Accuracy of intracranial pressure monitoring: systematic review and meta-analysis

被引:62
作者
Zacchetti, Lucia [1 ]
Magnoni, Sandra [2 ]
Di Corte, Federica [1 ]
Zanier, Elisa R. [3 ]
Stocchetti, Nino [1 ,2 ]
机构
[1] Univ Milan, Dept Physiopathol & Transplant, I-20122 Milan, Italy
[2] Ca Granda Osped Maggiore Policlin, Fdn IRCCS, Dept Anesthesiol & Intens Care, I-20122 Milan, Italy
[3] IRCCS Ist Mario Negri, Dept Neurosci, I-20156 Milan, Italy
来源
CRITICAL CARE | 2015年 / 19卷
关键词
CLINICAL-EVALUATION; CODMAN MICROSENSOR; ZERO DRIFT; SENSOR; RELIABILITY; TRANSDUCER; EXPERIENCE; DEVICE;
D O I
10.1186/s13054-015-1137-9
中图分类号
R4 [临床医学];
学科分类号
1002 ; 100602 ;
摘要
Introduction: Intracranial pressure (ICP) measurement is used to tailor interventions and to assist in formulating the prognosis for traumatic brain injury patients. Accurate data are therefore essential. The aim of this study was to verify the accuracy of ICP monitoring systems on the basis of a literature review. Methods: A PubMed search was conducted from 1982 to 2014, plus additional references from the selected papers. Accuracy was defined as the degree of correspondence between the pressure read by the catheter and a reference "real" ICP measurement. Studies comparing simultaneous readings from at least two catheters were included. Drift was defined as the loss of accuracy over the monitoring period. Meta-analyses of data from the studies were used to estimate the overall mean difference between simultaneous ICP measurements and their variability. Individual studies were weighted using both a fixed and a random effects model. Results: Of 163 articles screened, 83 compared two intracranial catheters: 64 reported accuracy and 37 drift (some reported both). Of these, 10 and 17, respectively, fulfilled the inclusion criteria for accuracy and zero drift analysis. The combined mean differences between probes were 1.5 mmHg (95 % confidence interval (CI) 0.7-2.3) with the random effects model and 1.6 mmHg (95 % CI 1.3-1.9) with the fixed effects model. The reported mean drift over a long observation period was 0.75 mmHg. No relation was found with the duration of monitoring or differences between various probes. Conclusions: This study confirms that the average error between ICP measures is clinically negligible. The random effects model, however, indicates that a high percentage of readings may vary over a wide range, with clinical implications both for future comparison studies and for daily care.
引用
收藏
页数:8
相关论文
共 33 条
[1]   ASSESSMENT OF ZERO DRIFT IN THE CODMAN INTRACRANIAL PRESSURE MONITOR: A STUDY FROM 2 NEUROINTENSIVE CARE UNITS [J].
Al-Tamimi, Yahia Z. ;
Helmy, Adel ;
Bavetta, Seb ;
Price, Stephen J. .
NEUROSURGERY, 2009, 64 (01) :94-98
[2]   Is intracranial pressure monitoring in the epidural space reliable?: Fact and fiction [J].
Antonia Poca, Maria ;
Sahuquillo, Juan ;
Topczewski, Thomaz ;
Jesus Penarrubia, Maria ;
Muns, Asuncion .
JOURNAL OF NEUROSURGERY, 2007, 106 (04) :548-556
[3]   Intracranial pressure and clinical status: assessment of two intracranial pressure transducers [J].
Banister, K ;
Chambers, IR ;
Siddique, MS ;
Fernandes, HM ;
Mendelow, AD .
PHYSIOLOGICAL MEASUREMENT, 2000, 21 (04) :473-479
[4]   Prospective study of zero drift in fiberoptic pressure monitors used in clinical practice [J].
Bavetta, S ;
Norris, JS ;
Wyatt, M ;
Sutcliffe, JC ;
Hamlyn, PJ .
JOURNAL OF NEUROSURGERY, 1997, 86 (06) :927-930
[5]   Is supratentorial pressure difference clinically relevant? Analysis of 55 consecutive cases by bilateral intracranial pressure monitoring [J].
Bekar, Ahmet ;
Taskapilioglu, Ozgur ;
Yilmazlar, Selcuk ;
Ender, Korfali ;
Aksoy, Kaya .
NEUROLOGICAL RESEARCH, 2008, 30 (05) :465-470
[6]  
Bratton S.L., 2007, J NEUROTRAUM, V24, pS45S54, DOI [10.1089/neu.2007.9989, DOI 10.1089/NEU.2007.9989]
[7]  
Chambers IR, 1998, ACT NEUR S, V71, P42
[8]   A CLINICAL-EVALUATION OF THE CAMINO SUBDURAL SCREW AND VENTRICULAR MONITORING KITS [J].
CHAMBERS, IR ;
MENDELOW, AD ;
SINAR, EJ ;
MODHA, P .
NEUROSURGERY, 1990, 26 (03) :421-423
[9]  
CHAMBERS IR, 1993, NEUROSURGERY, V33, P866
[10]   Clinical comparison of the Spiegelberg parenchymal transducer and ventricular fluid pressure [J].
Chambers, IR ;
Siddique, MS ;
Banister, K ;
Mendelow, AD .
JOURNAL OF NEUROLOGY NEUROSURGERY AND PSYCHIATRY, 2001, 71 (03) :383-385