A Review: Photonic Devices Used for Dosimetry in Medical Radiation

被引:51
作者
Damulira, Edrine [1 ]
Yusoff, Muhammad Nur Salihin [1 ]
Omar, Ahmad Fairuz [2 ]
Taib, Nur Hartini Mohd [3 ]
机构
[1] Univ Sains Malaysia, Sch Hlth Sci, Med Radiat Programme, Kubang Kerian 16150, Malaysia
[2] Univ Sains Malaysia, Sch Phys, Engn Phys Lab, George Town 11800, Malaysia
[3] Univ Sains Malaysia, Sch Med Sci, Dept Radiol, Kubang Kerian 16150, Malaysia
关键词
radiation-induced current; dosimetry; photodiodes; phototransistors (MOSFETs); photovoltaic sensor; CCD; CMOS; SILICON SOLAR-CELLS; CERENKOV RADIATION; MOSFET DETECTORS; PIN PHOTODIODE; THIN-FILMS; LUMINESCENCE; GAMMA;
D O I
10.3390/s19102226
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Numerous instruments such as ionization chambers, hand-held and pocket dosimeters of various types, film badges, thermoluminescent dosimeters (TLDs) and optically stimulated luminescence dosimeters (OSLDs) are used to measure and monitor radiation in medical applications. Of recent, photonic devices have also been adopted. This article evaluates recent research and advancements in the applications of photonic devices in medical radiation detection primarily focusing on four types; photodiodes - including light-emitting diodes (LEDs), phototransistorsincluding metal oxide semiconductor field effect transistors (MOSFETs), photovoltaic sensors/solar cells, and charge coupled devices/charge metal oxide semiconductors (CCD/CMOS) cameras. A comprehensive analysis of the operating principles and recent technologies of these devices is performed. Further, critical evaluation and comparison of their benefits and limitations as dosimeters is done based on the available studies. Common factors barring photonic devices from being used as radiation detectors are also discussed; with suggestions on possible solutions to overcome these barriers. Finally, the potentials of these devices and the challenges of realizing their applications as quintessential dosimeters are highlighted for future research and improvements.
引用
收藏
页数:28
相关论文
共 81 条
[31]   From photovoltaics to medical imaging: Applications of thin-film CdTe in x-ray detection [J].
Kang, J. ;
Parsai, E. I. ;
Albin, D. ;
Karpov, V. G. ;
Shvydka, Diana .
APPLIED PHYSICS LETTERS, 2008, 93 (22)
[32]  
Kester W., 1999, SENS TECHNOL HDB, V2, P31
[33]  
Knoll G., 2010, Radiation detection and measurement
[34]   Dosimetry of ionising radiation in modern radiation oncology [J].
Kron, Tomas ;
Lehmann, Joerg ;
Greer, Peter B. .
PHYSICS IN MEDICINE AND BIOLOGY, 2016, 61 (14) :R167-R205
[35]   A prototype scintillation dosimeter customized for small and dynamic megavoltage radiation fields [J].
Lambert, Jamil ;
Yin, Yongbai ;
McKenzie, David R. ;
Law, Susan H. ;
Ralston, Anna ;
Suchowerska, Natalka .
PHYSICS IN MEDICINE AND BIOLOGY, 2010, 55 (04) :1115-1126
[36]  
Lischka H., 1993, P 2 IEEE EUR C RADEC, DOI [10.1109/RADECS.1993.316536, DOI 10.1109/RADECS.1993.316536]
[37]  
Manufacturer S.M.S.L., 2018, DIFF LED LIGHT COB S
[38]  
Mehta S., 2015, Communications and Signal Processing (ICCSP), 2015 International Conference on, DOI 10.1109/ICCSP.2015.7322890
[39]   Cherenkov imaging for linac beam shape analysis as a remote electronic quality assessment verification tool [J].
Miao, Tianshun ;
Bruza, Petr ;
Pogue, Brian W. ;
Jermyn, Michael ;
Krishnaswamy, Venkataramanan ;
Ware, William ;
Rafie, Frank ;
Gladstone, David J. ;
Williams, Benjamin B. .
MEDICAL PHYSICS, 2019, 46 (02) :811-821
[40]  
Mobley J., 2003, BIOMED PHOTONICS HDB, V2, P1