Dual-energy CT quantitative imaging: a comparison study between twin-beam and dual-source CT scanners

被引:111
作者
Almeida, Isabel P. [1 ]
Schyns, Lotte E. J. R. [1 ]
Ollers, Michel C. [1 ]
van Elmpt, Wouter [1 ]
Parodi, Katia [2 ]
Landry, Guillaume [2 ]
Verhaegen, Frank [1 ]
机构
[1] Maastricht Univ, Med Ctr, GROW Sch Oncol & Dev Biol, Dept Radiat Oncol MAASTRO, Maastricht, Netherlands
[2] Ludwig Maximilians Univ Munchen, Dept Med Phys, Fac Phys, Coulombwall 1, D-85748 Garching, Germany
关键词
Dual-energy dual-source CT; Dual-energy twin-beam CT; effective atomic number; relative electron density; CARLO DOSE CALCULATIONS; X-RAY TARGETS; TISSUE SEGMENTATION; ATOMIC NUMBERS; SIMULATION; QUALITY; IMAGES; TUBE;
D O I
10.1002/mp.12000
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess image quality and to quantify the accuracy of relative electron densities (rho(e)) and effective atomic numbers (Z(eff)) for three dual-energy computed tomography (DECT) scanners: a novel single-source split-filter (i. e., twin-beam) and two dual-source scanners. Methods: Measurements were made with a second generation dual-source scanner at 80/140Sn kVp, a third-generation twin-beam single-source scanner at 120 kVp with gold (Au) and tin (Sn) filters, and a third-generation dual-source scanner at 90/150Sn kVp. Three phantoms with tissue inserts were scanned and used for calibration and validation of parameterized methods to extract rho(e) and Z(eff), whereas iodine and calcium inserts were used to quantify Contrast-to-Noise-Ratio (CNR). Spatial resolution in tomographic images was also tested. Results: The third-generation scanners have an image resolution of 6.2, similar to 0.5 lp/cm higher than the second generation scanner. The twin-beam scanner has low imaging contrast for iodine materials due to its limited spectral separation. The parameterization methods resulted in calibrations with low fit residuals for the dual-source scanners, yielding values of rho(e) and Z(eff) close to the reference values (errors within 1.2% for rho(e) and 6.2% for Z(eff) for a dose of 20 mGy, excluding lung substitute tissues). The twin-beam scanner presented overall higher errors (within 3.2% for rho(e) and 28% for Z(eff), also excluding lung inserts) and also larger variations for uniform inserts. Conclusions: Spatial resolution is similar for the three scanners. The twin-beam is able to derive qe and Z(eff), but with inferior accuracy compared to both dual-source scanners. (C) 2016 American Association of Physicists in Medicine
引用
收藏
页码:171 / 179
页数:9
相关论文
共 21 条
[1]   Dual-energy CT-based material extraction for tissue segmentation in Monte Carlo dose calculations [J].
Bazalova, Magdalena ;
Carrier, Jean-Francois ;
Beaulieu, Luc ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2008, 53 (09) :2439-2456
[2]  
Christe A, 2013, WORLD J RADIOL, V5, P421, DOI 10.4329/wjr.v5.i11.421
[3]   Initial Results of a Single-Source Dual-Energy Computed Tomography Technique Using a Split-Filter Assessment of Image Quality, Radiation Dose, and Accuracy of Dual-Energy Applications in an In Vitro and In Vivo Study [J].
Euler, Andre ;
Parakh, Anushri ;
Falkowski, Anna L. ;
Manneck, Sebastian ;
Dashti, David ;
Krauss, Bernhard ;
Szucs-Farkas, Zsolt ;
Schindera, Sebastian T. .
INVESTIGATIVE RADIOLOGY, 2016, 51 (08) :491-498
[4]   Dealing with Uncertainty in CT Images [J].
Fletcher, Joel G. ;
Leng, Shuai ;
Yu, Lifeng ;
McCollough, Cynthia H. .
RADIOLOGY, 2016, 279 (01) :5-10
[5]   Dual-source spiral CT with pitch up to 3.2 and 75 ms temporal resolution: Image reconstruction and assessment of image quality [J].
Flohr, Thomas G. ;
Leng, Shuai ;
Yu, Lifeng ;
Allmendinger, Thomas ;
Bruder, Herbert ;
Petersilka, Martin ;
Eusemann, Christian D. ;
Stierstorfer, Karl ;
Schmidt, Bernhard ;
McCollough, Cynthia H. .
MEDICAL PHYSICS, 2009, 36 (12) :5641-5653
[6]   A simple approach to measure computed tomography (CT) modulation transfer function (MTF) and noise-power spectrum (NPS) using the American College of Radiology (ACR) accreditation phantom [J].
Friedman, Saul N. ;
Fung, George S. K. ;
Siewerdsen, Jeffrey H. ;
Tsui, Benjamin M. W. .
MEDICAL PHYSICS, 2013, 40 (05)
[7]   Experimental verification of ion stopping power prediction from dual energy CT data in tissue surrogates [J].
Huenemohr, Nora ;
Krauss, Bernhard ;
Tremmel, Christoph ;
Ackermann, Benjamin ;
Jaekel, Oliver ;
Greilich, Steffen .
PHYSICS IN MEDICINE AND BIOLOGY, 2014, 59 (01) :83-96
[8]  
Landry G, 2015, MED PHYS, V26, P75
[9]   Deriving effective atomic numbers from DECT based on a parameterization of the ratio of high and low linear attenuation coefficients [J].
Landry, Guillaume ;
Seco, Joao ;
Gaudreault, Mathieu ;
Verhaegen, Frank .
PHYSICS IN MEDICINE AND BIOLOGY, 2013, 58 (19) :6851-6866
[10]   Extracting atomic numbers and electron densities from a dual source dual energy CT scanner: Experiments and a simulation model [J].
Landry, Guillaume ;
Reniers, Brigitte ;
Granton, Patrick Vincent ;
van Rooijen, Bart ;
Beaulieu, Luc ;
Wildberger, Joachim E. ;
Verhaegen, Frank .
RADIOTHERAPY AND ONCOLOGY, 2011, 100 (03) :375-379