Design of Anthropomorphic Textured Phantoms for CT Performance Evaluation

被引:12
作者
Solomon, Justin [1 ,3 ]
Bochud, Francois [4 ]
Samei, Ehsan [1 ,2 ,3 ]
机构
[1] Duke Univ, Med Ctr, Dept Radiol, Carl E Ravin Adv Imaging Labs, Durham, NC 27706 USA
[2] Duke Univ, Pratt Sch Engn, Dept Biomed Engn, Dept Elect & Comp Engn, Durham, NC 27706 USA
[3] Duke Univ, Grad Sch, Med Phys Grad Program, Durham, NC 27706 USA
[4] Univ Lausanne Hosp, Inst Radiat Phys, Lausanne, Switzerland
来源
MEDICAL IMAGING 2014: PHYSICS OF MEDICAL IMAGING | 2014年 / 9033卷
关键词
D O I
10.1117/12.2043555
中图分类号
O43 [光学];
学科分类号
070207 ; 0803 ;
摘要
Commercially available computed tomography (CT) technologies such as iterative reconstruction (IR) have the potential to enable reduced patient doses while maintaining diagnostic image quality. However, systematically determining safe dose reduction levels for IR algorithms is a challenging task due to their non-linear nature. Most attempts to evaluate IR algorithms rely on measurements made in uniform phantoms. Such measurements may overstate the dose reduction potential of IR because they don't account for the complex relationship between anatomical variability and image quality. The purpose of this study was to design anatomically informed textured phantoms for CT performance evaluation. Two phantoms were designed to represent lung and soft-tissue textures. The lung phantom includes intricate vessel-like structures along with embedded nodules (spherical, lobulated, and spiculated). The soft tissue phantom was designed based on a three-dimensional clustered lumpy background with included low-contrast lesions (spherical and anthropomorphic). The phantoms were built using rapid prototyping (3D printing) technology and imaged on a modern multi-slice clinical CT scanner to assess the noise performance of a commercial IR algorithm in the context of uniform and textured backgrounds. Fifty repeated acquisitions were acquired for each background type and noise was assessed by measuring pixel standard deviation, across the ensemble of repeated acquisitions. For pixels in uniform areas, the IR algorithm reduced noise magnitude (STD) by 60% (compared to FBP). However, for edge pixels, the noise magnitude in the IR images ranged from 20% higher to 40% lower compared to FBP. In all FBP images and in IR images of the uniform phantom, noise appeared to be globally non-stationary (i.e., spatially dependent) but locally stationary (within a reasonably small region of interest). In the IR images of the textured phantoms, the noise was globally and locally non-stationary.
引用
收藏
页数:11
相关论文
共 10 条
[1]   Statistical texture synthesis of mammographic images with clustered lumpy backgrounds [J].
Bochud, FO ;
Abbey, CK ;
Eckstein, MP .
OPTICS EXPRESS, 1999, 4 (01) :33-43
[2]  
Chen B., 2012, SPIE MED IMAGING
[3]   Correlation between model observer and human observer performance in CT imaging when lesion location is uncertain [J].
Leng, Shuai ;
Yu, Lifeng ;
Zhang, Yi ;
Carter, Rickey ;
Toledano, Alicia Y. ;
McCollough, Cynthia H. .
MEDICAL PHYSICS, 2013, 40 (08)
[4]   Achieving Routine Submillisievert CT Scanning: Report from the Summit on Management of Radiation Dose in CT [J].
McCollough, Cynthia H. ;
Chen, Guang Hong ;
Kalender, Willi ;
Leng, Shuai ;
Samei, Ehsan ;
Taguchi, Katsuyuki ;
Wang, Ge ;
Yu, Lifeng ;
Pettigrew, Roderic I. .
RADIOLOGY, 2012, 264 (02) :567-580
[5]   3D printing in X-ray and gamma-ray imaging: A novel method for fabricating high-density imaging apertures [J].
Miller, Brian W. ;
Moore, Jared W. ;
Barrett, Harrison H. ;
Frye, Teresa ;
Adler, Steven ;
Sery, Joe ;
Furenlid, Lars R. .
NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2011, 659 (01) :262-268
[6]   Three-dimensional phantoms for curvature correction in spatial frequency domain imaging [J].
Nguyen, Thu T. A. ;
Le, Hanh N. D. ;
Vo, Minh ;
Wang, Zhaoyang ;
Luu, Long ;
Ramella-Roman, Jessica C. .
BIOMEDICAL OPTICS EXPRESS, 2012, 3 (06) :1200-1214
[7]   3D printing based on imaging data: review of medical applications [J].
Rengier, F. ;
Mehndiratta, A. ;
von Tengg-Kobligk, H. ;
Zechmann, C. M. ;
Unterhinninghofen, R. ;
Kauczor, H. -U. ;
Giesel, F. L. .
INTERNATIONAL JOURNAL OF COMPUTER ASSISTED RADIOLOGY AND SURGERY, 2010, 5 (04) :335-341
[8]   Are Uniform Phantoms Sufficient to Characterize the Performance of Iterative Reconstruction in CT? [J].
Solomon, Justin ;
Samei, Ehsan .
MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
[9]   A methodology for image quality evaluation of advanced CT systems [J].
Wilson, Joshua M. ;
Christianson, Olav I. ;
Richard, Samuel ;
Samei, Ehsan .
MEDICAL PHYSICS, 2013, 40 (03)
[10]   Prediction of human observer performance in a 2-alternative forced choice low-contrast detection task using channelized Hotelling observer: Impact of radiation dose and reconstruction algorithms [J].
Yu, Lifeng ;
Leng, Shuai ;
Chen, Lingyun ;
Kofler, James M. ;
Carter, Rickey E. ;
McCollough, Cynthia H. .
MEDICAL PHYSICS, 2013, 40 (04)