Improving the sensitivity of a multi-collector inductively coupled plasma mass spectrometer via expansion-chamber pressure reduction

被引:6
作者
Yuan, Honglin [1 ]
Bao, Zhian [1 ]
Chen, Kaiyun [1 ]
Zong, Chunlei [1 ]
Chen, Lu [1 ]
Zhang, Tao [2 ]
机构
[1] Northwest Univ, Dept Geol, Collaborat Innovat Ctr Continental Tecton, State Key Lab Continental Dynam, Xian 710069, Shaanxi, Peoples R China
[2] Northwest Univ, Sch Informat Sci & Technol, Xian 710069, Shaanxi, Peoples R China
基金
中国国家自然科学基金;
关键词
MC-ICP-MS; MAGNESIUM ISOTOPIC COMPOSITION; PRECISE DETERMINATION; ENHANCED SENSITIVITY; GEOLOGICAL-MATERIALS; HF; SR; NITROGEN; RATIOS; ND;
D O I
10.1039/c8ja00448j
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
In this work, we improved the sensitivity of a multi-collector-inductively coupled plasma mass spectrometry system by increasing the vacuum of the expansion chamber using a high-speed EH mechanical booster pump (EH500, 500 m(3) h(-1), BOC Edwards) in conjunction with a standard E2M80 pump (80 m(3) h(-1), BOC Edwards). The vacuum was improved from 1.0 mbar to less than 0.5 mbar. The sample ion yield for heavy isotopes, such as those of Pb, was improved from 0.047% to 0.072% in the wet-plasma mode (sample uptake rate, 113 mu L min(-1)), and from 1.35% to 1.84% in the dry-plasma mode (sample uptake rate 105 mu L min(-1)); the variation in the background intensity was less than 1.3 mV for all the Faraday cups. Thus, the sample ion yield in the wet-plasma mode was enhanced more effectively than that in the dry-plasma mode owing to a slightly larger orifice in the wet cone setup. The accuracies for all the isotopes tested, including those of non-traditional metal ones, in both modes were identical to previously published data within an error of 2 s. These results indicate that improving the vacuum at the interface has no effect on the accuracy of isotope measurements and is an effective method for improving the detectability of isotopic compositions for bulk solution analysis with low element contents and the spatial resolution for measurements performed using the laser-ablation sampling technique. This work could provide an alternative for laboratories with a similar device, when tackling samples with low contents.
引用
收藏
页码:1011 / 1017
页数:7
相关论文
共 44 条
  • [1] δ30Si and δ29Si determinations on USGS BHVO-1 and BHVO-2 reference materials with a new configuration on a nu plasma multi-collector ICP-MS
    Abraham, Kathrin
    Opfergelt, Sophie
    Fripiat, Francois
    Cavagna, Anne-Julie
    de Jong, Jeroen T. M.
    Foley, Stephen F.
    Andre, Luc
    Cardinal, Damien
    [J]. GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2008, 32 (02) : 193 - 202
  • [2] Pb isotopic analysis of standards and samples using a 207Pb-204Pb double spike and thallium to correct for mass bias with a double-focusing MC-ICP-MS
    Baker, J
    Peate, D
    Waight, T
    Meyzen, C
    [J]. CHEMICAL GEOLOGY, 2004, 211 (3-4) : 275 - 303
  • [3] Natural variations of lithium isotopes in a mammalian model
    Balter, Vincent
    Vigier, Nathalie
    [J]. METALLOMICS, 2014, 6 (03) : 582 - 586
  • [4] Determination of Hf–Sr–Nd isotopic ratios by MC-ICP-MS using rapid acid digestion after flux-free fusion in geological materials
    Bao Z.
    Zong C.
    Fang L.
    Yuan H.
    Chen K.
    Dai M.
    [J]. Acta Geochimica, 2018, 37 (2) : 244 - 256
  • [5] Magnesium Isotope Compositions of Natural Reference Materials
    Bolou-Bi, Emile B.
    Vigier, Nathalie
    Brenot, Agnes
    Poszwa, Anne
    [J]. GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2009, 33 (01) : 95 - 109
  • [6] Fractionation of Sr and Hf isotopes by mineral sorting in Cascadia Basin terrigenous sediments
    Carpentier, Marion
    Weis, Dominique
    Chauvel, Catherine
    [J]. CHEMICAL GEOLOGY, 2014, 382 : 67 - 82
  • [7] Precise and Accurate In Situ Determination of Lead Isotope Ratios in NIST, USGS, MPI-DING and CGSG Glass Reference Materials using Femtosecond Laser Ablation MC-ICP-MS
    Chen Kaiyun
    Yuan Honglin
    Bao Zhian
    Zong Chunlei
    Dai Mengning
    [J]. GEOSTANDARDS AND GEOANALYTICAL RESEARCH, 2014, 38 (01) : 5 - 21
  • [8] Enhanced sensitivity in laser ablation multi-collector inductively coupled plasma mass spectrometry
    Cottle, John M.
    Burrows, Andrew J.
    Kylander-Clark, Andrew
    Freedman, Philip A.
    Cohen, Roy S.
    [J]. JOURNAL OF ANALYTICAL ATOMIC SPECTROMETRY, 2013, 28 (11) : 1700 - 1706
  • [9] Doubling Sensitivity in Multicollector ICPMS Using High-Efficiency, Rapid Response Laser Ablation Technology
    Craig, Grant
    Managh, Amy J.
    Stremtan, Ciprian
    Lloyd, Nicholas S.
    Horstwood, Matthew S. A.
    [J]. ANALYTICAL CHEMISTRY, 2018, 90 (19) : 11564 - 11571
  • [10] In situ sulfur isotopes (δ34S and δ33S) analyses in sulfides and elemental sulfur using high sensitivity cones combined with the addition of nitrogen by laser ablation MC-ICP-MS
    Fu, Jiali
    Hu, Zhaochu
    Zhang, Wen
    Yang, Lu
    Liu, Yongsheng
    Li, Ming
    Zong, Keqing
    Gao, Shan
    Hu, Shenghong
    [J]. ANALYTICA CHIMICA ACTA, 2016, 911 : 14 - 26