In vitro fragmentation efficiency of holmium: yttrium-aluminum-garnet (YAG) laser lithotripsy - a comprehensive study encompassing different frequencies, pulse energies, total power levels and laser fibre diameters

被引:80
作者
Kronenberg, Peter [1 ]
Traxer, Olivier [2 ]
机构
[1] Hosp Prof Doutor Fernando Fonseca, Dept Urol, P-2720276 Amadora, Portugal
[2] Univ Paris 06, Hop Tenon, Paris, France
关键词
holmium:YAG laser lithotripsy; urinary calculi; high frequency lithotripsy; lithotripter settings; automated laser fragmentation testing system; CALCULUS RETROPULSION; STONE FRAGMENTATION; DEPENDENCE; VARIES;
D O I
10.1111/bju.12567
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objective To assess the fragmentation (ablation) efficiency of laser lithotripsy along a wide range of pulse energies, frequencies, power settings and different laser fibres, in particular to compare high- with low-frequency lithotripsy using a dynamic and innovative testing procedure free from any human interaction bias. Materials and Methods An automated laser fragmentation testing system was developed. The unmoving laser fibres fired at the surface of an artificial stone while the stone was moved past at a constant velocity, thus creating a fissure. The lithotripter settings were 0.2-1.2 J pulse energies, 5-40 Hz frequencies, 4-20 W power levels, and 200 and 550 mu m core laser fibres. Fissure width, depth, and volume were analysed and comparisons between laser settings, fibres and ablation rates were made. Results Low frequency-high pulse energy (LoFr-HiPE) settings were (up to six times) more ablative than high frequency-low pulse energy (HiFr-LoPE) at the same power levels (P < 0.001), as they produced deeper (P < 0.01) and wider (P < 0.001) fissures. There were linear correlations between pulse energy and fragmentation volume, fissure width, and fissure depth (all P < 0.001). Total power did not correlate with fragmentation measurements. Laser fibre diameter did not affect fragmentation volume (P = 0.81), except at very low pulse energies (0.2 J), where the large fibre was less efficient (P = 0.015). Conclusions At the same total power level, LoFr-HiPE lithotripsy was most efficient. Pulse energy was the key variable that drove fragmentation efficiency. Attention must be paid to prevent the formation of time-consuming bulky debris and adapt the lithotripter settings to one's needs. As fibre diameter did not affect fragmentation efficiency, small fibres are preferable due to better scope irrigation and manoeuvrability.
引用
收藏
页码:261 / 267
页数:7
相关论文
共 29 条
  • [1] In vitro comparison of fragmentation efficiency of flexible pneumatic lithotripsy using 2 flexible ureteroscopes
    Auge, BK
    Sekula, JJ
    Springhart, WP
    Zhu, SL
    Zhong, P
    Preminger, GM
    [J]. JOURNAL OF UROLOGY, 2004, 172 (03) : 967 - 970
  • [2] Comparison of holmium:YAG and thulium fiber laser lithotripsy: ablation thresholds, ablation rates, and retropulsion effects
    Blackmon, Richard L.
    Irby, Pierce B.
    Fried, Nathaniel M.
    [J]. JOURNAL OF BIOMEDICAL OPTICS, 2011, 16 (07)
  • [3] Holmium:YAG (λ=2,120nm) Versus Thulium Fiber (λ=1,908nm) Laser Lithotripsy
    Blackmon, Richard L.
    Irby, Pierce B.
    Fried, Nathaniel M.
    [J]. LASERS IN SURGERY AND MEDICINE, 2010, 42 (03) : 232 - 236
  • [4] Thulium Fiber Laser Lithotripsy Using Tapered Fibers
    Blackmon, Richard L.
    Irby, Pierce B.
    Fried, Nathaniel M.
    [J]. LASERS IN SURGERY AND MEDICINE, 2010, 42 (01) : 45 - 50
  • [5] Effectiveness of high-frequency holmium: YAG laser stone fragmentation: The "popcorn effect"
    Chawla, Sam N.
    Chang, Mark F.
    Chang, Andrew
    Lenoir, James
    Bagley, Demetrius H.
    [J]. JOURNAL OF ENDOUROLOGY, 2008, 22 (04) : 645 - 649
  • [6] CHUONG CJC, 1992, J UROLOGY, V148, P200
  • [7] Laser lithotripsy and cyanide
    Corbin, NS
    Teichman, JMH
    Nguyen, T
    Glickman, RD
    Rihbany, L
    Pearle, MS
    Bishoff, JT
    [J]. JOURNAL OF ENDOUROLOGY, 2000, 14 (02) : 169 - 173
  • [8] In vitro analysis of stone fragmentation ability of the FREDDY laser
    Delvecchio, FC
    Auge, BK
    Brizuela, RM
    Weizer, AZ
    Zhong, P
    Preminger, GM
    [J]. JOURNAL OF ENDOUROLOGY, 2003, 17 (03) : 177 - 179
  • [9] *EAU GUID OFF, 2013, 28 EAU ANN C MIL ARN
  • [10] *EAU GUID OFF, 2011, 26 EAU ANN C VIENN A