Fatigue behaviors of different materials for schanz screws in femoral fracture model using finite element analysis

被引:0
|
作者
Gok, Arif [1 ]
Inal, Sermet [2 ]
Taspinar, Ferruh
Gulbandilar, Eyyup
Gok, Kadir
机构
[1] Amasya Univ, Fac Technol, Dept Mech Engn, TR-05100 Amasya, Turkey
[2] Dumlupinar Univ, Sch Med, Dept Orthopaed & Traumatol, TR-43100 Kutahya, Turkey
关键词
Hip fracture; Bone screw; Titanium; Stainless steel; Materials failure; Finite element analysis; PERTROCHANTERIC FRACTURES; HIP-PROSTHESIS; COMPONENT; DESIGN;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In the daily life, human beings may be met with a number of undesired traumas. As a result, intertrochanteric fractures can occur in the skeletal system. Intertrochanteric femoral fractures are serious traumas that can lead to pneumonia, pulmonary embolism or death. Therefore, fixing as accuracy and stabilty of these fractures is necessary. The schanz screws with pertrochanteric fixator body are used for stabilizaiton of intertrochanteric fractures. Fatigue behaviors of the schanz screws must be longer than the duration of fracture consolidation. The aim of this study is to determine whether there are differences between fatigue behaviours of different materials for schanz screws when they are used for intertrochanteric femoral fracture fixation. Fatigue behaviors of schanz screws used for fixing of intertrochanteric fractures were analysed with ANSYS Workbench software using finite element analysis under dynamic loads for stainless steel and titanium. After the analysis, minimum fatigue life has been determined as approximately 21 months for the schanz screws having stainless steel while titanium has been determined as approximately 2 months. In addition, safety factor for schanz screws having stainless steel is calculated as 1.36 whereas safety factor for titanium is calculated as 1. Therefore, considering the consolidation of an intertrochanteric femoral fracture with a mean duration of three months, the titanium schanz screws shouldn't be used since their fatigue behavior is two months and safety factor is less than stainless steel.
引用
收藏
页码:576 / 580
页数:5
相关论文
共 50 条
  • [1] A finite element analysis of a low-profile femoral neck system of screws in sleeves in a vertical femoral neck fracture model
    Sun, Jun
    Wu, Le
    Fang, Nan
    Qiao, Wenze
    Liu, Lifeng
    BMC MUSCULOSKELETAL DISORDERS, 2024, 25 (01)
  • [2] Fixation effects of different types of cannulated screws on vertical femoral neck fracture: A finite element analysis and experimental study
    Zhan, Shi
    Jiang, Dajun
    Ling, Ming
    Ding, Jian
    Yang, Kai
    Duan, Lei
    Tsai, Tsung-Yuan
    Feng, Yong
    van Trigt, Bart
    Jia, Weitao
    Zhang, Changqing
    Hu, Hai
    MEDICAL ENGINEERING & PHYSICS, 2021, 97 : 32 - 39
  • [3] Predicting femoral neck fracture using finite element analysis
    Testi, D
    Viceconti, M
    Cappello, A
    Gnudi, S
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING - 3, 2001, : 9 - 14
  • [4] Predicting the stresses and displacements in femoral intramedullary nails and screws using a finite element model
    Kralovic, Barbara J.
    Fryman, J. Craig
    Longsworth, Serene
    PROCEEDING OF THE ASME SUMMER BIOENGINEERING CONFERENCE - 2007, 2007, : 297 - 298
  • [5] Biomechanical analysis of cement materials for vertebroplasty using a finite element fracture model
    Wilcox, RK
    PROCEEDINGS OF THE THIRD IASTED INTERNATIONAL CONFERENCE ON BIOMECHANICS, 2005, : 67 - 71
  • [6] Finite element analysis of the effect of different configurations of internal fixation screws on the treatment of femoral neck fractures
    Gao, Yangbo
    Hu, Zehua
    Gao, Weitao
    Zhang, Kairan
    Zhongnan Daxue Xuebao (Ziran Kexue Ban)/Journal of Central South University (Science and Technology), 2024, 55 (09): : 3306 - 3314
  • [7] Finite Element Analysis for the Treatment of Proximal Femoral Fracture
    Hsu, Ching-Chi
    Lin, Jinn
    Amaritsakul, Yongyut
    Antonius, Takalamesar
    Chao, Ching-Kong
    CMC-COMPUTERS MATERIALS & CONTINUA, 2009, 11 (01): : 1 - 13
  • [8] Analysis of orthopedic screws for bone fracture fixations with finite element method
    School of Mechanical Engineering, Universiti Sains Malaysia, Engineering Campus, 14300 Nibong Tebal Seberang Perai, Malaysia
    不详
    J. Appl. Sci., 2007, 13 (1748-1754):
  • [9] Influence of element discretization types to fatigue behaviors in finite element analysis
    Bintoro, Sefrian Rizki
    Prabowo, Aditya Rio
    Muhayat, Nurul
    Muhayat, Nurul
    MATERIALS TODAY-PROCEEDINGS, 2022, 57 : 531 - 538
  • [10] Finite element analysis of fixation effect for femoral neck fracture under different fixation configurations
    Cui, Haipo
    Wei, Wenqing
    Shao, Yinlin
    Du, Kewei
    COMPUTER METHODS IN BIOMECHANICS AND BIOMEDICAL ENGINEERING, 2022, 25 (02) : 132 - 139