Characterization of acoustic noise in a neonatal intensive care unit MRI system

被引:29
|
作者
Tkach, Jean A. [1 ]
Li, Yu [1 ]
Pratt, Ronald G. [1 ]
Baroch, Kelly A. [2 ]
Loew, Wolfgang [1 ]
Daniels, Barret R. [1 ]
Giaquinto, Randy O. [1 ]
Merhar, Stephanie L. [3 ]
Kline-Fath, Beth M. [4 ]
Dumoulin, Charles L. [1 ]
机构
[1] Cincinnati Childrens Hosp Med Ctr, Dept Radiol, Imaging Res Ctr, Cincinnati, OH 45229 USA
[2] Cincinnati Childrens Hosp Med Ctr, Div Audiol, Cincinnati, OH 45229 USA
[3] Cincinnati Childrens Hosp Med Ctr, Perinatal Inst, Div Neonatol & Pulm Biol, Cincinnati, OH 45229 USA
[4] Cincinnati Childrens Hosp Med Ctr, Dept Radiol, Cincinnati, OH 45229 USA
关键词
Magnetic resonance imaging; Neonatal intensive care unit; Acoustic noise; Safety; Neonates; CHLORAL HYDRATE; SEDATION; TRANSPORT; INFANTS; BRAIN; FETUS;
D O I
10.1007/s00247-014-2909-0
中图分类号
R72 [儿科学];
学科分类号
100202 ;
摘要
To eliminate the medical risks and logistical challenges of transporting infants from the neonatal intensive care unit (NICU) to the radiology department for magnetic resonance imaging, a small-footprint 1.5-T MRI scanner has been developed for neonatal imaging within the NICU. MRI is known to be noisy, and exposure to excessive acoustic noise has the potential to elicit physiological distress and impact development in the term and preterm infant. To measure and compare the acoustic noise properties of the NICU MRI system against those of a conventional 1.5-T MRI system. We performed sound pressure level measurements in the NICU MRI scanner and in a conventional adult-size whole-body 1.5-T MRI system. Sound pressure level measurements were made for six standard clinical MR imaging protocols. The average sound pressure level value, reported in unweighted (dB) and A-weighted (dBA) decibels for all six imaging pulse sequences, was 73.8 dB and 88 dBA for the NICU scanner, and 87 dB and 98.4 dBA for the conventional MRI scanner. The sound pressure level values measured on the NICU scanner for each of the six MR imaging pulse sequences were consistently and significantly (P = 0.03) lower, with an average difference of 14.2 dB (range 10-21 dB) and 11 dBA (range 5-18 dBA). The sound pressure level frequency response of the two MR systems showed a similar harmonic structure above 200 Hz for all imaging sequences. The amplitude, however, was appreciably lower for the NICU scanner, by as much as 30 dB, for frequencies below 200 Hz. The NICU MRI system is quieter than conventional MRI scanners, improving safety for the neonate and facilitating siting of the unit within the NICU.
引用
收藏
页码:1011 / 1019
页数:9
相关论文
共 50 条
  • [31] Noise in a Neonatal Intensive Care Unit: Exploring Its State and Solutions
    Sabetsarvestani, Raheleh
    Kose, Semra
    Geckil, Emine
    Tosun, Esra Erturk
    Ozkilicaslan, Fatma Tokan
    Karaarslan, Fatih
    Altunhan, Huseyin
    ADVANCES IN NEONATAL CARE, 2022, 22 (06) : E183 - E190
  • [32] Determination of Noise Level and Its Sources in the Neonatal Intensive Care Unit and Neonatal Ward
    Blourchian, Mahdi Jahangir
    Sharafi, Reza
    IRANIAN JOURNAL OF NEONATOLOGY, 2015, 6 (04) : 21 - 24
  • [33] Nutritional characterization in patients hospitalized in the neonatal intensive care unit
    Espitia, Olga Lucia Pinzon
    Soriano, Luisa Fernanda Murcia
    Gonzalez, Claudia Lucia
    Corzo, Carlos Julio
    NUTRICION CLINICA Y DIETETICA HOSPITALARIA, 2024, 44 (01): : 194 - 200
  • [34] Neonatal pain: characterization of the physiotherapist's perception in the Neonatal Intensive Care Unit
    Gimenez, Isabelle Leandro
    Neves Moreira Arakaki, Vanessa da Silva
    Correa, Raquel Miranda
    dos Santos, Rosana Silva
    Peres, Rodrigo Tosta
    Sant'Anna, Clemax Couto
    Ferreira, Halina Cidrini
    REVISTA PAULISTA DE PEDIATRIA, 2020, 38
  • [35] Noise in an intensive care unit
    Salandin, Andrea
    Arnold, Joerg
    Kornadt, Oliver
    JOURNAL OF THE ACOUSTICAL SOCIETY OF AMERICA, 2011, 130 (06): : 3754 - 3760
  • [36] Noise survey of neonatal intensive care unit at a government tertiary-care centre
    Lokwani, P.
    Gupta, N.
    Choudhary, S. K.
    Singh, A. K.
    JOURNAL OF NEONATAL-PERINATAL MEDICINE, 2023, 16 (04) : 619 - 625
  • [37] MULTIORGAN SYSTEM FAILURE IN THE NEONATAL INTENSIVE-CARE UNIT
    STRING, DI
    TEPAS, JJ
    MOLLITT, DL
    CLINICAL RESEARCH, 1989, 37 (01): : A70 - A70
  • [38] Feasibility of and experience using a portable MRI scanner in the neonatal intensive care unit
    Sien, Maura E.
    Robinson, Amie L.
    Hu, Houchun H.
    Nitkin, Chris R.
    Hall, Ara S.
    Files, Marcie G.
    Artz, Nathan S.
    Pitts, John T.
    Chan, Sherwin S.
    ARCHIVES OF DISEASE IN CHILDHOOD-FETAL AND NEONATAL EDITION, 2023, 108 (01): : F45 - F50
  • [39] Palliative Care in the Neonatal Intensive Care Unit
    Lenington, Kathryn
    Dudding, Katherine M.
    Fazeli, Pariya L.
    Dick, Tracey
    Patrician, Patricia
    ADVANCES IN NEONATAL CARE, 2024, 24 (02) : 187 - 194
  • [40] Assessment of environmental noise and its effect on neonates in a Neonatal Intensive Care Unit
    Romeu, Jordi
    Cotrina, Laura
    Perapoch, Josep
    Lines, Marcos
    APPLIED ACOUSTICS, 2016, 111 : 161 - 169