Image quality in thoracic 4D cone-beam CT: A sensitivity analysis of respiratory signal, binning method, reconstruction algorithm, and projection angular spacing

被引:30
作者
Shieh, Chun-Chien [1 ,2 ]
Kipritidis, John [1 ]
O'Brien, Ricky T. [1 ]
Kuncic, Zdenka [2 ]
Keall, Paul J. [1 ]
机构
[1] Univ Sydney, Radiat Phys Lab, Sydney Med Sch, Sydney, NSW 2006, Australia
[2] Univ Sydney, Sch Phys, Inst Med Phys, Sydney, NSW 2006, Australia
关键词
4D-CBCT; reconstruction; image quality; LUNG-TUMOR MOTION; COMPUTED-TOMOGRAPHY; MARKERS;
D O I
10.1118/1.4868510
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: Respiratory signal, binning method, and reconstruction algorithm are three major controllable factors affecting image quality in thoracic 4D cone-beam CT (4D-CBCT), which is widely used in image guided radiotherapy (IGRT). Previous studies have investigated each of these factors individually, but no integrated sensitivity analysis has been performed. In addition, projection angular spacing is also a key factor in reconstruction, but how it affects image quality is not obvious. An investigation of the impacts of these four factors on image quality can help determine the most effective strategy in improving 4D-CBCT for IGRT. Methods: Fourteen 4D-CBCT patient projection datasets with various respiratory motion features were reconstructed with the following controllable factors: (i) respiratory signal (real-time position management, projection image intensity analysis, or fiducial marker tracking), (ii) binning method (phase, displacement, or equal-projection-density displacement binning), and (iii) reconstruction algorithm [Feldkamp-Davis-Kress (FDK), McKinnon-Bates (MKB), or adaptive-steepest-descent projection-onto-convex-sets (ASD-POCS)]. The image quality was quantified using signal-to-noise ratio (SNR), contrast-to-noise ratio, and edge-response width in order to assess noise/streaking and blur. The SNR values were also analyzed with respect to the maximum, mean, and root-meansquared-error (RMSE) projection angular spacing to investigate how projection angular spacing affects image quality. Results: The choice of respiratory signals was found to have no significant impact on image quality. Displacement-based binning was found to be less prone to motion artifacts compared to phase binning in more than half of the cases, but was shown to suffer from large interbin image quality variation and large projection angular gaps. Both MKB and ASD-POCS resulted in noticeably improved image quality almost 100% of the time relative to FDK. In addition, SNR values were found to increase with decreasing RMSE values of projection angular gaps with strong correlations (r approximate to -0.7) regardless of the reconstruction algorithm used. Conclusions: Based on the authors' results, displacement-based binning methods, better reconstruction algorithms, and the acquisition of even projection angular views are the most important factors to consider for improving thoracic 4D-CBCT image quality. In view of the practical issues with displacement-based binning and the fact that projection angular spacing is not currently directly controllable, development of better reconstruction algorithms represents the most effective strategy for improving image quality in thoracic 4D-CBCT for IGRT applications at the current stage. (c) 2014 American Association of Physicists in Medicine.
引用
收藏
页数:18
相关论文
共 24 条
  • [21] Common-mask guided image reconstruction (c-MGIR) for enhanced 4D cone-beam computed tomography
    Park, Justin C.
    Zhang, Hao
    Chen, Yunmei
    Fan, Qiyong
    Li, Jonathan G.
    Liu, Chihray
    Lu, Bo
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (23) : 9157 - 9183
  • [22] Accuracy and inter-observer variability of 3D versus 4D cone-beam CT based image-guidance in SBRT for lung tumors
    Sweeney, Reinhart A.
    Seubert, Benedikt
    Stark, Silke
    Homann, Vanessa
    Mueller, Gerd
    Flentje, Michael
    Guckenberger, Matthias
    RADIATION ONCOLOGY, 2012, 7
  • [23] Compressed-sensing (CS)-based 3D image reconstruction in cone-beam CT (CBCT) for low-dose, high-quality dental X-ray imaging
    Lee, M. S.
    Kim, H. J.
    Cho, H. S.
    Hong, D. K.
    Je, U. K.
    Oh, J. E.
    Park, Y. O.
    Lee, S. H.
    Cho, H. M.
    Choi, S. I.
    Koo, Y. S.
    JOURNAL OF THE KOREAN PHYSICAL SOCIETY, 2013, 63 (05) : 1066 - 1071
  • [24] Compressed-sensing (CS)-based 3D image reconstruction in cone-beam CT (CBCT) for low-dose, high-quality dental X-ray imaging
    M. S. Lee
    H. J. Kim
    H. S. Cho
    D. K. Hong
    U. K. Je
    J. E. Oh
    Y. O. Park
    S. H. Lee
    H. M. Cho
    S. I. Choi
    Y. S. Koo
    Journal of the Korean Physical Society, 2013, 63 : 1066 - 1071