The measurement of alkaline phosphatase at nanomolar concentration within 70 s using a disposable microelectrochemical transistor

被引:13
作者
Astier, Y [1 ]
Bartlett, PN [1 ]
机构
[1] Univ Southampton, Sch Chem, Southampton SO17 1BJ, Hants, England
关键词
alkaline phosphatase; nanomolar concentration; disposable microelectrochemical transistor;
D O I
10.1016/j.bioelechem.2004.01.004
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
We report a new approach to the measurement of alkaline phosphatase concentration based on the use of a disposable poly(aniline) microelectrochemical transistor. The measurement is carried out in a two cell configuration in which the poly(aniline) microelectrochemical transistor operates in acid solution and is connected to the alkaline buffer solution containing the analyte by a salt bridge. Disposable microelectrochemical transistors were reproducibly fabricated by electrochemical deposition of poly(aniline) onto photolithographically fabricated gold microband arrays. Using these devices alkaline phosphatase was detected by employing p-aminophenyl phosphate as the substrate for the enzyme and using glucose and glucose oxidase to recycle the p-aminophenol generated upon enzyme catalysed hydrolysis of the phosphate. Recycling the p-aminophenol with glucose and glucose oxidase amplified the detection of alkaline phosphatase approximately tenfold. Using this approach we obtain linear calibration curves for alkaline phosphatase up to 5 nM within 70 s on single use devices. (C) 2004 Elsevier B.V. All rights reserved.
引用
收藏
页码:53 / 59