Lesion formation and visualization using dual-mode ultrasound phased arrays

被引:0
作者
Ebbini, ES [1 ]
Bischof, JC [1 ]
Coad, JE [1 ]
机构
[1] Univ Minnesota, Dept Elect & Comp Engn, Minneapolis, MN 55455 USA
来源
2001 IEEE ULTRASONICS SYMPOSIUM PROCEEDINGS, VOLS 1 AND 2 | 2001年
关键词
D O I
暂无
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
A 1 MHz spherical-section 64-element linear piezo-composite phased array and a supporting driving and data acquisition system were recently tested for use as a dual-mode HIFU applicator system. In therapeutic mode, the army was shown to be capable of producing focal intensities in excess of 3500 W/cm(2) at its geometric center (100 mm radius of cur. vature). Imaging tests of quality assurance tissue-mimicking phantoms as well as computer simulations confirmed that the array has an oval-shaped imaging field of view (IxFOV) with nearly 50 dB dynamic range extending from 70 to 120 mm in the axial direction and 20 mm in the transverse direction [1]. This is larger than Its therapeutic operating field (ThxOF), which is defined as the region around the geometric focus where the loss in intensity gain is within 1 dB from the intensity gain at the geometric center. We have tested the array imaging capability in the visualization of the formation of discrete and volumetric lesions In freshly excised porcine liver samples. Experimental results clearly show that the enhancement of the tissue echogenecity at the lesion location is more pronounced in the harmonic images when compared to images at the fundamental. Results also clearly show that the harmonic imaging mode produces more accurate mapping of the lesion size and shape as determined by histologic evaluation.
引用
收藏
页码:1351 / 1354
页数:4
相关论文
共 10 条
  • [1] EBBINI E, 1998, SPIE BIOS SURG APPL
  • [2] Ebbini ES., 1990, THESIS U ILLINOIS
  • [3] HYNYNEN K, 1997, 1997 IEEE ULTR S, V2, P1305
  • [4] Jenne J, 1997, IEEE ULTR S
  • [5] Microwave imaging for tissue assessment: Initial evaluation in multitarget tissue-equivalent phantoms
    Meaney, PM
    Paulsen, KD
    Hartov, A
    Crane, RK
    [J]. IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1996, 43 (09) : 878 - 890
  • [6] Initial in vivo experience with EIT as a thermal estimator during hyperthermia
    Paulsen, KD
    Moskowitz, MJ
    Ryan, TP
    Mitchell, SE
    Hoopes, PJ
    [J]. INTERNATIONAL JOURNAL OF HYPERTHERMIA, 1996, 12 (05) : 573 - 591
  • [7] PHUKPATTARNONT P, 2001, IEEE ENG MED BIOL IS
  • [8] Noninvasive surgery of prostate tissue by high-intensity focused ultrasound
    Sanghvi, NT
    Fry, FJ
    Bihrle, R
    Foster, RS
    Phillips, MH
    Syrus, J
    Zaitsev, AV
    Hennige, CW
    [J]. IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 1996, 43 (06) : 1099 - 1110
  • [9] SIMON C, 1998, SPIE MED IMAGING 199, V3341
  • [10] Vanbaren P, 1996, IEEE ULTR S NOV, P1269