Performance of a new laser fluorescence device for the detection of occlusal caries in vitro

被引:113
作者
Lussi, A.
Hellwig, E.
机构
[1] Univ Bern, Sch Dent Med, Dept Operat Prevent & Paediat Dent, CH-3012 Bern, Switzerland
[2] Univ Clin Freiburg, Dept Operat Dent & Periodontol, Freiburg, Germany
[3] Sch Dent, Freiburg, Germany
关键词
caries detection; occlusal caries; laser fluorescence; DIAGNOdent; DIAGNOdent pen;
D O I
10.1016/j.jdent.2005.11.002
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
The new device DIAGNOdent pen based on red laser light induced fluorescence was introduced for the detection of approximal and occlusal caries. The aim of this study was to test its performance on occlusal surfaces. The new device comes with two different sapphire fibre tips: a cylindrical tip and a conical tip. The two new sapphire fibre tips were used and compared with the tip currently available with DIAGNOdent (DD). Methods: The teeth were selected from a pool of extracted permanent human molars, which were stored frozen at -20 degrees C, until use. Prior to being measured the teeth were defrosted and cleaned. One hundred and nineteen teeth were selected and measured with the old tip and with the two new tips of the new device by two independent assessments. The teeth were histologically prepared and assessed for caries extension. Results: Specificity values for D-1, D-2 and D-3 ranged between 0.69 and 0.89, sensitivity between 0.78 and 0.96. There were no statistically significant differences obtained between the two tips of the new and the one tip of the old device (p > 0.05). Intra-examiner reliability with kappa values of > 0.83 was high. Conclusions: in this study, the new laser fluorescence device performed on occlusal surfaces as well as the available device. (c) 2005 Elsevier Ltd. All rights reserved.
引用
收藏
页码:467 / 471
页数:5
相关论文
共 24 条
[1]  
[Anonymous], 2002, J DENT CHILD
[2]   Clinical study of the use of the laser fluorescence device DIAGNOdent for detection of occlusal caries in children [J].
Anttonen, V ;
Seppä, L ;
Hausen, H .
CARIES RESEARCH, 2003, 37 (01) :17-23
[3]   A systematic review of the performance of a laser fluorescence device for detecting caries [J].
Bader, JD ;
Shugars, DA .
JOURNAL OF THE AMERICAN DENTAL ASSOCIATION, 2004, 135 (10) :1413-1426
[4]   A systematic review of the performance of methods for identifying carious lesions [J].
Bader, JD ;
Shugars, DA ;
Bonito, AJ .
JOURNAL OF PUBLIC HEALTH DENTISTRY, 2002, 62 (04) :201-213
[5]   STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT [J].
BLAND, JM ;
ALTMAN, DG .
LANCET, 1986, 1 (8476) :307-310
[6]   An in vitro comparison of a combined FOTI/visual examination of occlusal caries with other caries diagnostic methods and the effect of stain on their diagnostic performance [J].
Côrtes, DF ;
Ellwood, RP ;
Ekstrand, KR .
CARIES RESEARCH, 2003, 37 (01) :8-16
[7]  
Downer M C, 1989, Int Dent J, V39, P241
[8]  
Ekstrand K R, 2004, J Dent Res, V83 Spec No C, pC67
[9]  
Fleiss IL, 1981, STAT METHODS RATES P, P212
[10]  
FRANCESCUT P, IN PRESS CARIES RES