Long-Term Performance of the Attain Model 4193 Left Ventricular Lead

被引:4
作者
Ben Johnson, W. [1 ]
Abraham, William T. [2 ]
Young, James B. [3 ]
Wheelan, Kevin [4 ]
Smith, Andrew L. [5 ]
Chang, Yanping [6 ]
Brinkman, Paula [6 ]
机构
[1] Iowa Heart Ctr, Des Moines, IA 50314 USA
[2] Ohio State Univ, Ctr Heart, Columbus, OH 43210 USA
[3] Cleveland Clin Fdn, Cleveland, OH 44195 USA
[4] Baylor Univ, Med Ctr, Dallas, TX USA
[5] Emory Univ, Crawford Long Hosp, Atlanta, GA 30365 USA
[6] Medtronic Inc, Minneapolis, MN USA
来源
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY | 2009年 / 32卷 / 09期
关键词
cardiac resynchronization therapy; long-term performance; left ventricular lead; CHRONIC HEART-FAILURE; CARDIAC-RESYNCHRONIZATION THERAPY; PACING LEADS; IMPLANTATION; EXPERIENCE; TRIAL;
D O I
10.1111/j.1540-8159.2009.02451.x
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Methods: The Attain Model 4193 LV lead (Medtronic, Inc, Minneapolis MN, USA) was successfully implanted in 1,070 patients with 286 patients completing 3 years of follow-up. Clinical data were collected at pre-implant, implant, and at 6-month intervals for 3 years. Results: Over 3 years, the mean chronic pacing threshold ranged from 1.9 V to 2.1 V, the mean R-wave sensing amplitudes ranged between 13.6 mV and 15.0 mV, and the mean pacing impedance ranged between 562 ohms and 590 ohms. Additionally, the observed freedom from first post-implant LV-lead-related complications was 90.4%. Of 1,070 total patients, 82 experienced 89 LV-lead-related adverse events requiring invasive interventions or resulting in the termination of the CRT therapy. The LV lead was repositioned in 31 patients, replaced in 21 patients, and explanted/capped in four patients. There were no deaths related to the LV lead during implantation or during the follow-up period. Conclusions: The data suggest that the 4193 LV lead is safe and effective over time. The LV lead electrical measurements remained stable through follow-up, demonstrating reliable long-term performance within the recommended value range at 36 months and had an acceptable complication rate. (PACE 2009; 32:1111-1116).
引用
收藏
页码:1111 / 1116
页数:6
相关论文
共 15 条
[1]   Cardiac resynchronization in chronic heart failure [J].
Abraham, WT ;
Fisher, WG ;
Smith, AL ;
Delurgio, DB ;
Leon, AR ;
Loh, E ;
Kocovic, DZ ;
Packer, M ;
Clavell, AL ;
Hayes, DL ;
Ellestad, M ;
Messenger, J ;
Trupp, RJ ;
Underwood, J ;
Pickering, F ;
Truex, C ;
McAtee, P .
NEW ENGLAND JOURNAL OF MEDICINE, 2002, 346 (24) :1845-1853
[2]   Six year experience of transvenous left ventricular lead implantation for permanent biventricular pacing in patients with advanced heart failure: technical aspects [J].
Alonso, C ;
Leclercq, C ;
d'Allonnes, FR ;
Pavin, D ;
Victor, F ;
Mabo, P ;
Daubert, JC .
HEART, 2001, 86 (04) :405-410
[3]   Cardiac-resynchronization therapy with or without an implantable defibrillator in advanced chronic heart failure [J].
Bristow, MR ;
Saxon, LA ;
Boehmer, J ;
Krueger, S ;
Kass, DA ;
De Marco, T ;
Carson, P ;
DiCarlo, L ;
DeMets, D ;
White, BG ;
DeVries, DW ;
Feldman, AM .
NEW ENGLAND JOURNAL OF MEDICINE, 2004, 350 (21) :2140-2150
[4]   Longer-term effects of cardiac resynchronization therapy on mortality in heart failure [the CArdiac REsynchronization-Heart Failure (CARE-HF) trial extension phase] [J].
Cleland, John G. F. ;
Daubert, Jean-Claude ;
Erdmann, Erland ;
Freemantle, Nick ;
Gras, Daniel ;
Kappenberger, Lukas ;
Tavazzi, Luigi .
EUROPEAN HEART JOURNAL, 2006, 27 (16) :1928-1932
[5]   Short and long-term single-centre experience with an S-shaped unipolar lead for left ventricular pacing [J].
Fatemi, M ;
Etienne, Y ;
Gilard, M ;
Mansourati, J ;
Blanc, JJ .
EUROPACE, 2003, 5 (02) :207-211
[6]  
Gregoratos G, 2002, J CARDIOVASC ELECTR, V13, P1183
[7]  
Hasdemir C, 2007, ANATOL J CARDIOL, V7, P53
[8]   Long-term performance of active-fixation pacing leads: A prospective study [J].
Kistler, PM ;
Liew, G ;
Mond, HG .
PACE-PACING AND CLINICAL ELECTROPHYSIOLOGY, 2006, 29 (03) :226-230
[9]   Safety of transvenous cardiac resynchronization system implantation in patients with chronic heart failure -: Combined results of over 2,000 patients from a multicenter study program [J].
León, AR ;
Abraham, WT ;
Curtis, AB ;
Daubert, JP ;
Fisher, WG ;
Gurley, J ;
Hayes, DL ;
Lieberman, R ;
Petersen-Stejskal, S ;
Wheelan, K .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (12) :2348-2356
[10]   Cardiac resynchronization with sequential biventricular pacing for the treatment of moderate-to-severe heart failure [J].
León, AR ;
Abraham, WT ;
Brozena, S ;
Daubert, JP ;
Fisher, WG ;
Gurley, JC ;
Liang, CS ;
Wong, G .
JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2005, 46 (12) :2298-2304