Transpulmonary thermodilution cardiac output measurement using the axillary artery in critically ill patients

被引:28
作者
Segal, E [1 ]
Katzenelson, R [1 ]
Berkenstadt, H [1 ]
Perel, A [1 ]
机构
[1] Tel Aviv Univ, Sackler Sch Med, Gen Intens Care Unit, Sheba Med Ctr,Dept Anesthesiol & Intens Care, IL-69978 Tel Aviv, Israel
关键词
axillary artery; cardiac output; monitoring; pulmonary artery catheter; thermodilution;
D O I
10.1016/S0952-8180(02)00345-8
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Study Objective: To compare cardiac output (CO) as measured by the arterial thermodilution technique using only a central venous catheter and an arterial catheter inserted into the axillary artery, with conventional CO measurement with thermodilution using a pulmony artery (PA) catheter (PAC). Design: Prospective clinical study in which each patient served as his/her own control. Setting: General ICU of a large tertiary-care teaching hospital. Patients: 22 patients who required invasive hemodynamic monitoring in the ICU. Interventions and Measurements: CO measurements made using the PAC (COpa) were compared to bolus arterial thermodilution measurements (COax). The significance of acute changes in the continuous CO measurements during acute hemodynamic episodes was observed. Main Results : The correlation between the two techniques (COpa and COax) was R-2 = 0. 82. There was a tendency for 5% overestimation of COpa by the COax. The SEM% (SEM/average CO) for COax and GOpa was 2.6% and 3.2%, respectively. The bias between measurements was 0.27 +/- 0.67 L/min, and the limits of agreement (mean difference +/- 2 SD) from minus 1.07 L/min to 1.63 L/min. Conclusions: In, critically ill patients, in whom the measurement of CO is required, arterial thermodilution, using a central vein and the axillary artery is accurate and reproducible. (C) 2002 by Elsevier Science Inc.
引用
收藏
页码:210 / 213
页数:4
相关论文
共 10 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] Buhre G, 1999, J CARDIOTHOR VASC AN, V13, P437
  • [3] The effectiveness of right heart catheterization in the initial care of critically ill patients
    Connors, AF
    Speroff, T
    Dawson, NV
    Thomas, C
    Harrell, FE
    Wagner, D
    Desbiens, N
    Goldman, L
    Wu, AW
    Califf, RM
    Fulkerson, WJ
    Vidaillet, H
    Broste, S
    Bellamy, P
    Lynn, J
    Knaus, WA
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11): : 889 - 897
  • [4] Is it time to pull the pulmonary artery catheter?
    Dalen, JE
    Bone, RC
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1996, 276 (11): : 916 - 918
  • [5] Continuous cardiac output by femoral arterial thermodilution calibrated pulse contour analysis: Comparison with pulmonary arterial thermodilution
    Goedje, O
    Hoeke, K
    Lichtwarck-Aschoff, M
    Faltchauser, A
    Lamm, P
    Reichart, B
    [J]. CRITICAL CARE MEDICINE, 1999, 27 (11) : 2407 - 2412
  • [6] CANNULATION OF BIG ARTERIES IN CRITICALLY ILL PATIENTS
    GURMAN, GM
    KRIEMERMAN, S
    [J]. CRITICAL CARE MEDICINE, 1985, 13 (04) : 217 - 220
  • [7] A MULTICENTER STUDY OF PHYSICIANS KNOWLEDGE OF THE PULMONARY-ARTERY CATHETER
    IBERTI, TJ
    FISCHER, EP
    LEIBOWITZ, AB
    PANACEK, EA
    SILVERSTEIN, JH
    ALBERTSON, TE
    [J]. JAMA-JOURNAL OF THE AMERICAN MEDICAL ASSOCIATION, 1990, 264 (22): : 2928 - 2932
  • [8] Comparison of pulmonary artery and arterial thermodilution cardiac output in critically ill patients
    Sakka, SG
    Reinhart, K
    Meier-Hellmann, A
    [J]. INTENSIVE CARE MEDICINE, 1999, 25 (08) : 843 - 846
  • [9] STETZ CW, 1982, AM REV RESPIR DIS, V126, P1001
  • [10] Cardiac output measurement by transpulmonary indicator dilution technique - An alternative to the pulmonary artery catheter
    vonSpiegel, T
    Wietasch, G
    Bursch, J
    Hoeft, A
    [J]. ANAESTHESIST, 1996, 45 (11): : 1045 - 1050