Fatigue behaviors of different materials for schanz screws in femoral fracture model using finite element analysis

被引:0
作者
Gok, Arif [1 ]
Inal, Sermet [2 ]
Taspinar, Ferruh
Gulbandilar, Eyyup
Gok, Kadir
机构
[1] Amasya Univ, Fac Technol, Dept Mech Engn, TR-05100 Amasya, Turkey
[2] Dumlupinar Univ, Sch Med, Dept Orthopaed & Traumatol, TR-43100 Kutahya, Turkey
来源
OPTOELECTRONICS AND ADVANCED MATERIALS-RAPID COMMUNICATIONS | 2014年 / 8卷 / 5-6期
关键词
Hip fracture; Bone screw; Titanium; Stainless steel; Materials failure; Finite element analysis; PERTROCHANTERIC FRACTURES; HIP-PROSTHESIS; COMPONENT; DESIGN;
D O I
暂无
中图分类号
T [工业技术];
学科分类号
08 ;
摘要
In the daily life, human beings may be met with a number of undesired traumas. As a result, intertrochanteric fractures can occur in the skeletal system. Intertrochanteric femoral fractures are serious traumas that can lead to pneumonia, pulmonary embolism or death. Therefore, fixing as accuracy and stabilty of these fractures is necessary. The schanz screws with pertrochanteric fixator body are used for stabilizaiton of intertrochanteric fractures. Fatigue behaviors of the schanz screws must be longer than the duration of fracture consolidation. The aim of this study is to determine whether there are differences between fatigue behaviours of different materials for schanz screws when they are used for intertrochanteric femoral fracture fixation. Fatigue behaviors of schanz screws used for fixing of intertrochanteric fractures were analysed with ANSYS Workbench software using finite element analysis under dynamic loads for stainless steel and titanium. After the analysis, minimum fatigue life has been determined as approximately 21 months for the schanz screws having stainless steel while titanium has been determined as approximately 2 months. In addition, safety factor for schanz screws having stainless steel is calculated as 1.36 whereas safety factor for titanium is calculated as 1. Therefore, considering the consolidation of an intertrochanteric femoral fracture with a mean duration of three months, the titanium schanz screws shouldn't be used since their fatigue behavior is two months and safety factor is less than stainless steel.
引用
收藏
页码:576 / 580
页数:5
相关论文
共 17 条
[1]   Clinical and finite element analysis of a modular femoral prosthesis consisting of a head and stem component in the treatment of pertrochanteric fractures [J].
Andress, HJ ;
Kahl, S ;
Kranz, C ;
Gierer, P ;
Schürmann, M ;
Lob, G .
JOURNAL OF ORTHOPAEDIC TRAUMA, 2000, 14 (08) :546-553
[2]   Fatigue analysis of cemented hip prosthesis: model definition and damage evolution algorithms [J].
Colombi, P .
INTERNATIONAL JOURNAL OF FATIGUE, 2002, 24 (08) :895-901
[3]   Material selection in the design of the femoral component of cemented total hip replacement [J].
El-Sheikh, HF ;
MacDonald, BJ ;
Hashmi, MSJ .
JOURNAL OF MATERIALS PROCESSING TECHNOLOGY, 2002, 122 (2-3) :309-317
[4]   The importance of lag screw position for the stabilization of trochanteric fractures with a sliding hip screw: A subject-specific finite element study [J].
Goffin, Jerome M. ;
Pankaj, Pankaj ;
Simpson, A. Hamish .
JOURNAL OF ORTHOPAEDIC RESEARCH, 2013, 31 (04) :596-600
[5]  
HALSEY D, 1992, CLIN ORTHOP RELAT R, P305
[6]   Finite element analysis of four different implants inserted in different positions to stabilize an idealized trochanteric femoral fracture [J].
Helwig, Peter ;
Faust, Gunter ;
Hindenlang, Ulrich ;
Hirschmueller, Anja ;
Konstantinidis, Lukas ;
Bahrs, Christian ;
Suedkamp, Norbert ;
Schneider, Ralf .
INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2009, 40 (03) :288-295
[7]  
Kasman R., 1984, J ORTHOP RES, V4, P377
[8]   Static, dynamic and fatigue behaviors of dental implant using finite element method [J].
Kayabasi, Oguz ;
Yuezbasioglu, Emir ;
Erzincanli, Fehmi .
ADVANCES IN ENGINEERING SOFTWARE, 2006, 37 (10) :649-658
[9]   Comparison of in situ and in vitro CT scan-based finite element model predictions of proximal femoral fracture load [J].
Keyak, JH ;
Falkinstein, Y .
MEDICAL ENGINEERING & PHYSICS, 2003, 25 (09) :781-787
[10]  
Mazzocca A. D., 2003, SKELETAL TRAUMA, V1