Evaluation of a pulse oximeter and customized probe for pulp vitality testing

被引:26
作者
Kahan, RS
Gulabivala, K
Snook, M
Setchell, DJ
机构
[1] UCL EASTMAN DENT INST, DEPT CONSERVAT DENT, LONDON, ENGLAND
[2] CLIN RES CTR, DIV BIOENGN, HARROW HA1 3UJ, MIDDX, ENGLAND
关键词
D O I
10.1016/S0099-2399(96)80283-4
中图分类号
R78 [口腔科学];
学科分类号
1003 ;
摘要
Current routine methods for assessment of pulp vitality rely on stimulation of A delta nerve fibers and give no direct indication of blood flow within the pulp. Recent papers have suggested that pulse oximeters may be used to diagnose pulp vitality by detection of blood flow. In this study, an optimized pulse oximeter probe for teeth was designed, built, and tested using the Biox 3740 Oximeter (Ohmeda, Louisville, CO). Following preliminary in vitro tests, the probe was tested clinically. Pulse waveforms from maxillary and mandibular anterior teeth were noted. Simultaneous readings from the subjects' finger were used as controls. Pulse wave readings from the teeth were found to be synchronous with the finger probe, but not consistently. It was easier to maintain continuous readings from mandibular incisors than from maxillary incisors. The average percentage synchronization with the pulse was 28.95% for maxillary incisors and 50.28% for mandibular incisors. This difference was significant (p = 0.05). The overall accuracy of the commercial instrument was disappointing, and in its present form it was not considered to have predictable diagnostic value.
引用
收藏
页码:105 / 109
页数:5
相关论文
共 15 条
[1]  
ARMITAGE P, 1987, STAT METHODS MED RES, P526
[2]   THE ROLE AND METHODS OF PULP TESTING IN ORAL-DIAGNOSIS - A REVIEW [J].
CHAMBERS, IG .
INTERNATIONAL ENDODONTIC JOURNAL, 1982, 15 (01) :1-15
[3]   FALSELY NORMAL SATURATION READING WITH THE PULSE OXIMETER [J].
COSTARINO, AT ;
DAVIS, DA ;
KEON, TP .
ANESTHESIOLOGY, 1987, 67 (05) :830-831
[4]   OPTICAL ASSESSMENT OF DENTAL-PULP VITALITY [J].
DALEY, J ;
BOYD, E ;
COOPER, J ;
ODRISCOLL, P .
JOURNAL OF BIOMEDICAL ENGINEERING, 1988, 10 (02) :146-148
[5]  
GAZELIUS B, 1986, Endodontics and Dental Traumatology, V2, P219
[6]   NONINVASIVE PULSE OXIMETRY UTILIZING SKIN REFLECTANCE PHOTOPLETHYSMOGRAPHY [J].
MENDELSON, Y ;
OCHS, BD .
IEEE TRANSACTIONS ON BIOMEDICAL ENGINEERING, 1988, 35 (10) :798-805
[7]   DUAL WAVELENGTH SPECTROPHOTOMETRY AS A DIAGNOSTIC-TEST OF THE PULP CHAMBER CONTENTS [J].
NISSAN, R ;
TROPE, M ;
ZHANG, CD ;
CHANCE, B .
ORAL SURGERY ORAL MEDICINE ORAL PATHOLOGY ORAL RADIOLOGY AND ENDODONTOLOGY, 1992, 74 (04) :508-514
[8]  
Payne JP, 1986, PULSE OXIMETRY
[9]  
Poswillo D, 1990, General anaesthesia, sedation and resuscitation in dentistry: report of an expert working party of the Standing Dental Advisory Committee
[10]   THE ASSESSMENT OF PULPAL VITALITY [J].
ROWE, AHR ;
FORD, TRP .
INTERNATIONAL ENDODONTIC JOURNAL, 1990, 23 (02) :77-83