An evaluation of automated blood collection mixers

被引:2
作者
van der Meer, P. F. [1 ]
Pietersz, R. N. [1 ]
机构
[1] Sanquin Blood Bank NW Reg, NL-1066 CX Amsterdam, Netherlands
关键词
collection mixers; whole blood;
D O I
10.1111/j.1423-0410.2006.00831.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background and Objectives We investigated the mixing capacity of two whole blood (WB) collection mixers. Material and Methods WB was simulated by using a 25% glycerol solution warmed to 35 degrees C. Citrate-phosphate-dextrose (CPD) anticoagulant of a collection system was stained with toluidine blue, and simulated WB was added at 30, 60 or 90 ml/min, respectively (n = 3 per flow speed). The optical density (OD) of 10-ml fractions was measured, and results are expressed as percentage of a well-mixed '100%-sample'. Results CompoGuard showed adequate mixing at all three flow speeds (average ODs 96-103%). HemoLight showed good mixing at 60 and 90 ml/min (ODs from 97 to 101%). At 30 ml/min, mixing appeared suboptimal, but still conformed to our requirements with ODs from 96% to 104%. Conclusions Both mixers give sufficient mixing of whole blood with anticoagulant.