Pulsed amperometric detection of ethanol in breath by gold electrodes supported on ion exchange membranes (solid polymer electrolytes)

被引:40
作者
Schiavon, G
Comisso, N
Toniolo, R
Bontempelli, G
机构
[1] CNR,IPELP,I-35020 PADUA,ITALY
[2] UNIV UDINE,DEPT CHEM SCI & TECHNOL,I-33100 UDINE,ITALY
关键词
ethanol; breath; solid polymer electrolyte; pulsed amperometric detection;
D O I
10.1002/elan.1140080609
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
A sensitive and fast-responding amperometric sensor for the determination of ethanol in breath which eliminates oxygen interferences is described. It consists of a porous gold working electrode (facing the sample) supported on one face of an ion-exchange membrane which serves as a solid polymer electrolyte (SPE). The other side of the membrane faces an internal electrolyte solution (1 M aqueous NaOH) containing the counter and reference electrodes. The fouling of the gold working electrode, due to the ethanol oxidation process, is overcome by continuously cleaning and reactivating its surface by applying a triple-pulse potential waveform. The performance of this sensor has been tested in a closed-loop flow apparatus fed with synthetic ethanol-vapor samples obtained by evaporating aqueous solutions with known ethanol contents. The sensor displays a high current sensitivity and a low background noise, so that a low detection limit (2 mu g L(-1), 1 ppm v./v.) is estimated for a signal-to-noise ratio of 3. The responses are characterized by both a good reproducibility and a linear dependence on the ethanol concentration over a fairly wide range (more than three orders of magnitude), as well as by a short response time (ca. 1 s). This fast response time arises from the lack of a gas-permeable membrane and direct gas contact to the triple interphase among the gaseous analyte, the porous working electrode and the solid polymer electrolyte. The absence of significant interfering effects and the possibility of adopting such a device for the detection of ethanol in fermentation processes are also discussed.
引用
收藏
页码:544 / 548
页数:5
相关论文
共 36 条
[1]  
ANDREWS SA, 1989, ANAL TOXICOL, V113, P113
[2]   CONDUCTIVITY RESPONSE OF POROUS-ELECTRODES SUPPORTED ON PERFLUOROSULFONIC ACID MEMBRANES TO ACIDIC GAS-MIXTURES [J].
BETTELHEIM, A ;
HARTH, R ;
OZER, D ;
MOR, U ;
SEGAL, B .
ANALYTICAL CHEMISTRY, 1991, 63 (23) :2724-2727
[3]   A STUDY OF THE ELECTROCATALYTIC BEHAVIOR OF GOLD IN BASE USING AC VOLTAMMETRY [J].
BURKE, LD ;
OSULLIVAN, JF .
ELECTROCHIMICA ACTA, 1992, 37 (04) :585-594
[4]   MGO-DOPED CR2O3 - SOLUBILITY LIMIT AND THE EFFECT OF DOPING ON THE RESISTIVITY AND ETHANOL SENSITIVITY [J].
CHEN, ZX ;
COLBOW, K .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 9 (01) :49-53
[5]  
CRIDDLE W J, 1991, Analytical Proceedings, V28, P338
[6]   APPLICATION OF NAFION PLATINUM-ELECTRODES (SOLID POLYMER ELECTROLYTE STRUCTURES) TO VOLTAMMETRIC INVESTIGATIONS OF HIGHLY RESISTIVE SOLUTIONS [J].
DEWULF, DW ;
BARD, AJ .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1988, 135 (08) :1977-1985
[7]   APPLICATION OF NAFION AS A POLYMER SOLID ELECTROLYTE FOR VOLTAMMETRY IN THE ABSENCE OF A CONTACTING ELECTROLYTE SOLUTION [J].
HARTH, R ;
MOR, U ;
OZER, D ;
BETTELHEIM, A .
JOURNAL OF THE ELECTROCHEMICAL SOCIETY, 1989, 136 (12) :3863-3867
[8]   GAS-DEPENDENT RESPONSE IN THE TEMPERATURE TRANSIENT OF SNO2 GAS SENSORS [J].
HIRANAKA, Y ;
ABE, T ;
MURATA, H .
SENSORS AND ACTUATORS B-CHEMICAL, 1992, 9 (03) :177-182
[9]   TRIPLE-PULSE AMPEROMETRIC DETECTION OF CARBOHYDRATES AFTER CHROMATOGRAPHIC-SEPARATION [J].
HUGHES, S ;
JOHNSON, DC .
ANALYTICA CHIMICA ACTA, 1983, 149 (MAY) :1-10
[10]   AMPEROMETRIC DETECTION OF SIMPLE ALCOHOLS IN AQUEOUS-SOLUTIONS BY APPLICATION OF A TRIPLE-PULSE POTENTIAL WAVEFORM AT PLATINUM-ELECTRODES [J].
HUGHES, S ;
MESCHI, PL ;
JOHNSON, DC .
ANALYTICA CHIMICA ACTA, 1981, 132 (DEC) :1-10