In vivo comparison of hip separation after metal-on-metal or metal-on-polyethylene total hip arthroplasty

被引:90
作者
Komistek, RD [1 ]
Dennis, DA [1 ]
Ochoa, JA [1 ]
Haas, BD [1 ]
Hammill, C [1 ]
机构
[1] Rocky Mtn Musculoskeletal Res Lab, Denver, CO 80222 USA
关键词
D O I
10.2106/00004623-200210000-00015
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
Background: Twenty subjects were analyzed in vivo with use of video fluoroscopy to determine if the femoral head separates from the acetabular component during normal gait and to determine if the amount of separation differs between metal-on-metal and meta I-on-polyethylene total hip prostheses. Methods: Ten subjects had been treated with a metal-on-metal total hip arthroplasty and ten, with a meta I-on-polyethylene total hip arthroplasty. All of the prostheses were implanted by the same surgeon utilizing the same surgical technique, and all were judged to be clinically successful (a Harris hip score of >90 points). Each subject walked with a normal gait on a level treadmill while under fluoroscopic surveillance. The two-dimensional fluoroscopic videotapes were then converted into three-dimensional images with use of a computer-automated model-fitting technique. Each implant was analyzed at various flexion angles to assess the amount of femoral head sliding. Results: No femoral head sliding was observed in the subjects with a metal-on-metal implant, whereas all ten subjects with a meta I-on-polyethylene implant had sliding that was greater than our threshold value of 0.75 mm. The average amount of femoral head sliding in these subjects was 2.0 mm, and the sliding was observed during the swing phase of gait. The sliding was typically seen medially while the femoral head remained in contact with the acetabular component superolaterally. Conclusions: Femoral head sliding commonly occurred following traditional meta I-on-polyethylene total hip arthroplasty but not after metal-on-metal arthroplasty. These kinematic data may be of value in future hip-simulation studies to better duplicate wear patterns observed in retrieval analyses, assist in the understanding of the lubrication and wear rates of metal-on-metal designs, and facilitate designing of prosthetic components that minimize wear and optimize hip kinematics.
引用
收藏
页码:1836 / 1841
页数:6
相关论文
共 38 条
[1]   HIP-JOINT LOADING DURING WALKING AND RUNNING, MEASURED IN 2 PATIENTS [J].
BERGMANN, G ;
GRAICHEN, F ;
ROHLMANN, A .
JOURNAL OF BIOMECHANICS, 1993, 26 (08) :969-990
[2]  
Bergmann G, 1997, CLIN ORTHOP RELAT R, P190
[3]  
Bono J V, 1994, J Arthroplasty, V9, P119, DOI 10.1016/0883-5403(94)90059-0
[4]   A MODEL OF LOWER-EXTREMITY MUSCULAR ANATOMY [J].
BRAND, RA ;
CROWNINSHIELD, RD ;
WITTSTOCK, CE ;
PEDERSEN, DR ;
CLARK, CR ;
VANKRIEKEN, FM .
JOURNAL OF BIOMECHANICAL ENGINEERING-TRANSACTIONS OF THE ASME, 1982, 104 (04) :304-310
[5]   Charnley wear model for validation of hip simulators - Ball diameter versus polytetrafluoroethylene and polyethylene wear [J].
Clarke, IC ;
Good, V ;
Anissian, L ;
Gustafson, A .
PROCEEDINGS OF THE INSTITUTION OF MECHANICAL ENGINEERS PART H-JOURNAL OF ENGINEERING IN MEDICINE, 1997, 211 (01) :25-36
[6]  
CLARKE IC, 1991, HIP ARTHROPLASTY, P535
[7]  
CLARKE IC, 1991, HIP ARTHROPLASTY, P555
[8]   BIOMECHANICAL INVESTIGATION OF HUMAN HIP [J].
CROWNINSHIELD, RD ;
JOHNSTON, RC ;
ANDREWS, JG ;
BRAND, RA .
JOURNAL OF BIOMECHANICS, 1978, 11 (1-2) :75-85
[9]   TELEMETRIC FORCE MEASUREMENTS ACROSS THE HIP AFTER TOTAL ARTHROPLASTY [J].
DAVY, DT ;
KOTZAR, GM ;
BROWN, RH ;
HEIPLE, KG ;
GOLDBERG, VM ;
HEIPLE, KG ;
BERILLA, J ;
BURSTEIN, AH .
JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 1988, 70A (01) :45-50
[10]   In vivo determination of hip joint separation and the forces generated due to impact loading conditions [J].
Dennis, DA ;
Komistek, RD ;
Northcut, EJ ;
Ochoa, JA ;
Ritchie, A .
JOURNAL OF BIOMECHANICS, 2001, 34 (05) :623-629