EFFECT OF WETNESS LEVEL ON THE SUITABILITY OF WET GAUZE AS A SUBSTITUTE FOR SUPERFLAB® AS A BOLUS MATERIAL FOR USE WITH 6 MV PHOTONS

被引:23
作者
Benoit, Jerome [1 ]
Pruitt, Amy F. [1 ]
Thrall, Donald E. [1 ]
机构
[1] N Carolina State Univ, Coll Vet Med, Dept Mol Biomed Sci, Raleigh, NC 27606 USA
关键词
air gap; bolus; radiation therapy; Superflab((R)); surface dose;
D O I
10.1111/j.1740-8261.2009.01573.x
中图分类号
S85 [动物医学(兽医学)];
学科分类号
0906 ;
摘要
Despite the availability of commercial tissue equivalent bolus material, wet gauze has an application in radiation therapy to provide superior conformance to irregular contours. Wet gauze bolus has the potential to reduce air gaps between the bolus and surface, which could decrease surface dose if sufficiently large to disrupt electronic equilibrium. Wet gauze bolus is often fabricated and wetness judged qualitatively. We assessed the effect of specific gauze wetness levels, quantified in terms of physical density, at various field sizes with respect to their effectiveness as bolus material compared with Superflab((R)). For large fields, > 7 x 7 cm(2) in this study, wet gauze sponges with a physical density of 1.02 g/cm(3) performed essentially identical to Superflab((R)); at a smaller field size the wet gauze was slightly less effective, likely due to the heterogeneity of the gauze-water matrix. Gauze that was wetter, with a physical density of 1.2 g/cm(3), or less wet, with a physical density of 0.75 g/cm(3) was not as effective either due to enhanced photon absorption in the wetter sponges, or less effective establishment of electronic equilibrium in the less wet sponges. The presence of an air gap under Superflab((R)) led to reduced surface dose, especially for small fields and large air gaps. Thus, if Superflab((R)) use leads to poor contact with the skin, wet gauze having a physical density of 1.02 g/cm(3) can be used as a substitute. Judging the water content of wet gauze subjectively is not acceptable as over- or under wetness can lead to decreased effectiveness of the bolus material.
引用
收藏
页码:555 / 559
页数:5
相关论文
共 6 条
[1]  
Bentel Gunilla C., 1996, RAD THERAPY PLANNING, P239
[2]   Effects on skin dose from unwanted air gaps under bolus in photon beam radiotherapy [J].
Butson, MJ ;
Cheung, T ;
Yu, P ;
Metcalfe, P .
RADIATION MEASUREMENTS, 2000, 32 (03) :201-204
[3]  
KHAN MK, 2003, PHYS RAD THERAPY, P259
[4]  
*NUCL ASS, 2005, SUP PLAST BOL MAT
[5]  
WAKELYN PJ, 2006, COTTON FIBER CHEM TE, P25
[6]   Play-Doh® and water-soaked gauze sponges as alternative bolus material for cobalt-60 teletherapy [J].
Walker, M ;
Cohen, N ;
Menchaca, D .
VETERINARY RADIOLOGY & ULTRASOUND, 2005, 46 (02) :179-181