Evaluation of frequency and type of errors detected by a computerized record and verify system during radiation treatment

被引:38
作者
Barthelemy-Brichant, N
Sabatier, J
Dewé, W
Albert, A
Deneufbourg, JM
机构
[1] Ctr Hosp Univ Liege, Dept Radiotherapy, B-4000 Liege, Belgium
[2] Ctr Hosp Univ Liege, Dept Med Informat & Stat, B-4000 Liege, Belgium
关键词
radiation therapy; errors; quality control; record and verify system;
D O I
10.1016/S0167-8140(99)00141-3
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Background: Computerized record and verify systems (RVS) have bean introduced to improve the precision of radiation treatment delivery. These systems prevent the delivery of ionizing radiations when the settings of the treatment machine do not match the intended parameters within some maximal authorized deviation. Purpose: To assess the potential alteration of the frequency of errors associated with the use of RVS during radiation treatment delivery. Materials and methods: The software of the RVS was altered in order to record the settings actually used for radiation treatment delivery whereas the verification function was suppressed. At the end of the study period, the settings used during daily administration of radiation treatment were compared to the parameters recorded in the RVS using the computer. They were also compared with the planned ones written in the patient treatment chart. Results: Out of the 147 476 parameters examined during the study period; 678 (0.46%) were set erroneously. At least one error occurred in 628 (3.22%) of the 19 512 treated fields, an erroneous parameter was introduced in the RVS memory in 22 (1.17%) of the 1885 fields. Conclusions: RVS has the potential to improve precision of radiation treatment delivery by detecting a significant number of setting errors. However, excessive confidence in RVS could lead to repeated errors as there is a potential for the entry of erroneous parameters into the RVS memory. (C) 1999 Elsevier Science Ireland Ltd. All rights reserved.
引用
收藏
页码:149 / 154
页数:6
相关论文
共 16 条
  • [1] WHEN AND HOW CAN WE IMPROVE PRECISION IN RADIOTHERAPY
    DUTREIX, A
    [J]. RADIOTHERAPY AND ONCOLOGY, 1984, 2 (04) : 275 - 292
  • [2] The impact of treatment complexity and computer-control delivery technology on treatment delivery errors
    Fraass, BA
    Lash, KL
    Matrone, GM
    Volkman, SK
    McShan, DL
    Kessler, ML
    Lichter, AS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (03): : 651 - 659
  • [3] EXPERIENCE WITH COMPUTER MONITORING, VERIFICATION AND RECORD KEEPING IN RADIOTHERAPY PROCEDURES USING A CLINAC-4
    FREDRICKSON, DH
    KARZMARK, CJ
    RUST, DC
    TUSCHMAN, M
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1979, 5 (03): : 415 - 418
  • [4] GREEN R, 1995, RECENT ADV ANAESTHES, P1
  • [5] *ICRU 24 INT COMM, 1993, 50 ICRU
  • [6] ACCURACY IN RADIOTHERAPY TREATMENT
    KARTHA, PKI
    CHUNGBIN, A
    WACHTOR, T
    HENDRICKSON, FR
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1977, 2 (7-8): : 797 - 799
  • [7] A change in treatment process with a modern record and verify system
    Klein, EE
    Drzymala, RE
    Williams, R
    Westfall, LA
    Purdy, JA
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1998, 42 (05): : 1163 - 1168
  • [8] HUMAN ERRORS IN DATA TRANSFER DURING THE PREPARATION AND DELIVERY OF RADIATION TREATMENT AFFECTING THE FINAL RESULT - GARBAGE IN, GARBAGE OUT
    LEUNENS, G
    VERSTRAETE, J
    VANDENBOGAERT, W
    VANDAM, J
    DUTREIX, A
    VANDERSCHUEREN, E
    [J]. RADIOTHERAPY AND ONCOLOGY, 1992, 23 (04) : 217 - 222
  • [9] Error rates in clinical radiotherapy
    Macklis, RM
    Meier, T
    Weinhous, MS
    [J]. JOURNAL OF CLINICAL ONCOLOGY, 1998, 16 (02) : 551 - 556
  • [10] A COMPUTERIZED RECORD AND VERIFY SYSTEM FOR RADIATION TREATMENTS
    MOHAN, R
    PODMANICZKY, KC
    CALEY, R
    LAPIDUS, A
    LAUGHLIN, JS
    [J]. INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1984, 10 (10): : 1975 - 1985