Effects of one-plane and two-plane external fixation on sheep osteotomy healing and complications

被引:31
作者
Krischak, GD
Janousek, A
Wolf, S
Augat, P
Kinzl, L
Claes, LE
机构
[1] Univ Ulm, Dept Orthopaed Surg, D-89075 Ulm, Germany
[2] Lorenz Boehler Trauma Ctr, Vienna, Austria
[3] Univ Ulm, Dept Orthopaed Res & Biomech, Ulm, Germany
关键词
external fixation; rigidity; fixation stiffness; biomechanics; fracture healing; osteogenesis;
D O I
10.1016/S0268-0033(02)00039-6
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Objective. Comparison of one-plane and two-plane external fixation in terms of successful healing, incidence of complications, and biomechanical stability in a sheep model. Background. Rigid fixation is preferred in open and comminuted fractures with a reduced blood supply, preventing infection and healing delay, but more often a flexible device is recommended even in unfavourable healing conditions, Methods. The left tibiae of fifteen sheep were osteotomized and laterally fixed with a four-screw unilateral fixator frame (axial stiffness 183 N/mm) to a 3 mm gap size. In 9 of 15 sheep, an additional four-screw unilateral external fixator was anterolaterally attached (total axial stiffness of both frames 388 N/mm). After sacrificing quality of osteotomy healing was assessed by mechanical and radiological evaluations. Osteogenesis was measured using fluorescence microscopy. Results. Two distal fractures through the pin-tracks, three non-unions and four deep infections occurred after two-plane fixation, These failures excluded, osteotomy healing showed inferior results after two-plane fixation with reduced callus formation, bone mineral content, and bending stiffness amounts, respectively. Osteogenesis was halved following two-plane fixation in the remaining sheep. Conclusions. Two-plane fixation was not sufficient to reach successful osteotomy healing in our study. While higher rigidity was expected to prevent complications, healing in this group might have been disturbed by reduced blood supply. The optimal stabilisation for a given fracture depends on many factors, including the biomechanical and biological environment.
引用
收藏
页码:470 / 476
页数:7
相关论文
共 32 条
[1]   Interspecies differences in bone composition, density, and quality:: Potential implications for in vivo bone research [J].
Aerssens, J ;
Boonen, S ;
Lowet, G ;
Dequeker, J .
ENDOCRINOLOGY, 1998, 139 (02) :663-670
[2]   Fracture healing in biological plate osteosynthesis [J].
Baumgaertel, F ;
Buhl, M ;
Rahn, BA .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1998, 29 :3-6
[3]  
BEHRENS F, 1989, CLIN ORTHOP RELAT R, P48
[4]   Treatment of open fractures of the shaft of the tibia - A systematic overview and meta-analysis [J].
Bhandari, M ;
Guyatt, GH ;
Swiontkowski, MF ;
Schemitsch, EH .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2001, 83B (01) :62-68
[5]   Influence of size and stability of the osteotomy gap on the success of fracture healing [J].
Claes, L ;
Augat, P ;
Suger, G ;
Wilke, HJ .
JOURNAL OF ORTHOPAEDIC RESEARCH, 1997, 15 (04) :577-584
[6]   Fixation technique influences osteogenesis of comminuted fractures [J].
Claes, L ;
Heitemeyer, U ;
Krischak, G ;
Braun, H ;
Hierholzer, G .
CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 1999, (365) :221-229
[7]  
CLAES L, 1991, CHIRURG, V62, P354
[8]   EFFECT OF DYNAMIZATION ON GAP HEALING OF DIAPHYSEAL FRACTURES UNDER EXTERNAL FIXATION [J].
CLAES, LE ;
WILKE, HJ ;
AUGAT, P ;
RUBENACKER, S ;
MARGEVICIUS, KJ .
CLINICAL BIOMECHANICS, 1995, 10 (05) :227-234
[9]  
COURTBROWN CM, 1985, CLIN ORTHOP RELAT R, P278
[10]   EXTERNAL FIXATION FOR TYPE-III OPEN TIBIAL FRACTURES [J].
COURTBROWN, CM ;
WHEELWRIGHT, EF ;
CHRISTIE, J ;
MCQUEEN, MM .
JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 1990, 72 (05) :801-804