Noninvasive Carotid Strain Imaging Using Angular Compounding at Large Beam Steered Angles: Validation in Vessel Phantoms

被引:99
作者
Hansen, Hendrik H. G. [1 ]
Lopata, Richard G. P. [1 ]
de Korte, Chris L. [1 ]
机构
[1] Radboud Univ Nijmegen, Med Ctr, Clin Phys Lab, Dept Pediat, NL-6500 HB Nijmegen, Netherlands
关键词
Angular compounding; beam steering; grating lobes; noninvasive ultrasound strain imaging; vulnerable plaque; INTRAVASCULAR ULTRASOUND ELASTOGRAPHY; VULNERABLE PLAQUE; IN-VIVO; DISPLACEMENT; ELASTICITY; SPEED; SOUND; SEGMENTATION; ARTERIES; NOISE;
D O I
10.1109/TMI.2008.2011510
中图分类号
TP39 [计算机的应用];
学科分类号
081203 ; 0835 ;
摘要
Stroke and myocardial infarction are initiated by rupturing vulnerable atherosclerotic plaques. With noninvasive ultrasound elastography, these plaques might be detected in carotid arteries. However, since the ultrasound beam is generally not aligned with the radial direction in which the artery pulsates, radial and circumferential strains need to be derived from axial and lateral data. Conventional techniques to perform this conversion have the disadvantage that lateral strain is required. Since the lateral strain has relatively poor accuracy, the quality of the radial and circumferential strains is reduced. In this study, the radial and circumferential strain estimates are improved by combining axial strain data acquired at multiple insonitication angles. Adaptive techniques to correct for grating lobe interference and other artifacts that occur when performing beam steering at large angles are introduced. Acquisitions at multiple angles are performed with a beam steered linear array. For each beam steered angle, there are two spatially restricted regions of the circular vessel cross section where the axial strain is closely aligned with the radial strain and two spatially restricted regions (different from the radial strain regions) where the axial strain is closely aligned with the circumferential strain. These segments with high quality strain estimates are compounded to form radial or circumferential strain images. Compound radial and circumferential strain images were constructed for a homogeneous vessel phantom with a concentric lumen subjected to different intraluminal pressures. Comparison of the elastographic signal-to-noise ratio (SNRe) and contrast-to-noise ratio (CNRe) revealed that compounding increases the image quality considerably compared to images from 0 degrees information only. SNRe and CNRe increase up to 2.7 and 6.6 dB, respectively. The highest image quality was achieved by projecting axial data, completed with a small segment determined by either principal component analysis or by application of a rotation matrix.
引用
收藏
页码:872 / 880
页数:9
相关论文
共 39 条
[1]   Finite element modeling and intravascular ultrasound elastography of vulnerable plaques: parameter variation [J].
Baldewsing, RA ;
de Korte, CL ;
Schaar, JA ;
Mastik, F ;
van der Steen, AFW .
ULTRASONICS, 2004, 42 (1-9) :723-729
[2]   A finite element model for performing intravascular ultrasound elastography of human atherosclerotic coronary arteries [J].
Baldewsing, RA ;
De Korte, CL ;
Schaar, JA ;
Mastik, F ;
Van Der Steen, AFW .
ULTRASOUND IN MEDICINE AND BIOLOGY, 2004, 30 (06) :803-813
[3]   Fully automatic luminal contour segmentation in intracoronary ultrasound imaging - A statistical approach [J].
Brusseau, E ;
de Korte, CL ;
Mastik, F ;
Schaar, J ;
van der Steen, AFW .
IEEE TRANSACTIONS ON MEDICAL IMAGING, 2004, 23 (05) :554-566
[4]   REDUCTION OF IMAGE NOISE IN ELASTOGRAPHY [J].
CESPEDES, I ;
OPHIR, J .
ULTRASONIC IMAGING, 1993, 15 (02) :89-102
[5]   DISTRIBUTION OF CIRCUMFERENTIAL STRESS IN RUPTURED AND STABLE ATHEROSCLEROTIC LESIONS - A STRUCTURAL-ANALYSIS WITH HISTOPATHOLOGICAL CORRELATION [J].
CHENG, GC ;
LOREE, HM ;
KAMM, RD ;
FISHBEIN, MC ;
LEE, RT .
CIRCULATION, 1993, 87 (04) :1179-1187
[6]   Stability and instability: Two faces of coronary atherosclerosis - The Paul Dudley White Lecture 1995 [J].
Davies, MJ .
CIRCULATION, 1996, 94 (08) :2013-2020
[7]   Intravascular ultrasound elastography: an overview [J].
de Korte, CL ;
van der Steen, AFW .
ULTRASONICS, 2002, 40 (1-8) :859-865
[8]   Identification of atherosclerotic plaque components with intravascular ultrasound elastography in vivo A Yucatan pig study [J].
de Korte, CL ;
Sierevogel, MJ ;
Mastik, F ;
Strijder, C ;
Schaar, JA ;
Velema, E ;
Pasterkamp, G ;
Serruys, PW ;
van der Steen, AFW .
CIRCULATION, 2002, 105 (14) :1627-1630
[9]   Influence of catheter position on estimated strain in intravascular elastography [J].
de Korte, CL ;
Céspedes, EI ;
van der Steen, AFW .
IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1999, 46 (03) :616-625
[10]   Elastic and acoustic properties of vessel mimicking material for elasticity imaging [J].
deKorte, CL ;
Cespedes, EI ;
vanderSteen, AFW ;
Norder, B ;
Nijenhuis, KT .