Low temperature magnetic susceptometer based upon a dc superconducting quantum interference device

被引:4
作者
Read, Dan
Terry, Ian
Giblin, Sean R.
机构
[1] Univ Durham, Dept Phys, Durham DH1 3LE, England
[2] ISIS, Muon Facil, Chilton OX11 0QX, Oxon, England
关键词
D O I
10.1063/1.2360885
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
A low temperature magnetic susceptometer was constructed by combining an Oxford Instruments Heliox probe sorption pumped He-3 cryostat with a Quantum Design model 5000 dc superconducting quantum interference device sensor, giving an effective temperature range from 300 mK to 4.2 K. The smallest resolvable change in magnetic moment detected by this system was found to be Delta similar to 1 x 10(-12) J T-1 (1 x 10(-9) emu). (c) 2006 American Institute of Physics.
引用
收藏
页数:3
相关论文
共 11 条
[1]   LOW-TEMPERATURE MAGNETIC SPECTROSCOPY OF A DILUTE MAGNETIC SEMICONDUCTOR [J].
AWSCHALOM, DD ;
WARNOCK, J ;
VONMOLNAR, S .
PHYSICAL REVIEW LETTERS, 1987, 58 (08) :812-815
[2]  
Björnsson PG, 2001, REV SCI INSTRUM, V72, P4153, DOI 10.1063/1.1406931
[3]   Probing the magnetic ground state of the DX- center in Cd1-xMnxTe:In using muon spin relaxation -: art. no. 054421 [J].
Giblin, SR ;
Terry, I ;
Read, DE ;
Hillier, AD ;
Becla, P .
PHYSICAL REVIEW B, 2006, 73 (05)
[4]  
Kossut J., 1993, HDB MAGNETIC MAT, V7
[5]   Automated and versatile SQUID magnetometer for the measurement of materials properties at millikelvin temperatures [J].
Morello, A ;
Angenent, WGJ ;
Frossati, G ;
de Jongh, LJ .
REVIEW OF SCIENTIFIC INSTRUMENTS, 2005, 76 (02) :023902-1
[6]   SPIN-GLASS BEHAVIOR OF CD1-XMNX TE BELOW THE NEAREST-NEIGHBOR PERCOLATION LIMIT [J].
NOVAK, MA ;
SYMKO, OG ;
ZHENG, DJ ;
OSEROFF, S .
JOURNAL OF APPLIED PHYSICS, 1985, 57 (08) :3418-3420
[7]  
ROUMENIN CS, 1994, HDB SENSORS ACTUATOR
[8]   BAND TAILS AND THE INSULATOR-METAL TRANSITION IN THE PERSISTENT PHOTOCONDUCTOR CD1-XMNXTEIN [J].
TERRY, I ;
PENNEY, T ;
VONMOLNAR, S ;
RIGOTTY, JM ;
BECLA, P .
SOLID STATE COMMUNICATIONS, 1992, 84 (1-2) :235-240
[9]   Single nanoparticle measurement techniques [J].
Wernsdorfer, W ;
Mailly, D ;
Benoit, A .
JOURNAL OF APPLIED PHYSICS, 2000, 87 (09) :5094-5096
[10]   MAGNETIZATION OF BOUND MAGNETIC POLARONS - DIRECT DETERMINATION VIA PHOTOMEMORY EFFECT [J].
WOJTOWICZ, T ;
KOLESNIK, S ;
MIOTKOWSKI, I ;
FURDYNA, JK .
PHYSICAL REVIEW LETTERS, 1993, 70 (15) :2317-2320