Experimental determination of energy resolution and transmission characteristics of an electrostatic toroidal spectrometer adapted to a standard scanning electron microscope

被引:6
作者
Berger, D
Filippov, M
Niedrig, H
Rau, EI
Schlichting, F
机构
[1] Tech Univ Berlin, Inst Opt, Sekr P 1 1, D-10623 Berlin, Germany
[2] Moscow MV Lomonosov State Univ, Dept Phys, Moscow 119899, Russia
关键词
toroidal energy analyzer; BSE spectroscopy in SEM; spectrometer calibration; spectrometer characteristics; BSE spectra correction;
D O I
10.1016/S0368-2048(99)00036-5
中图分类号
O433 [光谱学];
学科分类号
0703 ; 070302 ;
摘要
The main characteristics of sectorial toroidal energy analyzer for a new electrostatic electron spectrometer adapted to a standard scanning electron microscope are defined and determined experimentally. These transfer characteristics, i.e. intensity-energy response and transmission functions, energy resolution and coupling constant, are needed for spectrometer calibration, registration and energy correction of measured backscattered electron spectra and microtomographic analysis of multilayered structures. The spectrometer response to a monoenergetic primary electron beam and to a continuous energy distribution is discussed. Detector response functions for energy independent and linear energy dependent detectors are considered. For aperture slits of the spectrometer which allow reasonable electron intensities at the detector an energy resolution of about 2.5% is obtained. (C) 1999 Elsevier Science B.V. All rights reserved.
引用
收藏
页码:119 / 127
页数:9
相关论文
共 27 条
  • [1] ARISTOV VV, 1995, PHYS STATUS SOLIDI A, V150, P675
  • [2] Bauer H.-D., 1979, Experimentelle Technik der Physik, V27, P331
  • [3] Berger D, 1998, ELECTRON MICROSCOPY 1998, VOL 4, P449
  • [4] NOVEL CHARGED-PARTICLE ANALYZER FOR MOMENTUM DETERMINATION IN THE MULTI-CHANNELING MODE .1. DESIGN ASPECTS AND ELECTRON-ION OPTICAL-PROPERTIES
    ENGELHARDT, HA
    BACK, W
    MENZEL, D
    LIEBL, H
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1981, 52 (06) : 835 - 839
  • [5] REDUCTION OF SPURIOUS BACKGROUND PEAKS IN ELECTRON SPECTROMETERS
    FROITZHEIM, H
    IBACH, H
    LEHWALD, S
    [J]. REVIEW OF SCIENTIFIC INSTRUMENTS, 1975, 46 (10) : 1325 - 1328
  • [6] ANGLE, ENERGY, AND TIME-OF-FLIGHT FOCUSING WITH POLOIDAL TOROID ELECTROSTATIC ANALYZERS
    GHIELMETTI, AG
    SHELLEY, EG
    [J]. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1990, 298 (1-3) : 181 - 188
  • [7] POTENTIAL DISTRIBUTION AND FOCUSING PROPERTIES OF TOROIDAL DEFLECTION PLATES - APPLICATION TO SIMULTANEOUS ENERGY AND ANGLE-RESOLVED CHARGED-PARTICLE SPECTROSCOPY
    HELLINGS, GJA
    OTTEVANGER, H
    KNIBBELER, CLC
    VANENGELSHOVEN, J
    BRONGERSMA, HH
    [J]. JOURNAL OF ELECTRON SPECTROSCOPY AND RELATED PHENOMENA, 1989, 49 (04) : 359 - 382
  • [8] AN EMPIRICAL ELECTRON SPECTROMETER TRANSMISSION FUNCTION FOR APPLICATIONS IN QUANTITATIVE XPS
    HEMMINGER, CS
    LAND, TA
    CHRISTIE, A
    HEMMINGER, JC
    [J]. SURFACE AND INTERFACE ANALYSIS, 1990, 15 (05) : 323 - 327
  • [9] HUBBEL HH, 1964, NUCL INSTRUM METHODS, V31, P331
  • [10] IBACH H, 1977, TOPICS CURRENT PHYSI, V4