Continuous intraoperative noninvasive cardiac output monitoring using a new thoracic bioimpedance device

被引:23
作者
Thangathurai, D [1 ]
Charbonnet, C [1 ]
Roessler, P [1 ]
Wo, CCJ [1 ]
Mikhail, M [1 ]
Yoshida, R [1 ]
Shoemaker, WC [1 ]
机构
[1] UNIV SO CALIF,SCH MED,DEPT ANESTHESIOL,LOS ANGELES,CA
关键词
measurement of cardiac output; thoracic bioimpedance; noninvasive cardiac output measurement; thermodilution cardiac output; continuous hemodynamic monitoring; high-risk surgery; pulmonary artery catheterization; ELECTRICAL BIOIMPEDANCE; IMPEDANCE CARDIOGRAPHY; THERMODILUTION;
D O I
10.1016/S1053-0770(97)90052-5
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
Objectives: To compare a new noninvasive bioimpedance device with the standard thermodilution method during the intraoperative period in high-risk patients undergoing oncological surgery. Design: Prospectively collected data with retrospective analysis. Setting: The study was undertaken at a university hospital, single institution. Participants: Twenty-three selected adults undergoing extensive, ablative oncological surgery. Interventions: Simultaneous measurements of cardiac output by a new bioimpedance method and the standard thermodilution method during the intraoperative and immediate postoperative periods. Measurements and Main Results: The correlation coefficient between the two methods was r = 0.89, p < 0.001. Bias and precision analysis between the two techniques showed a mean bias of 0.1 L/min and SD of the bias (precision) of 1.0 L/min (95% level of agreement +2.1 L/min to -1.9 L/min). After software enhancement, data from the last 11 monitored patients showed improved correlation between the two methods; r = 0.93, mean bias -0.1 L/min, and precision 0.8 L/min,Electrical and motion induced interference only transiently impaired the performance of the new impedance method. Conclusion: This new impedance device is a safe, reliable, clinically acceptable alternative to the invasive thermodilution method in the operating room environment. Copyright (C) 1997 by W.B. Saunders Company.
引用
收藏
页码:440 / 444
页数:5
相关论文
共 20 条
[1]   COMPARISON OF MEASUREMENTS OF CARDIAC-OUTPUT BY BIOIMPEDANCE AND THERMODILUTION IN SEVERELY ILL SURGICAL PATIENTS [J].
APPEL, PL ;
KRAM, HB ;
MACKABEE, J ;
FLEMING, AW ;
SHOEMAKER, WC .
CRITICAL CARE MEDICINE, 1986, 14 (11) :933-935
[2]   CONTINUOUS NONINVASIVE REAL-TIME MONITORING OF STROKE VOLUME AND CARDIAC-OUTPUT BY THORACIC ELECTRICAL BIOIMPEDANCE [J].
BERNSTEIN, DP .
CRITICAL CARE MEDICINE, 1986, 14 (10) :898-901
[3]   RELATIONSHIP BETWEEN SUPRANORMAL CIRCULATORY VALUES, TIME DELAYS, AND OUTCOME IN SEVERELY TRAUMATIZED PATIENTS [J].
BISHOP, MH ;
SHOEMAKER, WC ;
APPEL, PL ;
WO, CJ ;
ZWICK, C ;
KRAM, HB ;
MEADE, P ;
KENNEDY, F ;
FLEMING, AW .
CRITICAL CARE MEDICINE, 1993, 21 (01) :56-63
[4]  
BLAND JM, 1986, LANCET, V1, P407
[5]   MEASUREMENT OF CARDIAC-OUTPUT BY IMPEDANCE CARDIOGRAPHY UNDER VARIOUS CONDITIONS [J].
BOER, P ;
ROOS, JC ;
GEYSKES, GG ;
DORHOUTMEES, EJ .
AMERICAN JOURNAL OF PHYSIOLOGY, 1979, 237 (04) :H491-H496
[6]   DETERMINATION OF CARDIAC-OUTPUT DURING POSITIVE ENDEXPIRATORY PRESSURE - NONINVASIVE ELECTRICAL BIOIMPEDANCE COMPARED WITH STANDARD THERMODILUTION [J].
CASTOR, G ;
MOLTER, G ;
HELMS, J ;
NIEDERMARK, I ;
ALTMAYER, P .
CRITICAL CARE MEDICINE, 1990, 18 (05) :544-546
[7]   CARDIAC-OUTPUT MEASUREMENT IN CRITICAL CARE PATIENTS - THORACIC ELECTRICAL BIOIMPEDANCE VERSUS THERMODILUTION [J].
CLANCY, TV ;
NORMAN, K ;
REYNOLDS, R ;
COVINGTON, D ;
MAXWELL, JG .
JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1991, 31 (08) :1116-1121
[8]   PREDICTORS OF BETWEEN-METHOD DIFFERENCES IN CARDIAC-OUTPUT MEASUREMENT USING THORACIC ELECTRICAL BIOIMPEDANCE AND THERMODILUTION [J].
DOERING, L ;
LUM, E ;
DRACUP, K ;
FRIEDMAN, A .
CRITICAL CARE MEDICINE, 1995, 23 (10) :1667-1673
[9]   COMPARISON OF TRANS-THORACIC ELECTRICAL-IMPEDANCE AND THERMODILUTION METHODS FOR MEASURING CARDIAC-OUTPUT [J].
DONOVAN, KD ;
DOBB, GJ ;
WOODS, WPD ;
HOCKINGS, BE .
CRITICAL CARE MEDICINE, 1986, 14 (12) :1038-1044
[10]   MEASUREMENT OF CARDIAC-OUTPUT IN PREGNANCY BY THORACIC ELECTRICAL BIOIMPEDANCE AND THERMODILUTION - A PRELIMINARY-REPORT [J].
MASAKI, DI ;
GREENSPOON, JS ;
OUZOUNIAN, JG .
AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1989, 161 (03) :680-684