The Bimetric cementless total hip replacement: 7-18 year follow-up assessing the influence of acetabular design on survivorship

被引:2
作者
Russell, R. C.
Ghassemi, A. [2 ]
Dorrell, J. H. [1 ]
Powles, D. P. [1 ]
机构
[1] Lister Hosp, Stevenage, Herts, England
[2] Kings Coll Hosp London, London, England
关键词
FEMORAL COMPONENT; ARTHROPLASTY; PROSTHESIS; TITANIUM; WEAR; STEM;
D O I
10.1007/s00264-008-0592-1
中图分类号
R826.8 [整形外科学]; R782.2 [口腔颌面部整形外科学]; R726.2 [小儿整形外科学]; R62 [整形外科学(修复外科学)];
学科分类号
摘要
The purpose of this study was to evaluate the mid- to long-term survivorship of Bimetric cementless total hip replacement and assess how it is affected by the acetabular design. This was a retrospective analysis of 127 Bimetric cementless total hip replacements in 110 patients with a follow-up of 7-18 years. A single design stem and three different cementless metal-backed acetabular designs were used. Patients were assessed clinically using the Harris hip score and radiologically by independent review of current hip radiographs. There was only one case of aseptic loosening of the femoral stem. The earliest acetabular design showed a high failure rate whilst the latter two designs showed a 96% survivorship at a mean of 9.5 years. We conclude that a combination of the bimetric stem with either of the latter acetabular cup designs has a good mid- to long-term performance.
引用
收藏
页码:933 / 937
页数:5
相关论文
共 14 条
  • [1] Cardiac output during hemiarthroplasty of the hip - A prospective controlled trial of cemented and uncemented prostheses
    Clark, DI
    Ahmed, AB
    Baxendale, BR
    Moran, CG
    [J]. JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (03): : 414 - 418
  • [2] Hozack W, 1994, J Arthroplasty, V9, P489, DOI 10.1016/0883-5403(94)90094-9
  • [3] Heterotopic ossification after uncemented hydroxyapatite-coated primary total hip arthroplasty
    Kasetti, RJ
    Shetty, AA
    Rand, C
    [J]. JOURNAL OF ARTHROPLASTY, 2001, 16 (08) : 1038 - 1042
  • [4] Primary total hip arthroplasty with a cementless porous-coated anatomic total hip prosthesis - 10-to 12-year results of prospective and consecutive series
    Kim, YH
    Kim, JS
    Cho, SH
    [J]. JOURNAL OF ARTHROPLASTY, 1999, 14 (05) : 538 - 548
  • [5] Clinical and radiographic outcome of a cementless, titanium, plasma spray-coated total hip arthroplasty femoral component - Justification for continuance of use
    Mallory, TH
    Head, WC
    Lombardi, AV
    Emerson, RH
    Eberle, RW
    Mitchell, MB
    [J]. JOURNAL OF ARTHROPLASTY, 1996, 11 (06) : 653 - 660
  • [6] Cementless titanium tapered-wedge femoral stem - 10-to 15-year follow-up
    Marshall, AD
    Mokris, JG
    Reitman, RD
    Dandar, A
    Mauerhan, DR
    [J]. JOURNAL OF ARTHROPLASTY, 2004, 19 (05) : 546 - 552
  • [7] Mella-Sousa M, 2001, ORTHOPEDICS, V24, P1053
  • [8] Moroni A, 2000, ANN CHIR GYNAECOL FE, V89, P113
  • [9] Primary total hip arthroplasty with an uncemented femoral component - A long-term study of the Taperloc stem
    Parvizi, J
    Keisu, KS
    Hozack, WJ
    Sharkey, PF
    Rothman, RH
    [J]. JOURNAL OF ARTHROPLASTY, 2004, 19 (02) : 151 - 156
  • [10] Polyethylene wear and periprosthetic osteolysis in metal-backed acetabular components with cylindrical liners
    Perez, RE
    Rodriguez, JA
    Deshmukh, RG
    Ranawat, CS
    [J]. JOURNAL OF ARTHROPLASTY, 1998, 13 (01) : 1 - 7