Fixation of Extra-Articular Distal Humerus Fractures Using One Locking Plate Versus Two Reconstruction Plates: A Laboratory Study

被引:35
|
作者
Tejwani, Nirmal C. [1 ]
Murthy, Anjali [1 ]
Park, Jason [1 ]
McLaurin, Toni M. [1 ]
Egol, Kenneth A. [1 ]
Kummer, Fred J. [1 ]
机构
[1] NYU, Hosp Joint Dis, Dept Orthopaed, New York, NY 10016 USA
来源
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE | 2009年 / 66卷 / 03期
关键词
Distal humerus fracture; Osteosynthesis; Locking plate; Plating; INVASIVE STABILIZATION SYSTEM; BIOMECHANICAL EVALUATION; INTERNAL-FIXATION; SHAFT FRACTURES; DIAPHYSEAL FRACTURES; SURGICAL EXPERIENCE; FEMUR FRACTURES; TIBIA FRACTURES;
D O I
10.1097/TA.0b013e318181e53c
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Objective: To compare the strength and stiffness of standard double-plate fixation with that of a single-locking plate for comminuted extra-articular distal humeral fractures. Materials: Eight matched pairs of humeri were used: One of each pair was fixed with two 3.5-mm standard reconstruction plates applied dorsally; the other was fixed with one precontoured locking plate applied to the posterior aspect of the lateral column. A 1-cm gap was created to simulate a distal fracture model with 100% metaphyseal comminution with no bony apposition. Stiffness testing of these constructs was performed in axial compression, anterior, posterior and lateral bending, and torsion. They were then cyclically loaded for 40110 cycles with 60 N in posterior bending and the stiffnesses re-tested. Finally, each construct was posteriorly loaded to failure to determine its ultimate strength. Results: The double-plate construct was significantly stiffer than the one locking plate construct in anterior bending (39%; p = 0.02), posterior bending (23%; p = 0.04), and lateral bending (60%; p = 0.01). No significant stiffness differences were seen in axial compression and torsion (p = 0.64 and 0.25, respectively). After cyclic loading, all construct stiffnesses were relatively unchanged. Both construct types had similar failure strengths (p = 0.76) of approximately 400 N. Conclusions: Double-plating provides a more rigid fixation than a single-locked plate for fixation of extra-articular comminuted distal humeral fractures. This could be clinically relevant in situations with 100% comminution as seen in gun shot injuries.
引用
收藏
页码:795 / 799
页数:5
相关论文
共 50 条
  • [1] Locking plate fixation versus antegrade intramedullary nailing for the treatment of extra-articular distal femoral fractures
    Ocalan, E.
    Ustun, C. C.
    Aktuglu, K.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2019, 50 : 55 - 62
  • [2] Locking versus non-locking plates in fixation of extra-articular distal humerus fracture: a randomized controlled study
    Galal, Sherif
    Mattar, Yasser
    Solyman, Alaa Mohy Eldin
    Ezzat, Mostafa
    INTERNATIONAL ORTHOPAEDICS, 2020, 44 (12) : 2761 - 2767
  • [3] Functional outcome of extra-articular distal humerus fracture fixation using a single locking plate: A retrospective study
    Trikha, Vivek
    Agrawal, Prabhat
    Das, Saubhik
    Gaba, Sahil
    Kumar, Arvind
    JOURNAL OF ORTHOPAEDIC SURGERY, 2017, 25 (03):
  • [4] Minimally-invasive internal fixation of extra-articular distal femur fractures using a locking plate: Tricks of the trade
    Ehlinger, M.
    Adam, P.
    Abane, L.
    Arlettaz, Y.
    Bonnomet, F.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2011, 97 (02) : 201 - 205
  • [5] Management of Extra-articular Fractures of the Distal Tibia: Intramedullary Nailing Versus Plate Fixation
    Casstevens, Chris
    Le, Toan
    Archdeacon, Michael T.
    Wyrick, John D.
    JOURNAL OF THE AMERICAN ACADEMY OF ORTHOPAEDIC SURGEONS, 2012, 20 (11) : 675 - 683
  • [6] Plate selection for fixation of extra-articular distal humerus fractures: A biomechanical comparison of three different implants
    Scolaro, John A.
    Hsu, Jason E.
    Svach, David J.
    Mehta, Samir
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2014, 45 (12): : 2040 - 2044
  • [7] Fixation of extra-articular distal humeral fractures with a lateral approach and a locked plate: an alternative method
    Parmaksizoglu, Atilla Sancar
    Ozkaya, Ufuk
    Bilgili, Fuat
    Mutlu, Harun
    Cetin, Umit
    ACTA ORTHOPAEDICA ET TRAUMATOLOGICA TURCICA, 2016, 50 (02) : 132 - 138
  • [8] Volar Locking Plate Versus Dorsal Locking Nail-Plate Fixation for Dorsally Displaced Unstable Extra-Articular Distal Radial Fractures
    Eikrem, Morten
    Brannsten, Hege
    Bjorkoy, Dagfinn
    Lian, Tom
    Madsen, Jan Erik
    Figved, Wender
    JBJS OPEN ACCESS, 2021, 6 (04)
  • [9] Risk factors for hardware-related complications after extra-articular distal humerus fracture fixation using an anatomical locking plate
    Jung, Hyoung-Seok
    Won, Yoo-Sun
    Choi, Yang-Seon
    Lee, Jae-Sung
    EUROPEAN JOURNAL OF TRAUMA AND EMERGENCY SURGERY, 2023, 49 (01) : 125 - 131
  • [10] Minimally-invasive fixation of distal extra-articular femur fractures with locking plates: Limitations and failures
    Ehlinger, M.
    Adam, P.
    Arlettaz, Y.
    Moor, B. K.
    DiMarco, A.
    Brinkert, D.
    Bonnomet, F.
    ORTHOPAEDICS & TRAUMATOLOGY-SURGERY & RESEARCH, 2011, 97 (06) : 668 - 674