Dosimetric characteristics of a newly designed grid block for megavoltage photon radiation and its therapeutic advantage using a linear quadratic model

被引:44
作者
Meigooni, Ali S. [1 ]
Dou, Kai [1 ]
Meigooni, Navid J. [1 ]
Gnaster, Michael [1 ]
Awan, Shahid [1 ]
Dini, Sharifeh [1 ]
Johnson, Ellis L. [1 ]
机构
[1] Univ Kentucky, Albert B Chandler Med Ctr, Dept Radiat Med, Lexington, KY 40535 USA
关键词
grid therapy; megavoltage photon radiation; dosimetry; cell survival; therapeutic advantace;
D O I
10.1118/1.2241998
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Grid radiation therapy with megavoltage x-ray beam has been proven to be an effective technique for management of large, bulky malignant tumors. The clinical advantage of GRID therapy, combined with conventional radiation therapy, has been demonstrated using a prototype GRID block [Mohiuddin, Curtis, Grizos, and Komarnicky, Cancer 66, 114-118 (1990)]. Recently, a new GRID block design with improved dosimetric properties has become commercially available from Radiation Product Design, Inc. (Albertive, MN). This GRID collimator consists of an array of focused apertures in a cerrobend block arranged in a hexagonal pattern having a circular cross-section with a diameter and center-to-center spacing of 14.3 and 21.1 mm, respectively, in the plane of isocenter. In this project, dosimetric characteristics of the newly redesigned GRID block have been investigated for a Varian 2 1 EX linear accelerator (Varian Associates, Palo Alto, CA). These determinations were performed using radiographic films, thermoluminescent dosimeters in Solid Water (TM) phantom materials, and an ionization chamber in water. The output factor, percentage depth dose, beam profiles, and isodose distributions of the GRID radiation as a function of field size and beam energy have been measured using both 6 and 18 MV x-ray beams. In addition, the therapeutic advantage obtained from this treatment modality with the new GRID block design for a high, single fraction of dose has been calculated using the linear quadratic model with alpha/beta ratios for typical tumor and normal cells. These biological characteristics of the new GRID block design will also be presented. (c) 2006 American Association of Physicists in Medicine.
引用
收藏
页码:3165 / 3173
页数:9
相关论文
共 21 条
[1]   DOSE FRACTIONATION, DOSE-RATE AND ISO-EFFECT RELATIONSHIPS FOR NORMAL TISSUE RESPONSES [J].
BARENDSEN, GW .
INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 1982, 8 (11) :1981-1997
[2]   Dosimetric characteristics of a new 125I brachytherapy source [J].
Gearheart, DM ;
Drogin, A ;
Sowards, K ;
Meigooni, AS ;
Ibbott, GS .
MEDICAL PHYSICS, 2000, 27 (10) :2278-2285
[3]   External audits of electron beams using mailed TLD dosimetry: preliminary results [J].
Gomola, I ;
Van Dam, J ;
Isern-Verdum, J ;
Verstraete, J ;
Reymen, R ;
Dutreix, A ;
Davis, B ;
Huyskens, D .
RADIOTHERAPY AND ONCOLOGY, 2001, 58 (02) :163-168
[4]   Monte Carlo and TLD dosimetry of an 192Ir high dose-rate brachytherapy source [J].
Karaiskos, P ;
Angelopoulos, A ;
Sakelliou, L ;
Sandilos, P ;
Antypas, C ;
Vlachos, L ;
Koutsouveli, E .
MEDICAL PHYSICS, 1998, 25 (10) :1975-1984
[5]  
LEE DE, 1946, ACTIONS RAD LIVING C
[6]  
Meigooni A S, 2002, Med Dosim, V27, P37, DOI 10.1016/S0958-3947(02)00086-9
[7]   INSTRUMENTATION AND DOSIMETER-SIZE ARTIFACTS IN QUANTITATIVE THERMOLUMINESCENCE DOSIMETRY OF LOW-DOSE FIELDS [J].
MEIGOONI, AS ;
MISHRA, V ;
PANTH, H ;
WILLIAMSON, J .
MEDICAL PHYSICS, 1995, 22 (05) :555-561
[8]   Dosimetric characteristics of the InterSource103 palladium brachytherapy source [J].
Meigooni, AS ;
Sowards, K ;
Soldano, M .
MEDICAL PHYSICS, 2000, 27 (05) :1093-1100
[9]  
MOHIUDDIN M, 1990, CANCER, V66, P114, DOI 10.1002/1097-0142(19900701)66:1<114::AID-CNCR2820660121>3.0.CO
[10]  
2-L