Two-photon fluorescence imaging of subsurface tissue structures with volume holographic microscopy

被引:4
作者
Zhai, Xiaomin [1 ]
Vyas, Sunil [1 ]
Yeh, J. Andrew [2 ]
Luo, Yuan [1 ,3 ,4 ]
机构
[1] Natl Taiwan Univ, Inst Med Device & Imaging, Taipei, Taiwan
[2] Natl Tsing Hua Univ, Dept Power Mech Engn, Hsinchu, Taiwan
[3] Natl Taiwan Univ, Mol Imaging Ctr, Taipei, Taiwan
[4] Natl Taiwan Univ, YongLin Inst Hlth, Taipei, Taiwan
关键词
two photon microscopy; volume gratings; holography; imaging systems; microscopy; fluorescence microscopy; GRATINGS;
D O I
10.1117/1.JBO.25.12.123705
中图分类号
Q5 [生物化学];
学科分类号
071010 ; 081704 ;
摘要
Significance: Two-photon (2P) fluorescence imaging can provide background-free high-contrast images from the scattering tissues. However, obtaining a multiplane image is not straightforward. We present a two-photon volume holographic imaging (2P-VHI) system for multiplane imaging. Aim: Our goal was to design and implement a 2P-VHI system that can provide the high-contrast optically sectioned images at multiple planes. Approach: A 2P-VHI system is presented that incorporates angularly multiplexed volume holographic gratings and a femtosecond laser source for fluorescence excitation for multiplane imaging. A volume hologram with multiplexed gratings provides multifocal observation, whereas nonlinear excitation using a femtosecond laser helps in significantly enhancing both depth resolution and contrast of images. Results: Standard fluorescent beads are used to demonstrate the imaging performance of the 2P-VHI system. Two-depth resolved optical-sectioning images of fluorescently labeled thick mice intestine samples were obtained. In addition, the optical sectioning capability of our system is measured and compared with that of a conventional VHI system. Conclusions: Results demonstrated that 2P excitation in VHI systems provided the optical sectioning ability that helps in reducing background noise in the images. Integration of nonlinear fluorescence excitation in the VHI provides some unique advantages to the system and has potential to design multidepth optical sectioned spatial-spectral imaging systems. (C) The Authors. Published by SPIE under a Creative Commons Attribution 4.0 Unported License.
引用
收藏
页数:8
相关论文
共 19 条
[1]   Improvement of axial resolution and contrast in temporally focused widefield two-photon microscopy with structured light illumination [J].
Choi, Heejin ;
Yew, Elijah Y. S. ;
Hallacoglu, Bertan ;
Fantini, Sergio ;
Sheppard, Colin J. R. ;
So, Peter T. C. .
BIOMEDICAL OPTICS EXPRESS, 2013, 4 (07) :995-1005
[2]  
Coufal H., 2002, HOLOGRAPHIC DATA STO
[3]   2-PHOTON LASER SCANNING FLUORESCENCE MICROSCOPY [J].
DENK, W ;
STRICKLER, JH ;
WEBB, WW .
SCIENCE, 1990, 248 (4951) :73-76
[4]   Characterizing point spread functions of two-photon fluorescence microscopy in turbid medium [J].
Dong, CY ;
Koenig, K ;
So, P .
JOURNAL OF BIOMEDICAL OPTICS, 2003, 8 (03) :450-459
[5]  
Goodman J., 1996, Introduction to Fourier optics
[6]   Deep tissue two-photon microscopy [J].
Helmchen, F ;
Denk, W .
NATURE METHODS, 2005, 2 (12) :932-940
[7]   COUPLED WAVE THEORY FOR THICK HOLOGRAM GRATINGS [J].
KOGELNIK, H .
BELL SYSTEM TECHNICAL JOURNAL, 1969, 48 (09) :2909-+
[8]   Real-time spectral imaging in three spatial dimensions [J].
Liu, WH ;
Psaltis, D ;
Barbastathis, G .
OPTICS LETTERS, 2002, 27 (10) :854-856
[9]   Laser-induced fluorescence imaging of subsurface tissue structures with a volume holographic spatial-spectral imaging system [J].
Luo, Yuan ;
Gelsinger-Austin, Paul J. ;
Watson, Jonathan M. ;
Barbastathis, George ;
Barton, Jennifer K. ;
Kostuk, Raymond K. .
OPTICS LETTERS, 2008, 33 (18) :2098-2100
[10]   Optimization of multiplexed holographic gratings in PQ-PMMA for spectral-spatial imaging filters [J].
Luo, Yuan ;
Gelsinger, Paul J. ;
Barton, Jennifer K. ;
Barbastathis, George ;
Kostuk, Raymond K. .
OPTICS LETTERS, 2008, 33 (06) :566-568