Evaluating the effect of pupil dilation on spectral-domain optical coherence tomography measurements and their quality score

被引:14
|
作者
Tanga, Lucia [1 ]
Roberti, Gloria [1 ]
Oddone, Francesco [1 ]
Quaranta, Luciano [2 ]
Ferrazza, Manuela [1 ]
Berardo, Francesca [1 ]
Manni, Gianluca [3 ]
Centofanti, Marco [1 ,3 ]
机构
[1] IRCCS Fdn GB Bietti, I-00198 Rome, Italy
[2] Univ Brescia, USVD Ctr Studio Glaucoma, DSMC, I-25123 Brescia, Italy
[3] Univ Roma Tor Vergata, DSCMT, I-00133 Rome, Italy
关键词
NERVE-FIBER LAYER; MACULAR THICKNESS MEASUREMENTS; TIME-DOMAIN; REPRODUCIBILITY; CATARACT; SIZE; HEAD; EYES;
D O I
10.1186/s12886-015-0168-y
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
Background: Spectral-domain optical coherence tomography (SD-OCT) provides fast scan speed and high scan resolution improving its diagnostic accuracy. The purpose of this study was to evaluate if SD-OCT measurements and their quality score are influenced by pupil dilation. Methods: Retinal nerve fiber layer thickness (RNFL), ganglion cell complex (GCC) and optic nerve head (ONH) were measured in one eye of 57 glaucoma patients and 36 healthy subjects using spectral domain optical coherence tomography (SD-OCT) before and after pupil dilation. Comparisons were made between measurements and their quality score pre- and post dilation (Signal Strength Index, SSI). Overall RNFL, average GCC and ONH rim volume were considered in the analysis. Results: No statistically significant differences were found between pre- and post-dilation measurements in both groups (glaucoma: RNFL 80 +/- 15 mu m vs 80 +/- 16 mu m, p = 0.87; GCC 81.35 +/- 13.4 mu m vs 81.10 +/- 13.14 mu m, p = 0.92; ONH 0.05 +/- 0.11 mm(3) vs 0.04 +/- 0.07 mm(3), p = 0.74; controls RNFL 99 +/- 12 mu m vs 98 +/- 14 mu m, p = 0.70; GCC 92.12 +/- 6.7 mu m vs 91.54 +/- 7.05 mu m, p = 0.72; ONH 0.11 +/- 0.1 mm(3) vs 0.04 +/- 0.07 mm(3), p = 0.36) nor between pre- and post-dilation quality score (glaucoma SSI RNFL 54.3 +/- 10.3 vs 51.7 +/- 18.1, p = 0.12; SSI GCC 58 +/- 9.5 vs 57 +/- 8.09, p = 0.55; SSI ONH 48.5 +/- 7.6 vs 46.6 +/- 7.2, p = 0.16; controls SSI RNFL 57 +/- 10.3 vs 54 +/- 9.31, p = 0.2; SSI GCC 60.9 +/- 8.1 vs 58.8 +/- 7.3, p = 0.3; SSI ONH 51.5 +/- 8.9 vs 50.4 +/- 8.3, p = 0.59). Conclusion: Pupil dilation doesn't affect SD-OCT measurements and their quality score.
引用
收藏
页数:5
相关论文
共 50 条
  • [21] Image sensor for spectral-domain optical coherence tomography on a chip
    Vlaskovic, M.
    Zimmermann, H.
    Meinhardt, G.
    Kraft, J.
    ELECTRONICS LETTERS, 2020, 56 (24) : 1306 - 1309
  • [22] Correlation in foveal thickness measurements between spectral-domain and time-domain optical coherence tomography in normal individuals
    Carpineto, P.
    Nubile, M.
    Toto, L.
    Gnama, A. Aharrh
    Marcucci, L.
    Mastropasqua, L.
    Ciancaglini, M.
    EYE, 2010, 24 (02) : 251 - 258
  • [23] Spectral-domain Optical Coherence Tomography Measurement of Macular Volume for Diagnosing Glaucoma
    Mori, Satoshi
    Hangai, Masanori
    Sakamoto, Atsushi
    Yoshimura, Nagahisa
    JOURNAL OF GLAUCOMA, 2010, 19 (08) : 528 - 534
  • [24] Spectral-Domain Optical Coherence Tomography Findings in Patients With Macropsia
    Jeon, Sohee
    Jung, Byungju
    Lee, Won Ki
    OPHTHALMIC SURGERY LASERS & IMAGING RETINA, 2018, 49 (09) : 656 - 663
  • [25] Variability in Spectral-Domain Optical Coherence Tomography over 4 Weeks by Age
    Howard, Kerri P.
    Chandler, Charles S.
    Huang, Yijun
    Klein, Barbara E. K.
    Lee, Kristine E.
    Meuer, Stacy M.
    Danis, Ronald P.
    Danforth, Lorraine G.
    Gangnon, Ronald E.
    Klein, Ronald
    OPHTHALMIC EPIDEMIOLOGY, 2016, 23 (03) : 193 - 201
  • [26] Spectral-Domain Optical Coherence Tomography and Scanning Laser Polarimetry in Glaucoma Diagnosis
    Lee, Suhwan
    Sung, Kyung Rim
    Cho, Jung Woo
    Cheon, Mi Hyun
    Kang, Sung Yong
    Kook, Michael S.
    JAPANESE JOURNAL OF OPHTHALMOLOGY, 2010, 54 (06) : 544 - 549
  • [27] Comparison of the influence of cataract and pupil size on retinal nerve fibre layer thickness measurements with time-domain and spectral-domain optical coherence tomography
    Cheng, Clarissa S. M.
    Natividad, Maricel G.
    Earnest, Arul
    Yong, Vernon
    Lim, Boon A.
    Wong, Hon T.
    Yip, Leonard W.
    CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2011, 39 (03) : 215 - 221
  • [28] Macular spectral-domain optical coherence tomography values and correlations in healthy children
    Passani, Andrea
    Sframeli, Angela Tindara
    Posarelli, Chiara
    Lisi, Domenico
    Guidi, Gianluca
    Casini, Giamberto
    Ferreras, Antonio
    Figus, Michele
    INTERNATIONAL OPHTHALMOLOGY, 2019, 39 (11) : 2449 - 2457
  • [29] Macular Imaging for Glaucoma Using Spectral-domain Optical Coherence Tomography: A Review
    Wong, Jessica J.
    Chen, Teresa C.
    Shen, Lucy Q.
    Pasquale, Louis R.
    SEMINARS IN OPHTHALMOLOGY, 2012, 27 (5-6) : 160 - 166
  • [30] Comparison of manually corrected retinal thickness measurements from multiple spectral-domain optical coherence tomography instruments
    Heussen, Florian M.
    Ouyang, Yanling
    McDonnell, Emma C.
    Narala, Ramsudha
    Ruiz-Garcia, Humberto
    Walsh, Alexander C.
    Sadda, SriniVas R.
    BRITISH JOURNAL OF OPHTHALMOLOGY, 2012, 96 (03) : 380 - 385