter, James M. ;
Emery, Richard S. ;
Forster, Kenneth M. ;
Jiang, Steve B. ;
Kapatoes, Jeffrey M. ;
Low, Daniel A. ;
Murphy, Martin J. ;
Murray, Brad R. ;
Ramsey, Chester R. ;
Van Herk, Marcel B. ;
Vedam, S. Sastry ;
Wong, John W. ;
Yorke, Ellen .
MEDICAL PHYSICS, 2006, 33 (10) :3874-3900
[15]   Required transition from research to clinical application: Report on the 4D treatment planning workshops 2014 and 2015 [J].
Knopf, Antje-Christin ;
Stuetzer, Kristin ;
Richter, Christian ;
Rucinski, Antoni ;
da Silva, Joakim ;
Phillips, Justin ;
Engelsman, Martijn ;
Shimizu, Shinichi ;
Werner, Rene ;
Jakobi, Annika ;
Goksel, Orcun ;
Zhang, Ye ;
Oshea, Tuathan ;
Fast, Martin ;
Perrin, Rosalind ;
Bert, Christoph ;
Rinaldi, Ilaria ;
Korevaar, Erik ;
McClelland, Jamie .
PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2016, 32 (07) :874-882
[16]   Adequate margin definition for scanned particle therapy in the incidence of intrafractional motion [J].
Knopf, Antje-Christin ;
Boye, Dirk ;
Lomax, Antony ;
Mori, Shininchiro .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (17) :6079-6094
[17]   Scanned proton radiotherapy for mobile targets-the effectiveness of re-scanning in the context of different treatment planning approaches and for different motion characteristics [J].
Knopf, Antje-Christin ;
Hong, Theodore S. ;
Lomax, Antony .
PHYSICS IN MEDICINE AND BIOLOGY, 2011, 56 (22) :7257-7271
[18]   Simulation of dosimetric consequences of 4D-CT-based motion margin estimation for proton radiotherapy using patient tumor motion data [J].
Koybasi, Ozhan ;
Mishra, Pankaj ;
James, Sara St. ;
Lewis, John H. ;
Seco, Joao .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (04) :853-867
[19]   Treatment planning and verification of proton therapy using spot scanning:: Initial experiences [J].
Lomax, AJ ;
Böhringer, T ;
Bolsi, A ;
Coray, D ;
Emert, F ;
Goitein, G ;
Jermann, M ;
Lin, SX ;
Pedroni, E ;
Rutz, H ;
Stadelmann, O ;
Timmermann, B ;
Verwey, J ;
Weber, DC .
MEDICAL PHYSICS, 2004, 31 (11) :3150-3157
[20]   High precision bladder cancer irradiation by integrating a library planning procedure of 6 prospectively generated SIB IMRT plans with image guidance using lipiodol markers [J].
Meijer, Gert Johan ;
van der Toorn, Peter-Paul ;
Bal, Matthieu ;
Schuring, Danny ;
Weterings, Jan ;
de Wildt, Michel .
RADIOTHERAPY AND ONCOLOGY, 2012, 105 (02) :174-179