The accuracy of transcutaneous bilirubin to identify hyperbilirubinemia in jaundiced neonates

被引:8
作者
Ying, Qian [1 ]
You, Xueqin [1 ]
You, Jingye [1 ]
Wang, Jimei [1 ]
机构
[1] Fudan Univ, Obstet & Gynecol Hosp, Shanghai, Peoples R China
关键词
Transcutaneous bilirubin; jaundice; JM-103; hyperbilirubinemia; neonate; SERUM BILIRUBIN; OUTPATIENT; NEWBORNS;
D O I
10.1080/14767058.2020.1849112
中图分类号
R71 [妇产科学];
学科分类号
100211 ;
摘要
Objective To compare transcutaneous bilirubin (TcB) to total serum bilirubin (TSB) in jaundiced neonates by using a JM-103 bilirubinometer. Methods We conducted a retrospective study in jaundiced infants >= 35 weeks' gestation. Infants' and maternal information, TcB measurements, and TSB levels were collected. The correlation and consistency between TcB and TSB were analyzed by Pearson's correlation analysis and the Bland-Altman plot. Multivariate regression analysis was used to identify characteristics associated with the difference between TcB and TSB, underestimation by TcB and overestimation by TcB. Results A total of 787 paired samples were collected. There was a strong correlation between TcB and TSB with a Pearson correlation coefficient of 0.708 (p < .001). In preterm infants, outpatients and infants with high TSB levels, the TcB measurements were less precise. The TcB-TSB difference was significantly associated with gender, maternal hypertension disease, ABO hemolytic disease, and outpatient setting (p < .05). Only the outpatient setting had statistical significance in the underestimation of TSB >= 2 mg/dL. Conclusions Generally, TcB by a JM-103 bilirubinometer had a good correlation and consistency with the TSB values. However, small for gestational age, postnatal age of hour, ABO hemolytic disease, and outpatient setting may lead to inconsistencies between the TcB and TSB.
引用
收藏
页码:4318 / 4325
页数:8
相关论文
共 26 条
[1]   Transcutaneous bilirubinometry is not influenced by term or skin color in neonates [J].
Afanetti, Mickael ;
Trolli, Sergio Eleni Dit ;
Yousef, Nadya ;
Jrad, Ikram ;
Mokhtari, Mostafa .
EARLY HUMAN DEVELOPMENT, 2014, 90 (08) :417-420
[2]   The accuracy of transcutaneous bilirubinometry in preterm infants [J].
Arman, Didem ;
Topcuoglu, Sevilay ;
Gursoy, Tugba ;
Ovali, Fahri ;
Karatekin, Guner .
JOURNAL OF PERINATOLOGY, 2020, 40 (02) :212-218
[3]   Limitations and Opportunities of Transcutaneous Bilirubin Measurements [J].
Bosschaart, Nienke ;
Kok, Joke H. ;
Newsum, Astrid M. ;
Ouweneel, Dagmar M. ;
Mentink, Rosaline ;
van Leeuwen, Ton G. ;
Aalders, Maurice C. G. .
PEDIATRICS, 2012, 129 (04) :689-694
[4]   PALMITATE BINDING TO SERUM-ALBUMIN, MEASURED BY RATE OF DIALYSIS [J].
BRODERSEN, R ;
HONORE, B ;
ANDERSEN, S .
EUROPEAN JOURNAL OF BIOCHEMISTRY, 1988, 174 (01) :45-50
[5]   Accuracy of transcutaneous bilirubin measurement in preterm low-birth-weight neonates [J].
Chawla, Deepak ;
Jain, Suksham ;
Kaur, Gurjit ;
Sinhmar, Vikas ;
Guglani, Vishal .
EUROPEAN JOURNAL OF PEDIATRICS, 2014, 173 (02) :173-179
[6]  
Chimhini GLT, 2018, MATERN HLTH NEONATOL, V4
[7]  
Dalia S, 2009, MEDICINA, V45
[8]   The accuracy of transcutaneous bilirubinometer measurements to identify the hyperbilirubinemia in outpatient newborn population [J].
Ercan, Serif ;
Ozgun, Gunay .
CLINICAL BIOCHEMISTRY, 2018, 55 :69-74
[9]   A Prospective Comparison of Transcutaneous and Serum Bilirubin Within Brief Time Intervals [J].
Jones, Denise F. ;
McRea, Abigail R. ;
Knowles, James D. ;
Lin, Feng-Chang ;
Burnette, Erin ;
Reller, Lara A. ;
Lohr, Jacob A. .
CLINICAL PEDIATRICS, 2017, 56 (11) :1013-1017
[10]   Accuracy and precision of transcutaneous bilirubinometry in postdischarge Asian neonates [J].
Kitsommart, Ratchada ;
Pornladnun, Pornpat ;
Chomchai, Chulathida ;
Urujchutchairut, Pornvilai ;
Paes, Bosco .
EUROPEAN JOURNAL OF PEDIATRICS, 2013, 172 (06) :781-786