Real-Time Flow MRI of the Aorta at a Resolution of 40 msec

被引:37
作者
Joseph, Arun [1 ]
Kowallick, Johannes T. [2 ]
Merboldt, Klaus-Dietmar [1 ]
Voit, Dirk [1 ]
Schaetz, Sebastian [1 ]
Zhang, Shuo [1 ]
Sohns, Jan M. [2 ]
Lotz, Joachim [2 ]
Frahm, Jens [1 ]
机构
[1] Max Planck Inst Biophys Chem, Biomed NMR Forsch GmbH, D-37070 Gottingen, Germany
[2] Univ Med, Inst Diagnost & Interventionelle Radiol, Gottingen, Germany
关键词
real-time MRI; aorta; blood flow; stroke volume; CARDIOVASCULAR MAGNETIC-RESONANCE; PHASE-CONTRAST FLOW; QUANTIFICATION; RECONSTRUCTION; VALIDATION; ACCURACY;
D O I
10.1002/jmri.24328
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To evaluate a novel real-time phase-contrast magnetic resonance imaging (MRI) technique for the assessment of through-plane flow in the ascending aorta. Materials and Methods: Real-time MRI was based on a radial fast low-angle shot (FLASH) sequence with about 30-fold undersampling and image reconstruction by regularized nonlinear inversion. Phase-contrast maps were obtained from two (interleaved or sequential) acquisitions with and without a bipolar velocity-encoding gradient. Blood flow in the ascending aorta was studied in 10 healthy volunteers at 3 T by both real-time MRI (15 sec during free breathing) and electrocardiogram (ECG)-synchronized cine MRI (with and without breath holding). Flow velocities and stroke volumes were evaluated using standard postprocessing software. Results: The total acquisition time for a pair of phase-contrast images was 40.0 msec (TR/TE = 2.86/1.93 msec, 10 degrees flip angle, 7 spokes per image) for a nominal in-plane resolution of 1.3 mm and a section thickness of 6 mm. Quantitative evaluations of spatially averaged flow velocities and stroke volumes were comparable for real-time and cine methods when real-time MRI data were averaged across heartbeats. For individual heartbeats real-time phase-contrast MRI resulted in higher peak velocities for values above 120 cm s(-1). Conclusion: Real-time phase-contrast MRI of blood flow in the human aorta yields functional parameters for individual heartbeats. When averaged across heartbeats real-time flow velocities and stroke volumes are comparable to values obtained by conventional cine MRI.
引用
收藏
页码:206 / 213
页数:8
相关论文
共 26 条
  • [1] STATISTICAL METHODS FOR ASSESSING AGREEMENT BETWEEN TWO METHODS OF CLINICAL MEASUREMENT
    BLAND, JM
    ALTMAN, DG
    [J]. LANCET, 1986, 1 (8476) : 307 - 310
  • [2] Phase-contrast echo-planar MR imaging: Real-time quantification of flow and velocity patterns in the thoracic vessels induced by Valsalva's maneuver
    Eichenberger, AC
    Schwitter, J
    Mckinnon, GC
    Debatin, JF
    vonSchulthess, GK
    [J]. JMRI-JOURNAL OF MAGNETIC RESONANCE IMAGING, 1995, 5 (06): : 648 - 655
  • [3] Applications of phase-contrast flow and velocity imaging in cardiovascular MRI
    Gatehouse, PD
    Keegan, J
    Crowe, LA
    Masood, S
    Mohiaddin, RH
    Kreitner, KF
    Firmin, DN
    [J]. EUROPEAN RADIOLOGY, 2005, 15 (10) : 2172 - 2184
  • [4] REAL-TIME FLOOD FLOW IMAGING BY SPIRAL SCAN PHASE-VELOCITY MAPPING
    GATEHOUSE, PD
    FIRMIN, DN
    COLLINS, S
    LONGMORE, DB
    [J]. MAGNETIC RESONANCE IN MEDICINE, 1994, 31 (05) : 504 - 512
  • [5] Flow measurement by cardiovascular magnetic resonance: a multi-centre multi-vendor study of background phase offset errors that can compromise the accuracy of derived regurgitant or shunt flow measurements
    Gatehouse, Peter D.
    Rolf, Marijn P.
    Graves, Martin J.
    Hofman, Mark B. M.
    Totman, John
    Werner, Beat
    Quest, Rebecca A.
    Liu, Yingmin
    von Spiczak, Jochen
    Dieringer, Matthias
    Firmin, David N.
    van Rossum, Albert
    Lombardi, Massimo
    Schwitter, Juerg
    Schulz-Menger, Jeanette
    Kilner, Philip J.
    [J]. JOURNAL OF CARDIOVASCULAR MAGNETIC RESONANCE, 2010, 12
  • [6] Real-time phase-contrast MRI of cardiovascular blood flow using undersampled radial fast low-angle shot and nonlinear inverse reconstruction
    Joseph, Arun A.
    Merboldt, Klaus-Dietmar
    Voit, Dirk
    Zhang, Shuo
    Uecker, Martin
    Lotz, Joachim
    Frahm, Jens
    [J]. NMR IN BIOMEDICINE, 2012, 25 (07) : 917 - 924
  • [7] Magnetic resonance flow measurements in real time: Comparison with a standard gradient-echo technique
    Klein, C
    Schalla, S
    Schnackenburg, B
    Bornstedt, A
    Fleck, E
    Nagel, E
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2001, 14 (03) : 306 - 310
  • [8] Flow volume and shunt quantification in pediatric congenital heart disease by real-time magnetic resonance velocity mapping -: A validation study
    Körperich, H
    Gieseke, J
    Barth, P
    Hoogeveen, R
    Esdorn, H
    Peterschröder, A
    Meyer, H
    Beerbaum, P
    [J]. CIRCULATION, 2004, 109 (16) : 1987 - 1993
  • [9] In vitro validation of phase-contrast flow measurements at 3 T in comparison to 1.5 T:: Precision, accuracy, and signal-to-noise ratios
    Lotz, J
    Döker, R
    Noeske, R
    Schüttert, M
    Felix, R
    Galanski, M
    Gutberlet, M
    Meyer, GP
    [J]. JOURNAL OF MAGNETIC RESONANCE IMAGING, 2005, 21 (05) : 604 - 610
  • [10] Cardiovascular flow measurement with phase-contrast MR imaging: Basic facts and implementation
    Lotz, J
    Meier, C
    Leppert, A
    Galanski, M
    [J]. RADIOGRAPHICS, 2002, 22 (03) : 651 - 671