Effect of Change in Portal Venous Blood Flow Rates on the Performance of a 2.45-GHz Microwave Ablation Device

被引:8
作者
Dodd, Gerald D., III [1 ]
Kreidler, Sarah M. [1 ]
Lanctot, Anthony C. [1 ]
Glueck, Deborah H. [2 ]
机构
[1] Univ Colorado, Sch Med, Dept Radiol, 12401 E 17th Ave,Mail Stop L954,POB 6510, Aurora, CO 80045 USA
[2] Univ Colorado, Sch Publ Hlth, Aurora, CO 80045 USA
基金
美国国家科学基金会;
关键词
BOVINE LIVER MODEL; HEPATIC RADIOFREQUENCY LESIONS; VIVO PORCINE LIVER; IN-VIVO; TISSUE ABLATION; VASCULAR OCCLUSION; HEATING PATTERNS; TUMOR ABLATION; SIZE; PERFUSION;
D O I
10.1148/radiol.2015150102
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To investigate the effect of change in portal venous blood flow rates on the size and shape of ablations created by a 2.45-GHz microwave ablation device. Materials and Methods: This study was exempt from review by the institutional animal care and use committee. An in vitro bovine liver model perfused with autologous blood via the portal vein at five flow rates (60, 70, 80, 90, and 100 mL/min per 100 g of liver) was used to evaluate the effect of change in flow rates on the size and shape of coagulation created by a 2.45-GHz, 140-W microwave ablation device operated for 5 and 10 minutes. Three ablations per ablation time were conducted in each of 10 livers, with two livers perfused at each flow rate. Short-and long-axis diameters were measured from gross specimens, and volume and sphericity index were calculated. General linear mixed models that accounted for correlations within the liver were used to evaluate the effects of lobe, flow, and ablation time on size and sphericity index of ablations. Results: Flow did not have a significant effect on the size or shape of coagulation created at 5 or 10 minutes (P>.05 for all tests). The mean short-and long-axis diameters and volume were 3.2 cm (95% confidence interval [CI]: 3.1, 3.3), 5.6 cm (95% CI: 5.4, 5.8), and 30.2 cm(3) (95% CI: 28.4, 32.1) for the 5-minute ablations and 3.8 cm (95% CI: 3.7, 3.9), 6.5 cm (95% CI: 6.3, 6.7), and 49.3 cm3 (95% CI: 47.5, 51.2), for the 10-minute ablations, respectively. The mean sphericity index for both 5- and 10-minute ablations was 34.4% (95% CI: 32%, 36.7%). Conclusion: Change in portal venous blood flow rates did not have an effect on the size and shape of ablations created by a 2.45GHz microwave ablation device. (C) RSNA, 2015
引用
收藏
页码:727 / 732
页数:6
相关论文
共 36 条
[1]   Computer modeling of the combined effects of perfusion, electrical conductivity, and thermal conductivity on tissue heating patterns in radiofrequency tumor ablation [J].
Ahmed, Muneeb ;
Liu, Zhengjun ;
Humphries, Stanley ;
Goldberg, S. Nahum .
INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2008, 24 (07) :577-588
[2]  
[Anonymous], 1996, COMPILATION DIELECTR
[3]  
Awad Michael M, 2007, HPB (Oxford), V9, P357, DOI 10.1080/13651820701646222
[4]   Effects of vascular perfusion on coagulation size in radiofrequency ablation of ex vivo perfused bovine livers [J].
Bitsch, RG ;
Düx, M ;
Helmberger, T ;
Lubienski, A .
INVESTIGATIVE RADIOLOGY, 2006, 41 (04) :422-427
[5]   Microwave ablation with multiple simultaneously powered small-gauge triaxial antennas: Results from an in vivo swine liver model [J].
Brace, Christopher L. ;
Laeseke, Paul F. ;
Sampson, Lisa A. ;
Frey, Tina M. ;
van der Weide, Daniel W. ;
Lee, Fred T., Jr. .
RADIOLOGY, 2007, 244 (01) :151-156
[6]   Microwave ablation with a single small-gauge triaxial antenna: In vivo porcine liver model [J].
Brace, Christopher L. ;
Laeseke, Paul F. ;
Sampson, Lisa A. ;
Frey, Tina M. ;
van der Weide, Daniel W. ;
Lee, Fred T., Jr. .
RADIOLOGY, 2007, 242 (02) :435-440
[7]   Tissue Contraction Caused by Radiofrequency and Microwave Ablation: A Laboratory Study in Liver and Lung [J].
Brace, Christopher L. ;
Diaz, Teresa A. ;
Hinshaw, J. Louis ;
Lee, Fred T., Jr. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2010, 21 (08) :1280-1286
[8]   THE ESTIMATION OF HEPATIC BLOOD FLOW IN MAN [J].
BRADLEY, SE ;
INGELFINGER, FJ ;
BRADLEY, GP ;
CURRY, JJ .
JOURNAL OF CLINICAL INVESTIGATION, 1945, 24 (06) :890-897
[9]   3T Magnetic Resonance Imaging Accurately Depicts Radiofrequency Ablation Zones in a Blood-perfused Bovine Liver Model [J].
Burns, Stephanie K. ;
Dodd, Gerald D., III ;
McManus, Linda M. ;
Cardan, Rex A. ;
Peng, Qi ;
Orsi, Michael D. ;
Head, Hayden W. ;
Blakemore, Dana L. ;
Blume, Jeffrey D. ;
Fullerton, Gary D. ;
Green, Tyler J. .
JOURNAL OF VASCULAR AND INTERVENTIONAL RADIOLOGY, 2012, 23 (06) :801-808
[10]   CT versus sonography for monitoring radiofrequency ablation in a porcine liver [J].
Cha, CH ;
Lee, FT ;
Gurney, JM ;
Markhardt, BK ;
Warner, TF ;
Kelcz, F ;
Mahvi, DM .
AMERICAN JOURNAL OF ROENTGENOLOGY, 2000, 175 (03) :705-711